journal article·2017-05

In situ observation of lubricant film formation in THR considering real conformity: The effect of diameter, clearance and material

D. Nečas, M. Vrbka, F. Urban, J. Gallo, I. Křupka, M. Hartl

Publisher page

Abstract

The aim of the present study is to provide an analysis of protein film formation in hip joint replacements considering real conformity based on in situ observation of the contact zone. The main attention is focused on the effect of implant nominal diameter, diametric clearance and material. For this purpose, a pendulum hip joint simulator equipped with electromagnetic motors enabling to apply continuous swinging flexion-extension motion was employed. The experimental configuration consists of femoral component (CoCrMo, BIOLOX®forte, BIOLOX®delta) and acetabular cup from optical glass fabricated according to the dimensions of real cups. Two nominal diameters were studied, 28 and 36mm, respectively, while different diametric clearances were considered. Initially, a static test focused on the protein adsorption onto rubbing surfaces was performed with 36mm implants. It was found that the development of adsorbed layer is much more stable in the case of metal head, indicating that the adsorption forces are stronger compared to ceramic. A consequential swinging test revealed that the fundamental parameter influencing the protein film formation is diametric clearance. Independently of implant diameter, film was much thicker when a smaller clearance was considered. An increase of implant size from 28mm to 36mm did not cause a substantial difference in film formation; however, the total film thickness was higher for smaller implant. In terms of material, metal heads formed a thicker film, while this fact can be, among others, also attributed to clearance, which is more than two times higher in the case of ceramic implant.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Dental materials and restorations · Advanced materials and composites

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (32)

Show source-provided references
1. A comparison of friction in 28mm conventional and 55mm resurfacing metal-on-metal hip replacements10.1243/13506501JET234 ↗
2. The influence of clearance on friction, lubrication and squeaking in large diameter metal-on-metal hip replacements
3. Ideal femoral head size in total hip arthroplasty balances stability and volumetric wear10.1007/s11420-012-9287-7 ↗
4. A hip joint simulator study of the performance of metal-on-metal joints: part II: design
5. Biotribology of artificial hip joints10.5312/wjo.v6.i1.77 ↗
6. Inlet protein aggregation: a new mechanism for lubricating film formation with model synovial fluids10.1177/0954411911401306 ↗
7. (i) Head size, does it matter?10.1016/j.cuor.2008.05.003 ↗
8. Femoral head diameter considerations for primary total hip arthroplasty
9. Thin film colorimetric interferometry10.1080/10402000108982458 ↗
10. The elastohydrodynamic lubrication of heavily loaded point contacts10.1243/JMES_JOUR_1980_022_036_02 ↗
11. Prediction of lubricating film thickness in UHMWPE hip joint replacements10.1016/S0021-9290(00)00181-0 ↗
12. Long-term results of Charnley low-friction arthroplasty in young patients10.1302/0301-620X.75B4.8331119 ↗
13. Importance of head diameter, clearance, and cup wall thickness in elastohydrodynamic lubrication analysis of metal-on-metal hip resurfacing prostheses10.1243/09544119JEIM172 ↗
14. Formation of adsorbed protein layers10.1006/jcis.1998.5763 ↗
15. Friction and lubricant film thickness measurements on simulated synovial fluids10.1243/13506501JET580 ↗
16. Lubricating film thickness measurements with bovine serum10.1016/j.triboint.2010.07.008 ↗
17. In contact observation of model synovial fluid lubricating mechanisms10.1016/j.triboint.2012.04.029 ↗
18. On the matter of synovial fluid lubrication: implications for metal-on-metal hip tribology10.1016/j.jmbbm.2013.12.016 ↗
19. The effect of transient conditions on synovial fluid protein aggregation lubrication10.1016/j.jmbbm.2014.02.005 ↗
20. Lubrication of metal-on-metal hip joints: the effect of protein content and load on film formation and wear10.1016/j.jmbbm.2011.09.008 ↗
21. The effect of lubricant constituents on lubrication mechanisms in hip joint replacements10.1016/j.jmbbm.2015.11.006 ↗
22. Lubrication within hip replacements–implication for ceramic-on-hard bearing couples10.1016/j.jmbbm.2016.04.003 ↗
23. The effect of buffer solution choice on protein adsorption and lubrication10.1016/j.triboint.2013.12.005 ↗
24. Wettability analysis of orthopaedic materials using optical contact angle methods10.4028/www.scientific.net/KEM.309-311.1199 ↗
25. Influence of protein and lipid concentration of the test lubricant on the wear of ultra high molecular weight polyethylene10.1016/j.triboint.2007.11.010 ↗
26. The effect of diametral clearance, motion and loading cycles upon lubrication of metal-on-metal total hip replacements10.1243/0954406011520454 ↗
27. The effect of femoral head diameter upon lubrication and wear of metal-on-metal total hip replacements10.1243/0954411011533724 ↗
28. Study of film formation in bovine serum lubricated contacts under rolling/sliding conditions10.1177/1350650112471000 ↗
29. In situ measurements of thin films in bovine serum lubricated contacts using optical interferometry10.1177/0954411913517498 ↗
30. Determination of a friction coefficient for THA bearing couples10.55095/achot2015/057 ↗
31. Visualization of lubricating films between artificial head and cup with respect to real geometry10.1016/j.biotri.2015.05.002 ↗
32. Wusylko, A., Freed, R., Brandt, J. M., et al., 2014. Comparison Between Surface Roughness and Wettability on Retrieved Metal and Ceramic Femoral Heads. In: Society for Biomaterials Annual Meeting and Exposition 2014: Transactions Proceedings of the 38th Annual Meeting, Society For Biomaterials, USA, s. 277. ISBN 978-1-63266-236-1. ISSN 1526-7547.