journal article·2021-06Open access

Non-invasive early detection of failure modes in total hip replacements (THR) via acoustic emission (AE)

Christine Lee, Lu Zhang, Dalton Morris, Kai Yuan Cheng, Remya Ampadi Ramachandran, Mark Barba, Divya Bijukumar, Didem Ozevin, Mathew T. Mathew

Abstract

Total hip replacements (THR) are becoming an common orthopedic surgucal procedure in the United States (332 K/year in 2017) to relieve pain and improve the mobility of those that are affected by osteoarthritis, ankylosing spondylitis, or injury. However, complications like tribocorrosion, or material degradation due to friction and corrosion, may result in THR failure. Unfortunately, few strategies to non-invasively diagnose early-stage complications are reported in literature, leading to implant complications being detected after irreversible damage. Therefore, the main objective of this study proposes the utilization of acoustic emission (AE) to continuously monitor implant materials, CoCrMo and Ti6Al4V, and identify degradations formed during cycles of sleeping, standing, and walking by correlating them to potential and friction coefficient behavior. AE activity detected from the study correlates with the friction coefficient and open-circuit potential observed during recreated in-vitro standing, walking, and sleeping cycles. It was found that the absolute energy level obtained from AE increased as the friction coefficient increased, potential decreased, and wear volume loss increased. Through the results, higher friction coefficient and AE activity were observed in Ti6Al4V alloys while there was also a significant drop in potential, indicating increased tribocorrosion activity. Therefore, AE can be utilized to predict material degradations as a non-invasive method based on the severity of abnormality of the absolute energy and hits emitted. The correlation between potential, friction coefficient, and AE activity was further confirmed through profilometry which showed more material degradation in Ti6Al4V than CoCrMo. Through these evaluations, it was demonstrated that AE could be utilized to identify the deformations and failure modes of implant materials caused by tribocorrosion.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Dental materials and restorations · Orthopedic Infections and Treatments

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (16)

Show source-provided references
1. Adhesive and abrasive wear studies using acoustic emission techniques10.1016/0043-1648(91)90363-Y ↗
2. Altering the course of technologies to monitor loosening states of endoprosthetic implants
3. Monitoring the Integrity of the C e m e n t - M e t a l Interface of Total Joint C o m p o n e n t s in vitro Using Acoustic Emission and Ultrasound10.1016/S0883-5403(96)80115-X ↗
4. Acta Biomaterialia Tribo-electrochemical characterization of metallic biomaterials for total joint replacement10.1016/j.actbio.2011.09.034 ↗
5. The role of synovial fluid in joint lubrication
6. Flexible molybdenum electrodes towards designing affinity based protein biosensors10.3390/bios6030036 ↗
7. Acoustic emission in orthopaedics : a state of the art review10.1016/j.jbiomech.2016.10.038 ↗
8. Karras, K., Pullin, R., Grosvenor, R. I. & Clarke, A. Damage Detection of a Composite Bearing Liner Using Acoustic Emission. 2–3.
9. Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 203010.2106/00004623-200704000-00012 ↗
10. CoCrMo metal-on-metal hip replacements10.1039/C2CP42968C ↗
11. Bio-functional nano-coatings on metallic biomaterials10.1016/j.msec.2015.05.018 ↗
12. Tribocorrosion behavior of CoCrMo alloy for hip prosthesis as a function of loads: a comparison between two testing systems10.1016/j.wear.2011.01.086 ↗
13. Real time monitoring of progressive damage during loading of a simplified total hip stem construct using embedded acoustic emission sensors10.1016/j.medengphy.2010.10.025 ↗
14. Multifunctional coatings to simultaneously promote osseointegration and prevent infection of orthopaedic implants10.1016/j.biomaterials.2016.01.016 ↗
15. Hip implant performance prediction by acoustic emission techniques: a review