Quantification of in vitro wear of a synthetic meniscus implant using gravimetric and micro-CT measurements
Source abstract
Abstract
A synthetic meniscus implant was recently developed for the treatment of patients with mild to moderate osteoarthritis with knee pain associated with medial joint overload. The implant is distinctively different from most orthopedic implants in its pliable construction, and non-anchored design, which enables implantation through a mini-arthrotomy without disruption to the bone, cartilage, and ligaments. Due to these features, it is important to show that the material and design can withstand knee joint conditions. This study evaluated the long-term performance of this device by simulating loading for a total of 5 million gait cycles (Mc), corresponding to approximately five years of service in-vivo. All five implants remained in good condition and did not dislodge from the joint space during the simulation. Mild abrasion was detected by electron microscopy, but µ-CT scans of the implants confirmed that the damage was confined to the superficial surfaces. The average gravimetric wear rate was 14.5 mg/Mc, whereas volumetric changes in reconstructed µ-CT scans point to an average wear rate of 15.76 mm(3)/Mc (18.8 mg/Mc). Particles isolated from the lubricant had average diameter of 15 µm. The wear performance of this polycarbonate-urethane meniscus implant concept under ISO-14243 loading conditions is encouraging.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Knee injuries and reconstruction techniques · Total Knee Arthroplasty Outcomes · Orthopaedic implants and arthroplasty
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (26)
Show source-provided references
1. Unicompartmental knee prostheses: in vitro wear assessment of the menisci tibial insert after two different fixation methods10.1088/0031-9155/53/19/006 ↗
2. Cushion form bearings for total knee joint replacement. Part 2: wear and durability10.1243/PIME_PROC_1995_209_324_02 ↗
3. The Unispacer knee implant: early clinical results10.1302/0301-620X.90B4.20319 ↗
4. An analysis of rotating-platform total knee replacements10.1097/01.blo.0000063562.90853.6a ↗
5. Condello, V., Arbel, R., Agar, G., Rozen, N., Angele, P., Victor, J., Brittberg, M., Verdonk, P., 2014. A novel synthetic meniscus implant for the treatment of middle aged patients: results of 118 patients in a prospective, multi-center study. 16th ESSKA Congress Amsterdam NL.
6. In-vivo evaluation of the kinematic behavior of an artificial medial meniscus implant: a pilot study using open-MRI10.1016/j.clinbiomech.2014.07.001 ↗
7. Design considerations for cushion form bearings in artificial hip joints10.1243/PIME_PROC_1991_205_269_02 ↗
8. MRI-based characterization of bone anatomy in the human knee for size matching of a medial meniscal implant10.1115/1.4002490 ↗
9. Long-term evaluation of a compliant cushion form acetabular bearing for hip joint replacement: a 20 million cycles wear simulation10.1002/jor.21471 ↗
10. Wear debris classification
11. Biotribology of alternative bearing materials for unicompartmental knee arthroplasty10.1016/j.actbio.2010.04.003 ↗
12. Measurements of constraint of total knee replacement10.1016/j.jbiomech.2004.02.014 ↗
13. Prospective single-arm, multi-center trial of a patient-specific interpositional knee implant: early clinical results10.2174/1874325001105010037 ↗
14. Bio-ferrography to capture and separate polyethylene wear debris from hip simulator fluid and compared with conventional filter method10.1115/1.2162554 ↗
15. Oatis · 2004 · 1992
16. The effect of loading and material on the biomechanical properties and vitality of bovine cartilage in vitro
17. Measurement of wear in orthopedic prosthesis10.12693/APhysPolA.121.131 ↗
18. Tribological assessment of a flexible carbon-fibre reinforced poly(ether–ether–ketone) acetabular cup articulating against an alumina femoral head10.1243/09544119JEIM334 ↗
19. Polyurethane unicondylar knee prostheses: simulator wear tests and lubrication studies10.1088/0031-9155/52/1/013 ↗
20. Wear studies on the likely performance of CFR-PEEK/CoCrMo for use as artificial joint bearing materials
21. Contributions of muscles, ligaments, and the ground-reaction force to tibiofemoral joint loading during normal gait10.1002/jor.20255 ↗
22. Viscoelastic properties of a synthetic meniscus implant10.1016/j.jmbbm.2013.08.021 ↗
23. CMM-based procedure for polyethylene non-congruous unicompartmental knee prosthesis wear assessment10.1016/j.wear.2008.12.049 ↗
24. Evaluation of the wear performance of a polycarbonate-urethane acetabular component in a hip joint simulator and comparison with UHMWPE and cross-linked UHMWPE10.1177/0885328210394471 ↗
25. Muscle and external load contribution to knee joint contact loads during normal gait10.1016/j.jbiomech.2009.06.019 ↗
26. Chondroprotective effects of a polycarbonate-urethane meniscal implant: histopathological results in a sheep model10.1007/s00167-010-1210-5 ↗