journal article·2015-11

Contact damage failure analyses of fretting wear behavior of the metal stem titanium alloy–bone cement interface

Lanfeng Zhang, Shirong Ge, Hongtao Liu, Qingliang Wang, Liping Wang, Cory J. Xian

Publisher page

Abstract

Although cemented titanium alloy is not favored currently in the Western world for its poor clinical and radiography outcomes, its lower modulus of elasticity and good biocompatibility are instrumental for its ability supporting and transforming physical load, and it is more suitable for usage in Chinese and Japanese populations due to their lower body weights and unique femoral characteristics. Through various friction tests of different cycles, loads and conditions and by examining fretting hysteresis loops, fatigue process curves and wear surfaces, the current study investigated fretting wear characteristics and wear mechanism of titanium alloy stem-bone cement interface. It was found that the combination of loads and displacement affected the wear quantity. Friction coefficient, which was in an inverse relationship to load under the same amplitude, was proportional to amplitudes under the same load. Additionally, calf serum was found to both lubricate and erode the wear interface. Moreover, cement fatigue contact areas appeared black/oxidative in dry and gruel in 25% calf serum. Fatigue scratches were detected within contact areas, and wear scars were found on cement and titanium surfaces, which were concave-shaped and ring concave/ convex-shaped, respectively. The coupling of thermoplastic effect and minimal torque damage has been proposed to be the major reason of contact damage. These data will be important for further studies analyzing metal-cement interface failure performance and solving interface friction and wear debris production issues.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Mechanical stress and fatigue analysis · Orthopedic Infections and Treatments

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (72)

Show source-provided references
1. Long-term results of cemented total hip arthroplasty in developmental dysplasia with acetabular bulk bone grafts after improving operative techniques10.1016/j.arth.2009.05.017 ↗
2. Clinical outcomes of cemented double-tapered titanium femoral stems: a minimum 5-year follow-up10.1007/s00776-011-0154-z ↗
3. In-vitro characteristics of cemented titanium femoral stems with a smooth surface finish10.1007/s00776-012-0298-5 ↗
4. Ageing and moisture uptake in polymethyl methacrylate (PMMA) bone cements10.1016/j.jmbbm.2013.12.010 ↗
5. Ageing and moisture uptake in polymethyl methacrylate (PMMA) bone cements10.1016/j.jmbbm.2013.12.010 ↗
6. Banaszkiewicz · 2014 · 81
7. The effect of mechanical loading on the size and shape of bone in pre-, peri-, and postpubertal girls: a study in tennis players10.1359/jbmr.2002.17.12.2274 ↗
8. The influence of contact conditions and micromotions on the fretting behavior of modular titanium alloy taper connections10.1016/j.medengphy.2012.07.013 ↗
9. Bhushan · 2013
10. What results in fretting wear on polished femoral stems10.1016/j.triboint.2008.11.007 ↗
11. The measurement of hip joints of the people in south of China and clinical significance
12. A review on the polymer properties of Hydrophilic, partially Degradable and Bioactive acrylic Cements (HDBC)10.1016/j.progpolymsci.2007.09.001 ↗
13. Is titanium so bad? Medium-term outcome of cemented titanium stems10.1302/0301-620X.83B5.10846 ↗
14. Breusch · 2006
15. Fretting corrosion of fully cemented polished collarless tapered stems: the influence of PMMA bone cement10.1016/j.wear.2012.12.042 ↗
16. Comparative study of PDGF-B and BMP-7 receptors distribution on histiocytes of interface membranes around aseptic loosened prostheses
17. Water resistance of bone-cements reinforced with bioceramics10.1016/j.matlet.2006.08.072 ↗
18. Water absorption characteristics of modified hydroxyapatite bone cements10.1016/0142-9612(95)98906-U ↗
19. Fretting wear study on the interface between titanium alloy ball and bone cement
20. Revision of the cemented femoral stem using a cement-in-cement technique. A FIVE-TO 15-YEAR REVIEW10.1302/0301-620X.91B5.21621 ↗
21. Effect of surgical approach for total hip replacement on hip function using Harris Hip scores and Trendelenburg׳s test. A retrospective analysis10.1016/j.surge.2010.08.014 ↗
22. Results of a cemented straight titanium alloy femoral stem after mean follow-up of 13 years10.1007/s00590-007-0232-2 ↗
23. Quantification of implant micromotion, strain shielding, and bone resorption with porous-coated anatomic medullary locking femoral prostheses10.1097/00003086-199212000-00005 ↗
24. Physical and mechanical characteristic of bone cement mixed with antibiotics
25. The Nicolas Andry award: comparative results of cemented and cementless total hip arthroplasty10.1097/00003086-199604000-00044 ↗
26. Evaluation of clinical outcomes and bone loss around titanium implants with oxidized surface: six-year follow-up results from a prospective case series study10.1111/j.1708-8208.2012.00466.x ↗
27. Fretting-corrosion of materials used as orthopaedic implants10.1016/j.wear.2004.11.027 ↗
28. Friction/fretting-corrosion mechanisms: current trends and outlooks for implants10.1016/j.matlet.2014.06.167 ↗
29. Atomic force microscopy investigations on pits and debris related to fretting-corrosion between 316L SS and PMMA10.1016/j.wear.2012.05.008 ↗
30. The future of biologic coatings for orthopaedic implants10.1016/j.biomaterials.2013.01.074 ↗
31. Is there still a place for the cemented titanium femoral stem? 10,108 cases from the Norwegian Arthroplasty Register10.3109/17453674.2011.645194 ↗
32. The problem is osteolysis
33. A comparative clinical outcome evaluation of smooth (10–13 year results) versus rough surface finish (5–8 year results) in an otherwise identically designed cemented titanium alloy stem10.1007/s00402-003-0515-y ↗
34. In vivo skeletal responses to porous-surfaced implants subjected to small induced motions10.2106/00004623-199705000-00010 ↗
35. Health outcome measures in the evaluation of total hip arthroplasties – a comparison between the Harris Hip Score and the Oxford Hip Score10.1016/j.arth.2005.04.017 ↗
36. Does early micromotion of femoral stem prostheses matter? 4–7-year stereoradiographic follow-up of 84 cemented prostheses10.1302/0301-620X.76B6.7983118 ↗
37. Cytokines in aseptic loosening of total hip replacement10.1016/S0268-0890(97)90050-5 ↗
38. Long-term survival of a cemented titanium–aluminium–vanadium alloy straight-stem femoral component10.1302/0301-620X.88B12.17796 ↗
39. Landolt · 2011
40. Evaluation of cement stresses in finite element analyses of cemented orthopaedic implants10.1115/1.1412452 ↗
41. Influence of micro-motion initiated by relative abrasions between prosthetic metallic component and bone cement: comparative studies
42. Fretting wear study on the interface between titanium alloy ball and bone cement
43. Fretting wear behavior of nitrogen ion implanted titanium alloys in bovine serum lubrication10.1016/j.triboint.2009.04.009 ↗
44. Fretting corrosion processes and wear mechanisms in medical implants
45. Toxicology of wear particles of cobalt-chromium alloy metal-on-metal hip implants Part I: Physicochemical properties in patient and simulator studies10.1016/j.nano.2014.12.005 ↗
46. The Swedish total hip replacement register10.2106/00004623-200200002-00002 ↗
47. Stress gradient effects in fretting fatigue10.1016/S0301-679X(02)00134-2 ↗
48. Evaluation of elution and mechanical properties of high-dose antibiotic-loaded bone cement: comparative “in vitro” study of the influence of vancomycin and cefazolin10.1016/j.arth.2015.02.040 ↗
49. Fretting-corrosion between 316L SS and PMMA: influence of ionic strength, protein and electrochemical conditions on material wear. Application to orthopaedic implants10.1016/j.wear.2011.01.082 ↗
50. The effects of different mixing speeds on the elution and strength of high-dose antibiotic-loaded bone cement created with the hand-mixed technique10.1016/j.arth.2014.12.003 ↗
51. CoCr wear particles generated from CoCr alloy metal-on-metal hip replacements, and cobalt ions stimulate apoptosis and expression of general toxicology-related genes in monocyte-like U937 cells10.1016/j.taap.2014.09.010 ↗
52. The morphology of the proximal femur. A three-dimensional radiographic analysis10.1302/0301-620X.74B1.1732260 ↗
53. Roentgen stereophotogrammetric analysis of prosthetic fixation in the hip and knee joint
54. Influence of roughness on surface instability of medical grade cobalt–chromium alloy (CoCrMo) during contact corrosion–fatigue10.1016/j.apsusc.2013.02.076 ↗
55. Influence of the physiological medium on the mechanical properties of bone cement: can current studies be extrapolated?
56. Radiographic analysis of a low-modulus titanium-alloy femoral total hip component. Two to six-year follow-up10.2106/00004623-198567010-00007 ↗
57. Five-year results of two cemented hip stem models each made of two different alloys10.1007/s00402-004-0772-4 ↗
58. Sosnovskiy · 2010
59. Effect of heterotopic ossification on hip range of motion and clinical outcome10.1016/j.arth.2014.09.019 ↗
60. Subsidence of THA stems due to acrylic cement creep is extremely sensitive to interface friction10.1016/S0021-9290(96)80008-X ↗
61. Contact of rough surfaces: a review of experimental work10.1016/0043-1648(80)90162-3 ↗
62. Yan · 2013
63. Study on torsional fretting corrosion behaviors of Ti6Al4V alloy in the serum solution
64. Dual motion fretting wear behaviors of titanium and its alloy in artificial saliva10.1016/S1003-6326(14)63033-8 ↗
65. The significance of the micropores at the stem–cement interface in total hip replacement10.1163/092050610X540495 ↗
66. The influence of bone cement type on production of fretting wear on the femoral stem surface: a preliminary study10.1016/j.clinbiomech.2012.02.008 ↗
67. Understanding initiation and propagation of fretting wear on the femoral stem in total hip replacement10.1016/j.wear.2008.04.076 ↗
68. The contribution of the micropores in bone cement surface to generation of femoral stem wear in total hip replacement10.1016/j.triboint.2010.11.007 ↗
69. Biotribological properties at the stem–cement interface lubricated with different media10.1016/j.jmbbm.2013.01.001 ↗
70. Bond strength analysis of the bone cement-stem interface of hip arthroplasties10.1016/S1995-7645(14)60013-4 ↗
71. Experiment investigation on fretting wear and wear debris performance for the stem–cement interface
72. The effect of acrylic bone cement solidification conditions on the compressive property10.4028/www.scientific.net/AMR.197-198.1694 ↗