Contact stress analysis of the anterior tibial post in bi-cruciate stabilized and mobile-bearing posterior stabilized total knee arthroplasty designs
Source abstract
Abstract
In posterior-stabilized (PS) total knee arthroplasty (TKA), unexpected wear and fracture of the tibial post due to anterior post impingement have been reported. The purpose of this study was to determine the contact stress on the anterior aspect of the tibial post in four contemporary TKA designs. We evaluated one bi-cruciate stabilized design (Journey II) and three mobile-bearing PS designs (Vanguard RP, PFC Sigma RP, and NexGen LPS Mobile). The contact conditions at the anterior aspect of the tibial post were determined upon application of a posterior force of 100N to individual implants. Each measurement was sequentially performed five times, and the data were compared within and across designs using analysis of variance and a post-hoc test. The contact stress of the Journey II and Vanguard RP was less than the compressive yield stress for polyethylene (10MPa) at all tested flexion angles and degrees of rotation. The PFC Sigma RP did not show anterior tibial post impingement under any experimental conditions. The NexGen LPS Mobile demonstrated bilateral edge loading at the anterior tibial post and exceeded 10MPa of contact stress in some test conditions. Thus, the differences among implants in terms of the dimensions of the femoral anterior cam or intercondylar notch and the anterior aspect of the tibial post in the axial and sagittal planes led to significant differences in contact conditions. The present study helps the surgeon to be more aware that various contact conditions of the anterior aspect of the tibial post can occur in individual TKA designs.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Total Knee Arthroplasty Outcomes · Orthopaedic implants and arthroplasty · Knee injuries and reconstruction techniques
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (48)
Show source-provided references
1. Contact stress analysis of the conforming post-cam mechanism in posterior-stabilized total knee arthroplasty10.1016/j.arth.2007.05.023 ↗
2. Fractography and oxidative analysis of gamma inert sterilized posterior-stabilized tibial insert post fractures: report of two cases10.1016/j.knee.2013.04.004 ↗
3. Mechanism of anterior impingement damage in total knee arthroplasty10.2106/00004623-200200002-00004 ↗
4. Patellofemoral forces after total knee arthroplasty: effect of extensor moment arm10.1016/j.knee.2004.05.006 ↗
5. Evaluation of contact stress in metal-backed patellar replacements. A predictor of survivorship
6. Mobile-bearing knee replacement: concepts and results
7. Tibial post impingement in posterior-stabilized total knee arthroplasty10.1097/00003086-200211000-00014 ↗
8. Cemented rotating-platform total knee replacement: a concise follow-up, at a minimum of twenty years, of a previous report10.2106/JBJS.I.01012 ↗
9. Fracture of the polyethylene tibial post in a NexGen posterior-stabilized knee prosthesis10.1016/j.arth.2004.04.013 ↗
10. Consecutive series of 226 journey bicruciate substituting total knee replacements: early complication and revision rates10.1186/1471-2474-15-395 ↗
11. Patellofemoral crepitus after total knee arthroplasty: etiology and preventive measures10.4055/cios.2014.6.1.9 ↗
12. Implant design influences tibial post wear damage in posterior-stabilized knees10.1007/s11999-010-1515-1 ↗
13. The incidence of the patellar clunk syndrome in a recently designed mobile-bearing posteriorly stabilised total knee replacement10.1302/0301-620X.91B4.21494 ↗
14. Tibial post wear in posterior-stabilized knee replacements is design-dependent10.1007/s11999-008-0422-1 ↗
15. In vivo kinematics of mobile-bearing total knee arthroplasty during deep knee bending under weight-bearing conditions10.1007/s00167-010-1262-6 ↗
16. Tibial post impingement in posterior-stabilized total knee arthroplasty
17. The “screw-home” movement in the knee-joint10.3109/17453676608989407 ↗
18. Evaluation of impingement of the anterior tibial post during gait in a posteriorly-stabilised total knee replacement10.1302/0301-620X.90B9.20298 ↗
19. Contact stress at the anterior aspect of the tibial post in posterior-stabilized total knee replacement10.2106/JBJS.I.00479 ↗
20. Continuous sagittal radiological evaluation of stair-climbing in cruciate-retaining and posterior-stabilized total knee arthroplasties using image-matching techniques10.1016/j.arth.2014.12.027 ↗
21. In vivo anterior tibial post contact after posterior stabilizing total knee arthroplasty10.1002/jor.20417 ↗
22. Stress analysis of the anterior tibial post in posterior stabilized knee prostheses10.1002/jor.20336 ↗
23. Post damage in contemporary posterior-stabilized tibial inserts: influence of implant design and clinical relevance10.1016/j.arth.2010.05.011 ↗
24. Long-term results of posterior cruciate-substituting total knee arthroplasty10.1097/00003086-200211000-00009 ↗
25. Fracture of the polyethylene tibial post in a posterior stabilized knee prosthesis: a case report and review of literature10.1016/j.jor.2015.01.002 ↗
26. In vivo knee kinematics during stair and deep flexion activities in patients with bicruciate substituting total knee arthroplasty10.1016/j.arth.2011.03.005 ↗
27. In vivo degradation of polyethylene liners after gamma sterilization in air10.2106/JBJS.D.02111 ↗
28. Early nontraumatic fracture of the polyethylene tibial post in a NexGen LPS-Flex posterior stabilized knee prosthesis
29. Anterior tibial post impingement in a posterior stabilized total knee arthroplasty10.1016/j.orthres.2004.09.005 ↗
30. Fracture of a polyethylene tibial post in a Scorpio posterior-stabilized knee prosthesis10.4055/cios.2009.1.2.118 ↗
31. A comparison of pressure sensitive film and digital electronic sensors to measure contact area and contact stress
32. Contact stress analysis in meniscal bearing total knee arthroplasty10.1016/S0883-5403(98)80016-8 ↗
33. Loosening and osteolysis with the press-fit condylar posterior-cruciate-substituting total knee replacement10.2106/00004623-200103000-00012 ↗
34. Prediction of total knee arthroplasty polyethylene wear using the wear index10.1054/arth.2002.33546 ↗
35. Cruciate coupling and screw-home mechanism in passive knee joint during extension-flexion10.1016/j.jbiomech.2004.05.033 ↗
36. Contact stress at the post-cam mechanism in posterior-stabilised total knee arthroplasty10.1302/0301-620X.87B4.15684 ↗
37. In vivo kinematic comparison between mobile-bearing and fixed-bearing total knee arthroplasty during step-up activity10.1016/j.arth.2014.02.022 ↗
38. Osteolysis associated with a cemented modular posterior-cruciate-substituting total knee design: five to eight-year follow-up10.2106/00004623-200208000-00012 ↗
39. Tibial post wear in posterior stabilized total knee arthroplasty. An unrecognized source of polyethylene debris10.2106/00004623-200103000-00011 ↗
40. Surface damage on open box posterior-stabilized polyethylene tibial inserts10.1097/01.blo.0000092998.90435.98 ↗
41. Posterior stabilized prosthesis. Results after follow-up of nine to twelve years10.2106/00004623-199274070-00004 ↗
42. Patellofemoral contact pressures exceed the compressive yield strength of UHMWPE in total knee arthroplasties10.1016/S0883-5403(05)80186-X ↗
43. Survivorship at minimum 10-year follow-up of a rotating-platform, mobile-bearing, posterior-stabilised total knee arthroplasty10.1007/s00167-014-3118-y ↗
44. Bicruciate substituting total knee replacement: how effective are the added kinematic constraints in vivo?10.1007/s00167-011-1796-2 ↗
45. In vivo kinematics after a cruciate-substituting TKA10.1007/s11999-009-1072-7 ↗
46. Bicruciate-stabilised total knee replacements produce more normal sagittal plane kinematics than posterior-stabilised designs10.1302/0301-620X.93B7.26208 ↗
47. Anterior dislocation of Insall Burstein II total knee arthroplasty secondary to polyethylene tibial post fracture: a case report10.1177/230949901001800328 ↗
48. In vivo determination of cam-post engagement in fixed and mobile-bearing TKA10.1007/s11999-013-3257-3 ↗