journal article·2014-10

Investigation of a fatigue failure in a stainless steel femoral plate

JB. Marcomini, C.A.R.P. Baptista, J.P. Pascon, R.L. Teixeira, F.P. Reis

Publisher page

Abstract

Surgical implants are exposed to severe working conditions and therefore a wide range of failure mechanisms may occur, including fatigue, corrosion, wear, fretting and combinations of them. The mechanical failures of metallic implants may also be influenced by several other factors, including the design, material, manufacturing, installation, postoperative complications and misuse. An 83-year-old patient suffered an oblique femoral shaft fracture due to a fall at home. A stainless steel locking compression plate (LCP) employed in the fracture reduction failed after four months and was sent back to the producer. A second LCP of the same type was implanted and also failed after six months. A failure analysis of the second femoral LCP is performed in this paper. The results demonstrate that poor material quality was decisive to the failure. The chemical analysis revealed a high P content in the steel, which is not in accordance to the standards. A combination of factors lead to LCP fracture and these include: brittle crack initiation due to phosphorus, segregation at grain boundaries, crack propagation due to cyclic loading and final fast fracture favored by the loss of ductility due to cold work.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Bone fractures and treatments · Fatigue and fracture mechanics

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (23)

Show source-provided references
1. ASME SA 240, 2010. Specification for Heat-resisting Chromium and Chromium–Nickel Stainless Steel Plate, Sheet, and Strip for Pressure Vessels.
2. ASTM A 240, 2007. Standard Specification for Chromium and Chromium–Nickel Stainless Steel Plate, Sheet, and Strip for Pressure Vessels and for General Applications.
3. ASTM A 276, 2006. Standard Specification for Stainless Steel Bars and Shapes.
4. ASTM A 370, 2007. Standard Test Methods and Definitions for Mechanical Testing of Steel Products.
5. ASTM A 262, 2005. Standard Practices for Detecting Susceptibility to Intergranular Attack in Austenitic Stainless Steels.
6. ASTM E 112, 2006. Standard Test Methods for Determining Average Grain Size.
7. ASTM F 138, 2013. Standard Specification for Wrought 18Chromium-14Nickel-2.5Molybdenum Stainless Steel Bar and Wire for Surgical Implants (UNS S31673).
8. ASTM F 139, 2012. Standard Specification for Wrought 18Chromium-14Nickel-2.5Molybdenum Stainless Steel Sheet and Strip for Surgical Implants(UNS S31673).
9. ASM HandbooK 2002. Failure Analysis and Prevention 11.
10. Failure analysis of surgical implants in Brazil10.1016/S1350-6307(02)00026-2 ↗
11. Carbide precipitation, grain boundary segregation and temper embrittlement in NiCrMoV rotor steels10.1007/BF02814824 ↗
12. Fractures of the shaft of the femur10.2106/00004623-199173100-00015 ↗
13. Castro, R. and Gueussier, A., Embrittlement of Austenitic Steels, Revue metallurgie, No. 8, 1960.
14. Diagnosing the cause of dynamic compression plate premature failure10.1016/j.engfailanal.2013.09.001 ↗
15. Effect of precipitates on the corrosion-fatigue crack initiation of ISSO 5832-9 stainless steel biomaterial10.1016/j.ijfatigue.2004.03.002 ↗
16. Epidemiology of diaphyseal femoral fracture10.3109/17453678609014762 ↗
17. Fatigue failure of an orthopedic implant—a locking compression plate10.1016/j.engfailanal.2007.04.001 ↗
18. Evolution of the internal fixation of long bone fractures10.1302/0301-620X.84B8.0841093 ↗
19. Case study: failure of high nitrogen stainless steel femoral stem10.1016/j.engfailanal.2011.11.010 ↗
20. Plate osteosynthesis of simple forearm fractures: LCP versus DC plates
21. Metallurgical investigation of a failure in 316L stainless steel orthopaedic implant10.1016/j.engfailanal.2004.05.004 ↗
22. National data of 6409 Swedish inpatients with femoral shaft fractures: stable incidence between 1998 and 200410.1016/j.injury.2008.07.017 ↗
23. Effect of cold working and sandblasting on the microhardness, tensile strength and corrosion resistance of AISI 316L stainless steel10.1007/s12613-012-0676-1 ↗