Wear characteristics of WSU total ankle replacement devices under shear and torsion loads
Source abstract
Abstract
Background There are several factors that contribute to the failure of total ankle replacement (TAR). Aseptic loosening is one of the primary mechanisms of failure in TAR. Since a cross-linked ultrahigh molecular weight polyethylene (UHMWPE) is used as liner material, there is a need to quantify and develop methods to estimate the wear rates of the liners. High contact stresses develop during the gait generates wear debris resulting in osteolysis and early loosening of the prostheses. Methods In this paper wear characteristics of Wright State University (WSU) TARs were determined by applying shear and torsion loads. Viscoelastic properties were used to model the liner component. Finite element analysis was conducted to determine the wear rate by deriving Von Mises and contact stresses generated in the liner and wear rate equation was used to predict the wear rate. Results Titanium alloy has shown less resistance towards shear forces when compared with other metal alloys. Under torsion, rotation angle plays a significant role in affecting the peak stress values. The maximum average contact stress was 14.46 MPa under torsion load which contributes to a wear rate of 0.67 (mm(3)/year) for one of the mobile bearing models. The maximum average contact stress and wear rate obtained from the analytical study were 10.55 MPa and 0.33 (mm(3)/year), respectively for mobile bearing models. When compared with mobile bearing model, fixed bearing model has shown higher stresses at different degrees of rotation. Conclusion Both shear and torsion loads cause significantly lower contact stresses and wear when compared to the axial load. Further studies are necessary to accurately determine the wear behavior of fixed bearing TAR models.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Prosthetics and Rehabilitation Robotics · Tribology and Wear Analysis
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (57)
Show source-provided references
1. Meniscal wear at a three-component total ankle prosthesis by a knee joint simulator10.1016/j.jbiomech.2006.08.002 ↗
2. Wear behaviour in total ankle replacement: a comparison between an in vitro simulation and retrieved prostheses10.1016/j.clinbiomech.2009.06.006 ↗
3. Intra-articular contact stress distributions at the ankle throughout stance phase-patient-specific finite element analysis as a metric of degeneration propensity10.1007/s10237-006-0025-2 ↗
4. Simulation of polyethylene wear in ankle joint prostheses10.1002/jbm.b.30649 ↗
5. Coughlin · 2014 · 3
6. The state of stress induced by the plane frictionless cylindrical contact. I. The case of elastic similarity10.1016/S0020-7683(00)00289-4 ↗
7. The state of stress induced by the plane frictionless cylindrical contact. II. The general case (elastic dissimilarity)10.1016/S0020-7683(00)00290-0 ↗
8. Modelling of the ankle joint complex. Reflections with regards to ankle prostheses10.1016/j.fas.2004.06.003 ↗
9. Plasticity-induced damage layer is a precursor to wear in radiation-cross-linked UHMWPE acetabular components for total hip replacement10.1016/S0883-5403(99)90086-4 ↗
10. The effect of sliding velocity on the friction and wear of UHMWPE for use in total artificial joints10.1016/0043-1648(94)90185-6 ↗
11. Fryman · 2011
12. Impact of talar component rotation on contact pressure after total ankle arthroplasty: a cadaveric study10.3113/FAI.2010.0404 ↗
13. 257 Ankle arthroplasties performed in Norway between 1994 and 200510.1080/17453670710014257 ↗
14. A controlled, comparative study of a fixed-bearing versus mobile-bearing ankle arthroplasty10.1177/1071100713517094 ↗
15. History and evolution in total ankle arthroplasty10.1093/bmb/ldn039 ↗
16. Management of specific complications after revision total ankle replacement
17. Thermal isotherms in PMMA and cell necrosis during total hip arthroplasty
18. Total Ankle Arthroplasty: The US Experience
19. The Swedish Ankle Arthroplasty Register: an analysis of 531 arthroplasties between 1993 and 200510.1080/17453670710014248 ↗
20. On the contact of elastic solids
21. Finite element analysis of stress and wear characterization in total ankle replacements10.1016/j.jmbbm.2014.01.020 ↗
22. Load-bearing pattern of the ankle joint10.1007/BF01246141 ↗
23. A comparison of the influence of global functional loads vs. local contact anatomy on articular cartilage thickness at the knee10.1016/j.jbiomech.2007.02.005 ↗
24. Revision rates after total joint replacement: cumulative results from worldwide joint register datasets10.1302/0301-620X.93B3.25467 ↗
25. Polyethylene wear in total hip and knee arthroplasties10.1002/(SICI)1097-4636(199721)38:1<55::AID-JBM8>3.0.CO;2-G ↗
26. Ultra-high molecular weight polyethylene. The material and its use in total joint implants10.2106/00004623-199407000-00020 ↗
27. Titanium alloys in total joint replacement – a materials science perspective10.1016/S0142-9612(97)00146-4 ↗
28. Joint loading in the lower extremities during elliptical exercise10.1249/mss.0b013e3180dc9970 ↗
29. Hip implant stem interfacial motion, a finite element analysis10.1504/IJECB.2011.039946 ↗
30. McDowell · 2005
31. Biomechanics of the ankle joint and clinical outcomes of total ankle replacement10.1016/j.jmbbm.2008.01.005 ↗
32. Stresses in polyethylene liners in a semiconstrained ankle prosthesis10.1115/1.1798011 ↗
33. A novel method of cross-linking ultra-high-molecular-weight polyethylene to improve wear, reduce oxidation, and retain mechanical properties: recipient of the 1999 HAP Paul Award10.1054/arth.2001.20540 ↗
34. Joint contact characteristics in Agility total ankle arthroplasty10.1097/01.blo.0000132461.34725.a7 ↗
35. 2010
36. Total ankle replacement in rheumatoid arthritis10.1007/s00264-003-0512-3 ↗
37. Effect of head size on wear properties of metal-on-metal bearings of hip prostheses, and comparison with wear properties of metal-on-polyethylene bearings using hip simulator10.1016/j.jmbbm.2013.10.021 ↗
38. Perry · 1992
39. Persson · 1964
40. Postak · 2008
41. Finite element analysis of a total ankle replacement during the stance phase of gait10.1016/j.jbiomech.2005.04.010 ↗
42. The prediction of muscular load sharing and joint forces in the lower extremities during walking10.1016/0021-9290(75)90089-5 ↗
43. Average patient walking activity approaches 2 million cycles per year: pedometers under-record walking activity10.1054/arth.2002.32699 ↗
44. Comparison of reoperation rates following ankle arthrodesis and total ankle arthroplasty10.2106/00004623-200710000-00007 ↗
45. Force and motion analysis of the normal, diseased, and prosthetic ankle joint10.1097/00003086-197709000-00027 ↗
46. The effects of irradiation and EtO-treatment on ultrahigh molecular weight polyethylene acetabular cups following accelerated aging: degradation of mechanical properties and morphology changes during hip simulator tests10.1016/j.molstruc.2007.04.042 ↗
47. Closed-form solution for the contact problem of reinforced pin-loaded joints used in glass structures10.1016/j.ijsolstr.2006.10.030 ↗
48. On the conforming contact problem in a reinforced pin-loaded structure with a non-zero second Dundurs׳ constant10.1016/j.ijsolstr.2008.02.010 ↗
49. Contribution of articular surface geometry to ankle stabilization10.2106/JBJS.E.00758 ↗
50. Contact mechanics analysis of metal-on-metal hip resurfacing prostheses10.1243/0954411041932854 ↗
51. Kinematic changes after fusion and total replacement of the ankle part 1: range of motion10.1177/107110070302401202 ↗
52. Kinematic changes after fusion and total replacement of the ankle part 2: movement transfer10.1177/107110070302401203 ↗
53. Kinematic changes after fusion and total replacement of the ankle part 3: talar movement10.1177/107110070302401204 ↗
54. Mobile-and fixed-bearing total ankle prostheses: is there really a difference?10.1016/j.fcl.2012.08.005 ↗
55. Analysis of retrieved Agility™ total ankle arthroplasty systems10.3113/FAI.2009.0815 ↗
56. The effect of real-time aging on the oxidation and wear of highly cross-linked UHMWPE acetabular liners10.1016/j.biomaterials.2005.10.002 ↗
57. Yaszemski · 2003