An approach to evaluating and benchmarking the mechanical behavior of a surgical mesh prototype designed for the repair of abdominal wall defects

C. García-García, M.T. Carrascal-Morillo, J.T. Castell Gómez, C. Bernal Guerrero, J.C. García Prada

Abstract

Ventral hernia repair is a common surgical procedure in abdominal surgery in which surgical mesh has become an essential tool to improve outcomes. To avoid recurrences the mesh needs to mimic the mechanical behavior of the abdominal wall. In this scenario the mechanical properties at the interface between the mesh and its surrounding tissue is critical for the performance of the device and, therefore, the success after surgery. We aimed to characterize and compare the mechanical behavior of the patented prototype mesh Spider and four commercial meshes at the mesh–tissue interface. The prototype mesh was designed based on the hypothesis that the best performance for a large-sized defect in a ventral hernia is obtained when the mesh presents an isotropic behavior. In contrast, commercial meshes presented significant anisotropic behavior. Mechanical properties of the meshes were characterized through uniaxial tensile tests. Longitudinal and transverse axes were defined for each mesh, and samples were cut in each axis orientation. Samples underwent uniaxial tensile testing, from which the elastic modulus in each axis was determined. The degree of anisotropy was calculated as the ratio between the elastic modulus in each axis. An in silico model of the ventral hernia defect was designed to simulate the mesh–tissue interface behavior via finite element method. Meshes were modeled by an hyperelastic orthotropic constitutive model, which allowed isotropic symmetry as particular case for the prototype mesh. Abdominal wall was modeled using a Neo-Hookean model. Once the simulations were launched, mesh–tissue interface behavior was evaluated through the difference between Von Mises stress values on either size of the interface, both on the external and the internal face of the mesh and abdominal wall. Mechanical response was anisotropic for all commercial meshes and isotropic for the Spider prototype. Among commercial, Ultrapro® was highly anisotropic. Tests revealed Gore-Tex® to be the stiffest, followed by Repol Angimesh®, Spider and Ultrapro®; Duramesh™ was found to be the most compliant. Concerning mesh–tissue behavior, simulation results revealed the Spider prototype and Duramesh™ to be the best; Spider due to its uniformity and lower stress difference thanks to its nearly isotropic behavior, and Duramesh™ due to its compliant behavior. Our results suggest that the compromise between stiffness and anisotropy must be considered in order to improve the mechanical performance of the meshes, bearing in mind that for large-sized ventral defects, nearly isotropic mesh ensures better performance.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Hernia repair and management · Pelvic and Acetabular Injuries · Abdominal Surgery and Complications

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (38)

Show source-provided references
1. Biomechanical compatibility of surgical mesh and fascia being reinforced: dependence of experimental hernia defect repair results on anisotropic surgical mesh positioning10.1007/s10029-011-0877-y ↗
2. Surgical anatomy of the aponeurotic expansions of the anterior abdominal wall
3. Collagen fibers in linea alba and rectus sheaths: I. General scheme and morphological aspects10.1006/jsre.2000.6070 ↗
4. Comparing the behavior of different polypropylene meshes (heavy and lightweight) in an experimental model of ventral hernia repair10.1002/jbm.b.31234 ↗
5. Biomechanical analyses of overlap and mesh dislocation in an incisional hernia model in vitro10.1016/j.surg.2007.04.024 ↗
6. Castell Gómez · 2016
7. Titanium-coated micromachined grooves of different dimensions affect epithelial and connective-tissue cells differentlyin vivo10.1002/jbm.820240906 ↗
8. Normal intrabdominal pressure in healthy adults10.1016/j.jss.2005.06.015 ↗
9. Pitfalls in retromuscular mesh repair for incisional hernia: The importance of the “fatty triangle”10.1007/s10029-004-0235-4 ↗
10. Physicomechanical evaluation of polypropylene, polyester, and polytetrafluoroethylene meshes for inguinal hernia repair10.1016/j.jamcollsurg.2010.09.012 ↗
11. Mechanical properties of the abdominal wall and biomaterials utilized for hernia repair10.1016/j.jmbbm.2017.05.008 ↗
12. Biaxial analysis of synthetic scaffolds for hernia repair demonstrates variability in mechanical anisotropy, non-linearity and hysteresis10.1016/j.jmbbm.2014.06.001 ↗
13. Multi-directional mechanical analysis of synthetic scaffolds for hernia repair10.1016/j.jmbbm.2017.02.009 ↗
14. Mechanical response of the herniated human abdomen to the placement of different prostheses10.1115/1.4023703 ↗
15. Mechanical behaviour of synthetic surgical meshes: Finite element simulation of the herniated abdominal wall10.1016/j.actbio.2011.06.033 ↗
16. Long-term anisotropic mechanical response of surgical meshes used to repair abdominal wall defects10.1016/j.jmbbm.2011.09.005 ↗
17. Mechanical and histological characterization of the abdominal muscle. A previous step to modelling hernia surgery10.1016/j.jmbbm.2010.11.012 ↗
18. A new and evolving paradigm for biocompatibility10.1002/term.4 ↗
19. A class of orthotropic and transversely isotropic hyperelastic constitutive models based on a polyconvex strain energy function10.1016/j.ijsolstr.2004.02.027 ↗
20. Elasticity of the anterior abdominal wall and impact for reparation of incisional hernias using mesh implants10.1007/s100290100019 ↗
21. Breaking strength and tissue elasticity after Shouldice repair10.1007/s10029-002-0103-z ↗
22. Pathophysiology of the abdominal wall
23. Modified mesh for hernia repair that is adapted to the physiology of the abdominal wall10.1080/110241598750005138 ↗
24. Shrinking of polypropylene mesh in vivo: an experimental study in dogs10.1080/110241598750005156 ↗
25. The effect of food consumption on the thickness of abdominal muscles, employing ultrasound measurements10.1007/s00586-011-1708-7 ↗
26. The tension-free hernioplasty10.1016/0002-9610(89)90526-6 ↗
27. Luenberger · 2008
28. Computation methods affect the reported values of in vivo human tendon stiffness10.1016/j.jmbbm.2011.08.008 ↗
29. Mechanical properties of mesh materials used for hernia repair and soft tissue augmentation10.1371/journal.pone.0046978 ↗
30. Abdominal muscle size and symmetry in normal subjects10.1002/mus.20589 ↗
31. Large elastic deformations of isotropic materials. IV. Further developments of the general theory
32. Uniaxial and biaxial mechanical characterization of a prosthetic mesh at different length scales10.1016/j.jmbbm.2013.07.021 ↗
33. Anisotropic evaluation of synthetic surgical meshes10.1007/s10029-010-0731-7 ↗
34. Elasticity of the living abdominal wall in laparoscopic surgery10.1016/j.jbiomech.2004.12.019 ↗
35. Modelling of abdominal wall under uncertainty of material properties
36. Computational modeling of abdominal hernia laparoscopic repair with a surgical mesh10.1007/s11548-017-1681-7 ↗
37. Synthetic surgical meshes used in abdominal wall surgery: Part II—Biomechanical aspects10.1002/jbm.b.33584 ↗
38. Walter · 2007