Clinically relevant mechanical testing of hernia graft constructs

Sambit Sahoo, Katherine R. DeLozier, Ahmet Erdemir, Kathleen A. Derwin

Publisher page

Abstract

Background To understand the mechanical behavior of grafts in the context of hernia repair, there is a need to develop and adopt methods for mechanical testing of grafts in a clinically-relevant manner with clinically-relevant outcomes. Materials and methods Ball-burst and planar-biaxial methods were used to test three commercially-available hernia grafts (DermaMatrix, Biodesign, VitaMesh Blue). Both load-to-failure and cyclic fatigue tests were performed (n=6-11/group/test). Grafts were tested as sutured constructs in patch geometry. Dilatational strain analysis was performed considering the construct (both test methods) or the graft (planar-biaxial only) as the area of interest. Results DermaMatrix, Biodesign, and VitaMesh grafts showed differences in mechanical properties at the point of construct failure (load, in-plane load-per-suture and membrane tension) in ball-burst tests and differences in sub-failure properties (stiffness, dilatational strain at 16N/cm and cyclic mechanical properties) in planar-biaxial tests. In both load-to-failure and cyclic fatigue tests, each graft construct tended to be stiffer in planar-biaxial than ball-burst testing. In biaxial testing, the strain analysis method influenced the mechanical properties with the construct being more compliant than the graft. Conclusions This study demonstrates that graft-fixation method, test mode and analysis method are important considerations for mechanical characterization of hernia grafts. Ball-burst tests can only estimate construct or material properties, whereas planar-biaxial tests capture anisotropy and can estimate construct, graft and material properties of the same test specimen. When the clinical performance of a graft in the context of hernia repair is of interest, testing a sutured construct and performing construct strain analysis arguably provides the most clinically-relevant assessment method.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Hernia repair and management · Intestinal and Peritoneal Adhesions · Surgical Sutures and Adhesives

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (41)

Show source-provided references
1. ASTM Standard D3787-07. Standard Test Method for Bursting Strength of Textiles-Constant-Rate-of-Traverse (CRT) Ball Burst Test. 2011, doi:10.1520/D3787-07R11, www.astm.org.10.1520/D3787-07R11 ↗
2. ASTM Standard D6797-07. Standard Test Method for Bursting Strength of Fabrics Constant-Rate-of-Extension (CRE) Ball Burst Test. 2011, doi:10.1520/D6797-07R11, www.astm.org.10.1520/D6797-07R11 ↗
3. Mechanisms of ventral hernia recurrence after mesh repair and a new proposed classification10.1016/j.jamcollsurg.2005.02.035 ↗
4. Biaxial mechanical modeling of the small intestine10.1016/j.jmbbm.2011.05.030 ↗
5. Abdominal wall repair using human acellular dermis10.1016/j.amjsurg.2006.11.012 ↗
6. Biaxial mechanical properties of the natural and glutaraldehyde treated aortic valve cusp—Part I: Experimental results10.1115/1.429624 ↗
7. Abdominal hernia repair with bridging acellular dermal matrix—an expensive hernia sac10.1016/j.amjsurg.2007.06.035 ↗
8. Extracellular matrix biomaterials for soft tissue repair10.1016/j.cpm.2009.08.001 ↗
9. De Silva GS, Krpata DM, Gao Y, Criss CN, Anderson JM, Soltanian HT et al. Lack of identifiable biologic behavior in a series of porcine mesh explants. Surgery. doi:10.1016/j.surg.2014.03.01110.1016/j.surg.2014.03.011 ↗
10. Physicomechanical evaluation of polypropylene, polyester, and polytetrafluoroethylene meshes for inguinal hernia repair10.1016/j.jamcollsurg.2010.09.012 ↗
11. Histologic and biomechanical evaluation of crosslinked and non-crosslinked biologic meshes in a porcine model of ventral incisional hernia repair10.1016/j.jamcollsurg.2011.01.006 ↗
12. Differentiation of biologic scaffold materials through physicomechanical, thermal, and enzymatic degradation techniques10.1097/SLA.0b013e3182445341 ↗
13. Evaluation of various prosthetic materials and newer meshes for hernia repairs
14. In vivo evaluation of acellular human dermis for abdominal wall repair10.1002/jbm.a.32636 ↗
15. Effect of repetitive loading on the mechanical properties of synthetic hernia repair materials10.1016/j.jamcollsurg.2011.05.018 ↗
16. Analytically derived material properties of multilaminated extracellular matrix devices using the ball-burst test10.1016/j.biomaterials.2005.01.070 ↗
17. The use of prosthetics in abdominal wall reconstruction10.1016/j.cps.2005.12.002 ↗
18. Are biologic grafts effective for hernia repair?: a systematic review of the literature10.1177/1553350609331397 ↗
19. Biaxial tension of fibrous tissue: using finite element methods to address experimental challenges arising from boundary conditions and anisotropy10.1115/1.4023503 ↗
20. Use of acellular dermal matrix for complicated ventral hernia repair: does technique affect outcomes?10.1016/j.jamcollsurg.2007.06.012 ↗
21. Acellular dermal matrix in the management of high-risk abdominal wall defects10.1016/j.amjsurg.2006.09.003 ↗
22. Modified mesh for hernia repair that is adapted to the physiology of the abdominal wall10.1080/110241598750005138 ↗
23. Foreign body reaction to meshes used for the repair of abdominal wall hernias10.1080/11024159950189726 ↗
24. Retrieval study at 623 human mesh explants made of polypropylene—impact of mesh class and indication for mesh removal on tissue reaction10.1002/jbm.b.32958 ↗
25. Biaxial mechanical/structural effects of equibiaxial strain during crosslinking of bovine pericardial xenograft materials10.1016/S0142-9612(98)00142-2 ↗
26. Long-term complications associated with prosthetic repair of incisional hernias10.1001/archsurg.133.4.378 ↗
27. Mechanical failure of a lightweight polypropylene mesh10.1007/s10029-012-0959-5 ↗
28. A comparison of suture repair with mesh repair for incisional hernia10.1056/NEJM200008103430603 ↗
29. Mechanical characterisation of porcine rectus sheath under uniaxial and biaxial tension10.1016/j.jbiomech.2014.03.009 ↗
30. Revascularization of human acellular dermis in full-thickness abdominal wall reconstruction in the rabbit model10.1097/01.SAP.0000044252.76804.6B ↗
31. A non-biological model system to simulate the in vivo mechanical behavior of prosthetic meshes10.1016/j.jmbbm.2013.01.029 ↗
32. Uniaxial and biaxial mechanical characterization of a prosthetic mesh at different length scales10.1016/j.jmbbm.2013.07.021 ↗
33. Characterization of collagen fiber architecture in the canine diaphragmatic central tendon10.1115/1.2891370 ↗
34. Effect of pretension and suture needle type on mechanical properties of acellular human dermis patches for rotator cuff repair10.1016/j.jse.2011.10.028 ↗
35. Fiber-reinforced dermis graft for ventral hernia repair10.1016/j.jmbbm.2014.03.001 ↗
36. A review of available prosthetics for ventral hernia repair10.1097/SLA.0b013e3181f9b6e6 ↗
37. Biological meshes: a review of their use in abdominal wall hernia repairs10.1016/j.surge.2012.02.006 ↗
38. Effects of boundary conditions on the estimation of the planar biaxial mechanical properties of soft tissues10.1115/1.1933931 ↗
39. Multilaminate resorbable biomedical device under biaxial loading10.1002/(SICI)1097-4636(199823)43:3<277::AID-JBM8>3.0.CO;2-K ↗
40. Optical strain fields in shear and tensile testing of textile reinforcements10.1016/j.compscitech.2007.08.018 ↗
41. Central rupture and bulging of low-weight polypropylene mesh following recurrent incisional sublay hernioplasty10.1007/s10029-013-1197-1 ↗