Clinically relevant mechanical testing of hernia graft constructs
Source abstract
Abstract
Background To understand the mechanical behavior of grafts in the context of hernia repair, there is a need to develop and adopt methods for mechanical testing of grafts in a clinically-relevant manner with clinically-relevant outcomes. Materials and methods Ball-burst and planar-biaxial methods were used to test three commercially-available hernia grafts (DermaMatrix, Biodesign, VitaMesh Blue). Both load-to-failure and cyclic fatigue tests were performed (n=6-11/group/test). Grafts were tested as sutured constructs in patch geometry. Dilatational strain analysis was performed considering the construct (both test methods) or the graft (planar-biaxial only) as the area of interest. Results DermaMatrix, Biodesign, and VitaMesh grafts showed differences in mechanical properties at the point of construct failure (load, in-plane load-per-suture and membrane tension) in ball-burst tests and differences in sub-failure properties (stiffness, dilatational strain at 16N/cm and cyclic mechanical properties) in planar-biaxial tests. In both load-to-failure and cyclic fatigue tests, each graft construct tended to be stiffer in planar-biaxial than ball-burst testing. In biaxial testing, the strain analysis method influenced the mechanical properties with the construct being more compliant than the graft. Conclusions This study demonstrates that graft-fixation method, test mode and analysis method are important considerations for mechanical characterization of hernia grafts. Ball-burst tests can only estimate construct or material properties, whereas planar-biaxial tests capture anisotropy and can estimate construct, graft and material properties of the same test specimen. When the clinical performance of a graft in the context of hernia repair is of interest, testing a sutured construct and performing construct strain analysis arguably provides the most clinically-relevant assessment method.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Hernia repair and management · Intestinal and Peritoneal Adhesions · Surgical Sutures and Adhesives
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (41)
Show source-provided references
1. ASTM Standard D3787-07. Standard Test Method for Bursting Strength of Textiles-Constant-Rate-of-Traverse (CRT) Ball Burst Test. 2011, doi:10.1520/D3787-07R11, www.astm.org.10.1520/D3787-07R11 ↗
2. ASTM Standard D6797-07. Standard Test Method for Bursting Strength of Fabrics Constant-Rate-of-Extension (CRE) Ball Burst Test. 2011, doi:10.1520/D6797-07R11, www.astm.org.10.1520/D6797-07R11 ↗
3. Mechanisms of ventral hernia recurrence after mesh repair and a new proposed classification10.1016/j.jamcollsurg.2005.02.035 ↗
4. Biaxial mechanical modeling of the small intestine10.1016/j.jmbbm.2011.05.030 ↗
5. Abdominal wall repair using human acellular dermis10.1016/j.amjsurg.2006.11.012 ↗
6. Biaxial mechanical properties of the natural and glutaraldehyde treated aortic valve cusp—Part I: Experimental results10.1115/1.429624 ↗
7. Abdominal hernia repair with bridging acellular dermal matrix—an expensive hernia sac10.1016/j.amjsurg.2007.06.035 ↗
8. Extracellular matrix biomaterials for soft tissue repair10.1016/j.cpm.2009.08.001 ↗
9. De Silva GS, Krpata DM, Gao Y, Criss CN, Anderson JM, Soltanian HT et al. Lack of identifiable biologic behavior in a series of porcine mesh explants. Surgery. doi:10.1016/j.surg.2014.03.01110.1016/j.surg.2014.03.011 ↗
10. Physicomechanical evaluation of polypropylene, polyester, and polytetrafluoroethylene meshes for inguinal hernia repair10.1016/j.jamcollsurg.2010.09.012 ↗
11. Histologic and biomechanical evaluation of crosslinked and non-crosslinked biologic meshes in a porcine model of ventral incisional hernia repair10.1016/j.jamcollsurg.2011.01.006 ↗
12. Differentiation of biologic scaffold materials through physicomechanical, thermal, and enzymatic degradation techniques10.1097/SLA.0b013e3182445341 ↗
13. Evaluation of various prosthetic materials and newer meshes for hernia repairs
14. In vivo evaluation of acellular human dermis for abdominal wall repair10.1002/jbm.a.32636 ↗
15. Effect of repetitive loading on the mechanical properties of synthetic hernia repair materials10.1016/j.jamcollsurg.2011.05.018 ↗
16. Analytically derived material properties of multilaminated extracellular matrix devices using the ball-burst test10.1016/j.biomaterials.2005.01.070 ↗
17. The use of prosthetics in abdominal wall reconstruction10.1016/j.cps.2005.12.002 ↗
18. Are biologic grafts effective for hernia repair?: a systematic review of the literature10.1177/1553350609331397 ↗
19. Biaxial tension of fibrous tissue: using finite element methods to address experimental challenges arising from boundary conditions and anisotropy10.1115/1.4023503 ↗
20. Use of acellular dermal matrix for complicated ventral hernia repair: does technique affect outcomes?10.1016/j.jamcollsurg.2007.06.012 ↗
21. Acellular dermal matrix in the management of high-risk abdominal wall defects10.1016/j.amjsurg.2006.09.003 ↗
22. Modified mesh for hernia repair that is adapted to the physiology of the abdominal wall10.1080/110241598750005138 ↗
23. Foreign body reaction to meshes used for the repair of abdominal wall hernias10.1080/11024159950189726 ↗
24. Retrieval study at 623 human mesh explants made of polypropylene—impact of mesh class and indication for mesh removal on tissue reaction10.1002/jbm.b.32958 ↗
25. Biaxial mechanical/structural effects of equibiaxial strain during crosslinking of bovine pericardial xenograft materials10.1016/S0142-9612(98)00142-2 ↗
26. Long-term complications associated with prosthetic repair of incisional hernias10.1001/archsurg.133.4.378 ↗
27. Mechanical failure of a lightweight polypropylene mesh10.1007/s10029-012-0959-5 ↗
28. A comparison of suture repair with mesh repair for incisional hernia10.1056/NEJM200008103430603 ↗
29. Mechanical characterisation of porcine rectus sheath under uniaxial and biaxial tension10.1016/j.jbiomech.2014.03.009 ↗
30. Revascularization of human acellular dermis in full-thickness abdominal wall reconstruction in the rabbit model10.1097/01.SAP.0000044252.76804.6B ↗
31. A non-biological model system to simulate the in vivo mechanical behavior of prosthetic meshes10.1016/j.jmbbm.2013.01.029 ↗
32. Uniaxial and biaxial mechanical characterization of a prosthetic mesh at different length scales10.1016/j.jmbbm.2013.07.021 ↗
33. Characterization of collagen fiber architecture in the canine diaphragmatic central tendon10.1115/1.2891370 ↗
34. Effect of pretension and suture needle type on mechanical properties of acellular human dermis patches for rotator cuff repair10.1016/j.jse.2011.10.028 ↗
35. Fiber-reinforced dermis graft for ventral hernia repair10.1016/j.jmbbm.2014.03.001 ↗
36. A review of available prosthetics for ventral hernia repair10.1097/SLA.0b013e3181f9b6e6 ↗
37. Biological meshes: a review of their use in abdominal wall hernia repairs10.1016/j.surge.2012.02.006 ↗
38. Effects of boundary conditions on the estimation of the planar biaxial mechanical properties of soft tissues10.1115/1.1933931 ↗
39. Multilaminate resorbable biomedical device under biaxial loading10.1002/(SICI)1097-4636(199823)43:3<277::AID-JBM8>3.0.CO;2-K ↗
40. Optical strain fields in shear and tensile testing of textile reinforcements10.1016/j.compscitech.2007.08.018 ↗
41. Central rupture and bulging of low-weight polypropylene mesh following recurrent incisional sublay hernioplasty10.1007/s10029-013-1197-1 ↗