Size and thickness effect on creep behavior in conventional and vitamin E-diffused highly crosslinked polyethylene for total hip arthroplasty

Yasuhito Takahashi, Toshiyuki Tateiwa, Takaaki Shishido, Toshinori Masaoka, Kosuke Kubo, Kengo Yamamoto

Publisher page

Abstract

Since the early 2000s, the use of large femoral heads is becoming increasingly popular in total hip arthroplasty (THA), which provides an improved range of motion and joint stability. Large femoral heads commonly necessitate to be coupled with thinner acetabular liners than the conventionally used because of the limited sizes of outer shells (especially for patients with small pelvic size). However, the influence of the liner thinning on the mechanical performance is still not clearly understood. The objective of this study was to experimentally clarify the size and thickness effect on the rates of compressive creep strain in conventional (virgin low-crosslinked) and vitamin E-diffused highly crosslinked, ultra-high molecular weight polyethylene (UHMWPE) acetabular liners. We applied uniaxial compression to these liners of various internal diameters (28, 32 and 36mm) and thicknesses (4.8, 6.8 and 8.9mm) up to 4320min under the constant load of 3000N. Vitamin E-diffused highly crosslinked UHMWPE components showed significantly greater creep resistance than the conventional ones. In the both types of UHMWPE, the rates of creep strain significantly decreased by increasing the internal diameter and thickness. Varying the component thickness contributed more largely to the creep behavior rather than the internal diameter. Our results suggest the positive mechanical advantage of using large femoral heads, but at the same time, a considerable liner thinning is not recommended for minimizing creep strain. Therefore, the further in-vitro as well as in-vivo research are necessary to conclude the optimal balance of head diameter and liner thickness within the limited sizes of outer shells.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Orthopedic Infections and Treatments

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (37)

Show source-provided references
1. Differences between the wear couples metal-on-polyethylene and ceramic-on-ceramic in the stability against dislocation of total hip replacement10.1023/B:JMSM.0000030214.79180.13 ↗
2. The effect of conformity, thickness, and material on stresses in ultra-high molecular weight components for total joint replacement10.2106/00004623-198668070-00010 ↗
3. Finite element simulation of early creep and wear in total hip arthroplasty10.1016/j.jbiomech.2004.10.022 ↗
4. Deformation mechanisms in crystalline polymers10.1007/BF00540553 ↗
5. Time resolved simultaneous small- and wide-angle X-ray scattering during polyethylene deformation 3. Compression of polyethylene10.1016/S0032-3861(97)00346-7 ↗
6. Serial measurement of polyethylene wear of well-fixed cementless metal-backed acetabular component in total hip arthroplasty: an over 10 year follow-up study10.1046/j.1525-1594.2000.06571-2.x ↗
7. The influence of head size on corrosion and fretting behavior at the head-neck interface of artificial hip joints10.1016/j.arth.2012.10.017 ↗
8. Plasticity-induced damage layer is a precursor to wear in radiation-cross-linked UHMWPE acetabular components for total hip replacement10.1016/S0883-5403(99)90086-4 ↗
9. The measurement of creep in ultrahigh molecular weight polyethylene. A comparision of conventional versus highly cross-linked polyethylene10.1016/j.arth.2004.08.008 ↗
10. Clinical fracture of cross-linked UHMWPE acetabular liners10.1016/j.biomaterials.2009.07.013 ↗
11. The creep and wear of highly cross-linked polyethylene: a three-year randomized, controlled trial using radiostereometric analysis10.1302/0301-620X.90B5.20545 ↗
12. J-integral fracture toughness and tearing modulus measurement of radiation cross-linked UHMWPE10.1016/S0736-0266(02)00073-6 ↗
13. Characterization of ultrahigh molecular weight polyethylene irradiated with γ-rays and electron beams to high doses10.1002/(SICI)1099-0518(19990115)37:2<159::AID-POLA6>3.0.CO;2-G ↗
14. Lower rates of dislocation with increased femoral head size after primary total hip replacement. A five-year analysis of NHS patients in England10.1302/0301-620X.93B7.26657 ↗
15. A parametric analysis of the contact stress in ultra-high molecular weight polyethylene acetabular cups10.1016/1350-4533(90)90006-T ↗
16. Enhanced computational prediction of polyethylene wear in hip joints by incorporating cross-shear and contact pressure in additional to load and sliding distance: effect of head diameter10.1016/j.jbiomech.2009.01.005 ↗
17. History and systematic review of wear and osteolysis outcomes for first-generation highly crosslinked polyethylene10.1007/s11999-011-1872-4 ↗
18. Advances in the processing, sterilization, and crosslinking of ultra-high molecular weight polyethylene for total joint arthroplasty10.1016/S0142-9612(99)00053-8 ↗
19. Compressive creep characteristics of extruded ultrahigh-molecular weight polyethylene10.1002/(SICI)1097-4636(199802)39:2<261::AID-JBM13>3.0.CO;2-G ↗
20. Structure and plastic deformation of polyethylene10.1007/BF01162485 ↗
21. Frictional torque in surface and conventional hip replacement10.2106/00004623-198365030-00012 ↗
22. Deformation of metal-backed acetabular components and the impact of liner thickness in a cadaveric model10.1007/s00264-010-1077-6 ↗
23. Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements10.1002/jor.1100170203 ↗
24. Hospital resources and implant cost management: a 2010 update
25. Vitamin E diffused, highly crosslinked UHMWPE: a review10.1007/s00264-010-1161-y ↗
26. Effect of crosslinking on the microstructure and mechanical properties of ultra-high molecular weight polyethylene10.1097/01.blo.0000185310.59202.e5 ↗
27. Wear versus thickness and other features of 5-Mrad crosslinked UHMWPE acetabular liners10.1007/s11999-010-1555-6 ↗
28. Large deformation compression induced crystallinity degradation of conventional and highly crosslinked UHMWPEs10.1016/j.biomaterials.2005.04.008 ↗
29. Strain-induced microstructural rearrangement in ultra-high molecular weight polyethylene for hip joints: a comparison between conventional and vitamin E-infused highly-crosslinked liners10.1016/j.jmbbm.2012.12.009 ↗
30. Raman spectroscopic study of remelting and annealing-induced effects on microstructure and compressive deformation behavior of highly crosslinked UHMWPE for total hip arthroplasty10.1002/jbm.b.33164 ↗
31. Mechanisms of plastic deformation in highly cross-linked UHMWPE for total hip components – The molecular physics viewpoint10.1016/j.jmbbm.2014.11.002 ↗
32. Effects of vitamin-E blending on plastic deformation mechanisms of highly cross-linked ultra-high molecular weight polyethylene (HXL-UHMWPE) in total hip arthroplasty10.1016/j.actbio.2014.12.022 ↗
33. Wang, A., Dumbleton, J.H., Essner, A., Yau, S.S., 2005. Sequentially cross-linked polyethylene. US Patent 20050043431.
34. Lubrication and wear of ultra-high molecular weight polyethylene in total joint replacements10.1016/S0301-679X(98)00005-X ↗
35. Orientation softening in the deformation and wear of ultra-high molecular weight polyethylene10.1016/S0043-1648(96)07362-0 ↗
36. Examination of the suitability of alpha-tocopherol as a stabilizer for ultra-high molecular weight polyethylene used for articulating surfaces in joint endoprostheses10.1023/A:1013834113967 ↗
37. Prospective clinical and joint simulator studies of a new total hip arthroplasty using alumina ceramic heads and cross-linked polyethylene cups10.1302/0301-620X.78B2.0780280 ↗