Wear performance of neat and vitamin E blended highly cross-linked PE under severe conditions: The combined effect of accelerated ageing and third body particles during wear test
Source abstract
Abstract
The objective of this study is to evaluate the effects of third-body particles on the in vitro wear behaviour of three different sets of polyethylene acetabular cups after prolonged testing in a hip simulator and accelerated ageing. Vitamin E-blended, cross-linked polyethylene (XLPE_VE), cross-linked polyethylene (XLPE) and conventional polyethylene (STD_PE) acetabular cups were simulator tested for two million cycles under severe conditions (i.e. by adding third-body particles to the bovine calf serum lubricant). Micro-Fourier Transform Infrared and micro-Raman spectroscopic analyses, differential scanning calorimetry, and crosslink density measurements were used to characterize the samples at a molecular level. The STD_PE cups had twice mass loss than the XLPE_VE components and four times than the XLPE samples; statistically significant differences were found between the mass losses of the three sets of cups. The observed wear trend was justified on the basis of the differences in cross-link density among the samples (XLPE>XLPE_VE>STD_PE). FTIR crystallinity profiles, bulk DSC crystallinity and surface micro-Raman crystallinity seemed to have a similar behaviour upon testing: all of them (as well as the all-trans and ortho-trans contents) revealed the most significant changes in XLPE and XLPE_VE samples. The more severe third-body wear testing conditions determined more noticeable changes in all spectroscopic markers with respect to previous tests. Unexpectedly, traces of bulk oxidation were found in both STD_PE (unirradiated) and XLPE (remelting-stabilized), which were expected to be stable to oxidation; on the contrary, XLPE_VE demonstrated a high oxidative stability in the present, highly demanding conditions.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Orthopedic Infections and Treatments
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (55)
Show source-provided references
1. In vitro effects on mobile polyethylene insert under highly demanding daily activities: stair climbing
2. The performance of gamma- and EtO-sterilised UHMWPE acetabulars cups tested under severe simulator conditions. Part 1: role of the third-body wear process10.1016/S0142-9612(02)00238-7 ↗
3. The predictive power of surface profile parameters on the amount of wear measured in vitro on metal-on-polyethylene artificial hip joints10.1243/09544119JEIM95 ↗
4. In vitro wear performance of standard, crosslinked, and vitamin-E-blended UHMWPE10.1002/jbm.a.33297 ↗
5. Does cyclic stress and accelerated ageing in fluence the wear behavior of highly crosslinked polyethylene ?10.1016/j.jmbbm.2016.02.029 ↗
6. ASTM F2102-3, 2013. ASTM F2102 − 13 – Standard Guide for Evaluating the Extent of Oxidation in Polyethylene Fabricated Forms Intended for Surgical Implants. 〈https://www.astm.org/Standards/F2102.htm〉.
7. The relative effects of radiation crosslinking and type of counterface on the wear resistance of ultrahigh-molecular-weight polyethylene10.1016/j.actbio.2011.05.018 ↗
8. Analysis of wear, wear particles, and reduced inflammatory potential of vitamin E ultrahigh-molecular-weight polyethylene for use in total joint replacement10.1002/jbm.b.32904 ↗
9. Vitamin E-stabilized UHMWPE for total joint implants: a review10.1007/s11999-010-1717-6 ↗
10. Oxidation behaviour in prosthetic UHMWPE components sterilised with high energy radiation in a low-oxygen environment10.1016/j.polymdegradstab.2006.02.003 ↗
11. Oxidation behaviour in prosthetic UHMWPE components sterilised with high-energy radiation in the presence of oxygen10.1016/j.polymdegradstab.2006.08.002 ↗
12. UHMWPE for arthroplasty: past or future?10.1007/s10195-008-0038-y ↗
13. Third-body wear of highly cross-linked polyethylene in a hip simulator10.1016/S0883-5403(03)00146-3 ↗
14. Effects of microtomy on the material properties of ultra high molecular weight polyethylene10.1016/S0142-9418(00)00088-X ↗
15. Comprehensive study of parameters for characterizing three-dimensional surface topography II: statistical properties of parameter variation10.1016/0043-1648(93)90050-V ↗
16. The basis for a second-generation highly cross-linked UHMWPE10.1097/01.blo.0000238856.61862.7d ↗
17. Zirconia phase transformation, metal transfer, and surface roughness in retrieved ceramic composite femoral heads in total hip arthroplasty10.1016/j.arth.2014.08.011 ↗
18. An activated energy approach for accelerated testing of the deformation of UHMWPE in artificial joints10.1016/j.jmbbm.2010.01.004 ↗
19. The effects of gamma radiation sterilization and ageing on the structure and morphology of medical grade ultra high molecular weight polyethylene10.1016/0032-3861(96)89386-4 ↗
20. Biotribology of a vitamin E-stabilized polyethylene for hip arthroplasty – influence of artificial ageing and third-body particles on wear10.1016/j.actbio.2014.02.052 ↗
21. Biological reactions to wear debris in total joint replacement10.1243/0954411001535219 ↗
22. James, S.P., Oldinski, R. (Kurkowski), Zhang, M., Schwartz, H., 2009. UHMWPE Biomaterials Handbook. 〈http://dx.doi.org/10.1016/B978-0-12-374721-1.00018-3〉.10.1016/B978-0-12-374721-1.00018-3 ↗
23. Friction and wear behavior of ultra-high molecular weight polyethylene as a function of polymer crystallinity10.1016/j.actbio.2008.02.022 ↗
24. The effect of gamma-irradiation on ultra-high molecular weight polyethylene recrystallized under different cooling conditions10.1016/S0969-806X(00)00333-9 ↗
25. Aggressive 3rd-body wear challenge to highly crosslinked polyethylene: a hip simulator model10.1016/j.wear.2009.01.029 ↗
26. Bearing surfaces in hip replacement – evolution and likely future10.1016/j.mjafi.2014.04.015 ↗
27. Kurtz · 2015
28. Trace concentrations of vitamin E protect radiation crosslinked UHMWPE from oxidative degradation10.1002/jbm.a.32122 ↗
29. Wear in highly crosslinked polyethylenes10.1016/j.cuor.2008.10.011 ↗
30. Interlaboratory reproducibility of standard accelerated aging methods for oxidation of UHMWPE10.1016/S0142-9612(00)00333-1 ↗
31. Evolution of morphology in UHMWPE following accelerated aging: the effect of heating rates10.1002/(SICI)1097-4636(199907)46:1<112::AID-JBM13>3.0.CO;2-U ↗
32. Accelerated aqueous aging simulation of in vivo oxidation for gamma-sterilized UHMWPE10.1002/jbm.b.31294 ↗
33. Morphological characterisation of the crystalline structure of cold-drawn HDPE used as a model material for the environmental stress cracking (ESC) phenomenon10.1016/S0032-3861(98)00500-X ↗
34. Use of vitamin E to protect cross-linked UHMWPE from oxidation10.1016/j.biomaterials.2010.01.076 ↗
35. Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements10.1002/jor.1100170203 ↗
36. Muratoglu, O.K., 2009. Highly crosslinked and melted UHMWPE. In: UHMWPE Biomaterials Handbook, Elsevier, pp. 197–204. doi: 〈http://dx.doi.org/10.1016/B978-0-12-374721-1.00013-4〉.10.1016/B978-0-12-374721-1.00013-4 ↗
37. A novel method of cross-linking ultra-high-molecular-weight polyethylene to improve wear, reduce oxidation, and retain mechanical properties. Recipient of the 1999 HAP Paul Award10.1054/arth.2001.20540 ↗
38. Ex vivo stability loss of irradiated and melted ultra-high molecular weight polyethylene10.2106/JBJS.I.01017 ↗
39. Surface cross-linked UHMWPE can enable the use of larger femoral heads in total joints10.1002/jor.22195 ↗
40. Organization, I.S. of, 2012. ISO 14242-1:2012(en), Implants for surgery — wear of total hip-joint prostheses — Part 1: loading and displacement parameters for wear-testing machines and corresponding environmental conditions for test.
41. Characterization of molecular-orientaiton in polyethyleen by Raman-spectroscopy10.1021/ma00020a032 ↗
42. Surface roughness quantification of CoCrMo implant alloys10.1002/(SICI)1097-4636(1999)48:5<705::AID-JBM16>3.0.CO;2-U ↗
43. Surface roughness of retrieved CoCrMo alloy femoral components from PCA artificial total knee joints
44. Correlation between in vivo stresses and oxidation of UHMWPE in total hip arthroplasty10.1007/s10856-014-5254-9 ↗
45. Oxidation and other property changes of a remelted highly crosslinked UHMWPE in retrieved tibial bearings10.1002/jbm.b.33240 ↗
46. Characterization and accelerated ageing of UHMWPE used in orthopedic prosthesis by peroxide10.3390/ma2020562 ↗
47. The combined effects of crosslinking and high crystallinity on the microstructural and mechanical properties of ultra high molecular weight polyethylene10.1016/j.biomaterials.2005.09.033 ↗
48. Ultra high molecular weight polyethylene: mechanics, morphology, and clinical behavior10.1016/j.jmbbm.2008.12.006 ↗
49. Raman spectroscopic method for determining the crystallinity of polyethylene10.1002/pol.1978.180160704 ↗
50. Wear behaviour, fluorescence and SEM investigations on nanocomposite zirconia-toughened alumina10.1007/s10853-006-0356-5 ↗
51. Highly cross-linked polyethylene in total hip and knee replacement: spatial distribution of molecular orientation and shape recovery behavior10.1155/2014/808369 ↗
52. The effects of contact area and applied load on the morphology of in vitro worn ultra-high molecular weight knee prostheses: a micro-Raman and gravimetric study10.1002/jrs.4545 ↗
53. Three-body wear of UHMWPE acetabular cups by PMMA particles against CoCr, alumina and zirconia heads in a hip joint simulator10.1016/S0043-1648(01)00643-3 ↗
54. Highly crosslinked polyethylenes in hip replacements: improved wear performance or paradox?10.1080/10402000701325204 ↗
55. Specific heat of synthetic high polymers. VIII. Low pressure polyethylene10.1002/pol.1957.1202410604 ↗