Study of two different thin film coating methods in transmission laser micro-joining of thin Ti-film coated glass and polyimide for biomedical applications
Source abstract
Abstract
Biomedical devices and implants require precision joining for hermetic sealing which can be achieved with low power lasers. The effect of two different thin metal film coating methods was studied in transmission laser micro-joints of titanium-coated glass and polyimide. The coating methods were cathodic arc physical vapor deposition (CA-PVD) and electron beam evaporation (EB-PVD). Titanium-coated glass joined to polyimide film can have neural electrode application. The improvement of the joint quality will be essential for robust performance of the device. Low power fiber laser (wave length = 1100 nm) was used for transmission laser micro-joining of thin titanium (Ti) film (approximately 200 nm) coated Pyrex borosilicate 7740 glass wafer (0.5 mm thick) and polyimide (Imidex) film (0.2 mm thick). Ti film acts as the coupling agent in the joining process. The Ti film deposition rate in the CA-PVD was 5-10 A/s and in the EB-PVD 1.5 A/s. The laser joint strength was measured by a lap shear test, the Ti film surfaces were analyzed by atomic force microscopy (AFM) and the lap shear tested joints were analyzed by optical microscopy and scanning electron microscopy (SEM). The film properties and the failure modes of the joints were correlated to joint strength. The CA-PVD produced around 4 times stronger laser joints than EB-PVD. The adhesion of the Ti film on glass by CA-PVD is better than that of the EB-PVD method. This is likely to be due to a higher film deposition rate and consequently higher adhesion or sticking coefficient for the CA-PVD particles arriving on the substrate compared to that of the EB-PVD film. EB-PVD shows poor laser bonding properties due to the development of thermal hotspots which occurs from film decohesion.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Metal and Thin Film Mechanics · Nanofabrication and Lithography Techniques · Synthesis and properties of polymers
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (26)
Show source-provided references
1. Bauer, I., Russek, U.A., Herfurth, H., Witte, R., Heinemann, S., Newaz, G., Mian, A., Georgiev, D., Auner, G., 2004. Laser micro-joining of dissimilar and biocompatible materials. In: Proceedings of SPIE - Photonics 2004, vol Lasers and Applications in Science and Engineering, pp. 45410.1117/12.530239 ↗
2. Fibre laser welding for packaging of disposable polymeric microfluidic-biochips10.1016/j.apsusc.2007.08.013 ↗
3. Surface structure and composition of flat titanium thin films as a function of film thickness and evaporation rate10.1016/j.apsusc.2005.01.013 ↗
4. A multielectrode implant device for the cerebral cortex10.1016/S0165-0270(99)00087-4 ↗
5. Stresses and deformation processes in thin films on substrates10.1080/10408438808243734 ↗
6. An xps study of laser-fabricated polyimide/titanium interfaces10.1016/j.apsusc.2004.03.261 ↗
7. Laser fabrication and characterization of sub-millimeter joints between polyimide and ti-coated borosilicate glass10.1007/s10853-005-1291-6 ↗
8. Gilleo, Ken, 2001. Overview of new packages, materials and processes. In: International Symposium on Advanced Packaging Materials, IMAPS, pp. 1–510.1109/ISAOM.2001.916539 ↗
9. Laser processing of components for polymer microfluidic and optoelectronic products
10. Thermal transport through thin films: Mirage technique measurements on aluminum/titanium multilayers10.1557/JMR.2000.0110 ↗
11. Grigorian, Ani, 2007. Metalworking: Making the cuts of tomorrow, today. Medical device linkcom
12. Polyimide-based intracortical neural implant with improved structural stiffness10.1088/0960-1317/14/1/305 ↗
13. Investigation of electronic packaging materials by using a 6-axis mini thermomechanical tester10.1016/S0020-7683(97)00281-3 ↗
14. Polyimide-based microfluidic devices10.1039/b103896f ↗
15. Performance of laser bonded glass/polyimide microjoints in cerebrospinal fluid
16. Laser bonded microjoints between titanium and polyimide for applications in medical implants
17. Thin film synthesis by energetic condensation10.1146/annurev.matsci.31.1.111 ↗
18. Laser fabrication and characterization of adhesive-free joints for encapsulation of biomedical implant devices
19. A comparison between glass/polyimide and titanium/polyimide microjoint performances in cerebrospinal fluid10.1002/jbm.a.30709 ↗
20. Methods to reduce the thermal conductivity of eb-pvd tbcs
21. Surface energy, surface topography and adhesion10.1016/S0143-7496(03)00068-X ↗
22. Single layer and multilayer wear resistant coatings of (ti,al) n: A review10.1016/S0921-5093(02)00259-9 ↗
23. Theoretical study of metal-polyimide interfacial properties10.1016/S0042-207X(01)00332-3 ↗
24. Application of usaxs analysis and non-interacting approximation to determine the influence of process parameters and ageing on the thermal conductivity of electron-beam physical vapor deposited thermal barrier coatings10.1016/j.surfcoat.2006.10.020 ↗
25. Nano-structural characteristics of ti/glass and ti/mo films as a function of deposition rate and angle of incidence10.1016/j.tsf.2005.12.283 ↗
26. Wise, K.D., Anderson, D.J., Hetke, J.F., Kipke, D.R., Najafi, K., 2004. Wireless implantable microsystems: High-density electronic interfaces to the nervous system. In: Proceedings of the IEEE, pp. 76–9710.1109/JPROC.2003.820544 ↗