Comprehensive computational analysis of the crimping procedure of PLLA BVS: effects of material viscous-plastic and temperature dependent behavior
Source abstract
Abstract
Recently, researchers focused their attention on the use of polymeric bioresorbable vascular scaffolds (BVSs) as alternative to permanent metallic drug-eluting stents (DESs) for the treatment of atherosclerotic coronary arteries. Due to the different mechanical properties, polymeric stents, if compared to DESs, are characterized by larger strut size and specific design. It implies that during the crimping phase, BVSs undergo higher deformation and the packing of the struts, making this process potentially critical for the onset of damage. In this work, a computational study on the crimping procedure of a PLLA stent, inspired by the Absorb GT1 (Abbott Vascular) design, is performed, with the aim of evaluating how different strategies (loading steps, velocities and temperatures) can influence the results in terms of damage risk and final crimped diameter. For these simulations, an elastic-viscous-plastic model was adopted, based on experimental results, obtained from tensile testing of PLLA specimens loaded according to ad hoc experimental protocols. Furthermore, the results of these simulations were compared with those obtained by neglecting strain rate and temperature dependence in the material model (as often done in the literature), showing how this lead to significant differences in the prediction of the crimped diameter and internal stress state.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Coronary Interventions and Diagnostics · Magnesium Alloys: Properties and Applications · Elasticity and Material Modeling
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (20)
Show source-provided references
1. An overview of mechanical properties and material modeling of polylactide (PLA) for medical applications10.1007/s10439-015-1455-8 ↗
2. Constitutive modeling of ultra-high molecular weight polyethylene under large-deformation and cyclic loading conditions10.1016/S0142-9612(01)00367-2 ↗
3. Characterisation and constitutive modelling of biaxially stretched poly(L-lactic acid)sheet for application in coronary stents10.1016/j.jmbbm.2019.05.039 ↗
4. Experimental mechanical testing of Poly (L-Lactide) (PLLA) to facilitate pre-degradation characteristics for application in cardiovascular stenting10.1016/j.polymertesting.2016.07.011 ↗
5. Computational analysis of the utilisation of the shape memory effect and balloon expansion in fully polymeric stent deployment10.1007/s13239-017-0333-y ↗
6. Computational bench testing to evaluate the short-term mechanical performance of a polymeric stent10.1007/s13239-015-0235-9 ↗
7. Understanding the requirements of self-expandable stents for heart valve replacement: radial force, hoop force and equilibrium10.1016/j.jmbbm.2017.02.006 ↗
8. A finite element strategy to investigate the free expansion behaviour of a biodegradable polymeric stent10.1016/j.jbiomech.2015.03.024 ↗
9. Eswaran
10. Jow · 2012
11. Simultaneous kissing stents to treat unprotected left main stem coronary artery bifurcation disease; stent expansion, vessel injury, hemodynamics, tissue healing, restenosis, and repeat revascularization10.1002/ccd.27640 ↗
12. Modeling the response of nonlinear viscoelastic biodegradable polymeric stents10.1016/j.ijsolstr.2011.12.007 ↗
13. Computational analysis of the radial mechanical performance of PLLA coronary artery stents10.1016/j.medengphy.2014.09.014 ↗
14. Effect of two-year degradation on mechanical interaction between a bioresorbable scaffold and blood vessel10.1016/j.jmbbm.2017.11.031 ↗
15. A computational study of crimping and expansion of bioresorbable polymeric stents10.1007/s11043-017-9371-y ↗
16. Computational analysis of mechanical stress–strain interaction of a bioresorbable scaffold with blood vessel10.1016/j.jbiomech.2016.05.035 ↗
17. Crimping and deployment of metallic and polymeric stents – finite element modelling10.20517/2574-1209.2016.03 ↗
18. Modeling of deformation-accelerated breakdown of polylactic acid biodegradable stents
19. Multimodal loading environment predicts bioresorbable vascular scaffolds' durability
20. Computational and experimental investigation into mechanical performances of Poly-L-Lactide Acid (PLLA) coronary stents10.1016/j.jmbbm.2016.08.033 ↗