Explant analysis of a Maestro™ wrist prosthesis and calculation of its lubrication regime
Source abstract
Abstract
Explant analysis can provide key insights to understanding failures of artificial joints and thus how they might be improved for the ultimate benefit of patients. There are no previous reports of explant analysis of an artificial wrist joint. In this study, an explanted metal-on-polymer Maestro wrist was analysed both in macro and nanoscales to estimate its biotribological performance. The articulation was formed between a cobalt chromium carpal head and an ultrahigh molecular weight polyethylene bearing. The surface roughness values of its articulating surfaces and the backside of the articulation were measured. On average, the articulating surface roughness values were calculated as 0.06 ± 0.02 μm and 1.29 ± 0.63 μm for the cobalt chromium carpal head and ultrahigh molecular weight polyethylene bearing, respectively. Both surfaces had negative skewness, indicating a preponderance of valleys. On the articulating surface of the carpal head, light scratches were observed, and no impingement was observed throughout the component. The polymeric surface had a polished appearance. It had unidirectional scratches at the centre of the articulation, pits of different sizes on its articulating surface, and matt white subsurface regions. The backside of the UHMWPE bearing and the convex surface of the radial body that it was sitting on, were found to have average surface roughness values of 4.23 ± 0.69 μm and 5.57 ± 1.05 μm, respectively. The difference in the means was not significant (p > 0.05). Taking the articulating surface roughness values, the lubrication regime that the explanted Maestro wrist operated under in vivo was estimated for varying physiological conditions, i.e. varying loads, entraining velocities and lubricant viscosities. In every case considered, the explant was found to operate under boundary lubrication.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopedic Surgery and Rehabilitation · Orthopaedic implants and arthroplasty · Engineering Technology and Methodologies
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (42)
Show source-provided references
1. Systematic review of total wrist arthroplasty and arthrodesis in wrist arthritis10.1055/s-0038-1646956 ↗
2. Wear particles and osteolysis in patients with total wrist arthroplasty
3. A biomechanical study of normal functional wrist motion10.1097/00003086-198407000-00004 ↗
4. Bs ISO 4287 · 1997
5. 3D patient imaging and retrieval analysis help understand the clinical importance of rotation in knee replacements10.1007/s00167-018-4891-9 ↗
6. S. Chacko Rajan, O. Bretcanu, D. J. Deehan, and T. J. Joyce, "Retrieval analysis of two contemporary total knee designs: influence of femoral component roughness and type of polyethylene," Journal of the Mechanical Behavior of Biomedical Materials, vol. 104, 2020, doi: 10.1016/j.jmbbm.2020.103620.10.1016/j.jmbbm.2020.103620 ↗
7. Total wrist replacement: a retrospective comparative study10.1055/s-0032-1326728 ↗
8. Analysis of wear and friction of total knee replacements part II: friction and lubrication as a function of wear10.1016/j.wear.2008.02.023 ↗
9. Elastohydrodynamic lubrication of elliptical contacts for materials of low elastic modulus10.1115/1.3453284 ↗
10. Goniometer measurement and computer analysis of wrist angles and movements applied to occupational repetitive work10.1016/1050-6411(95)00017-8 ↗
11. Analysis of the in-vivo failure of the adhesive interlayer for a DLC coated articulating metatarsophalangeal joint10.1016/j.diamond.2012.02.001 ↗
12. Changes in surface topography at the TKA backside articulation following in vivo service: a retrieval analysis10.1007/s00167-014-3197-9 ↗
13. Retrieval analysis of total knee prostheses: a method and its application to 48 condylar prostheses10.1002/jbm.820170510 ↗
14. The tribological behaviour of different clearance MOM hip joints with lubricants of physiological viscosities10.1177/0954411911419061 ↗
15. Quantification of third body damage to the tibial counterface in mobile bearing knees10.1243/0954411011533733 ↗
16. Prediction of lubrication regimes in two-piece metatarsophalangeal prostheses10.1016/S0021-9290(06)85171-7 ↗
17. Examination of failed ex vivo metal-on-metal metatarsophalangeal prosthesis and comparison with theoretically determined lubrication regimes10.1016/j.wear.2006.11.045 ↗
18. Sites of fractures in explanted NeuFlex silicone metacarpophalangeal joint prostheses10.1177/1753193418774500 ↗
19. Quantification of self-polishing in vivo from explanted metal-on-metal total hip replacements10.1016/j.triboint.2010.04.007 ↗
20. The determination of moments for extension of the wrist generated by muscles of the forearm10.1016/S0363-5023(78)80083-5 ↗
21. BIAX total wrist arthroplasty: management and results after 42 patients10.1055/s-2007-964878 ↗
22. Oxidation, damage mechanisms, and reasons for revision of sequentially annealed highly crosslinked polyethylene in total knee arthroplasty10.1016/j.arth.2017.09.036 ↗
23. NNAU · 2019
24. Nydick · 2012 · 37A · 1580
25. Tibial interface wear in retrieved total knee components and correlations with modular insert motion10.2106/00004623-200210000-00017 ↗
26. Comparison of gas plasma and gamma irradiation in air sterilization on the delamination wear of the ultra-high molecular weight polyethylene used in knee replacements10.1243/0954411001535741 ↗
27. Wrist arthroplasty: bone fixation, clinical development and mid to long term results10.3109/17453674.2014.900597 ↗
28. Intra-articular pressure measurement in the radioulnocarpal joint using a novel sensor: in vitro and in vivo results10.1016/j.jhsa.2006.10.007 ↗
29. Total wrist arthroplasty: a single-center study of 219 cases with 5-year follow-up10.1016/j.jhsa.2015.09.016 ↗
30. Effect of shelf versus accelerated aging of UHMWPE on delamination in knee wear simulation10.1016/j.triboint.2014.01.001 ↗
31. Combined replacements using the Maestro total wrist and uHead ulnar head implants10.1177/1753193414535721 ↗
32. The MaestroTM Wrist Reconstructive System for treatment of posttraumatic wrist joint osteoarthritis after a devasting course of a distal radius fracture: case presentation and technical note to the implant10.15761/TEC.1000123 ↗
33. Does total wrist arthroplasty for treatment of posttraumatic wrist joint osteoarthritis in Young patients always lead to restriction of high-demand activities of daily living? Case report and brief review of recent literature10.2174/1874325001711010439 ↗
34. Combined replacements of the wrist, ulnar head and thumb carpometacarpal joint. Case report, technical note and recent evidence to the ArpeTM prosthesis
35. RE-MOTION total wrist arthroplasty for treatment of advanced stage of scaphoid for treatment of advanced stage of the scaphoid non-union advanced collapse. Does excision of the entire scaphoid bone prevent impingement at terminal range of radial deviation?10.15761/TEC.1000127 ↗
36. A critical appraisal to the decision by the company Zimmer Biomet to withdraw the MaestroTM Wrist Reconstructive System from the marketplace10.15761/TEC.1000162 ↗
37. Functional outcomes after salvage procedures for wrist trauma and arthritis (Four-Corner fusion, proximal row carpectomy, total wrist arthroplasty, total wrist fusion, wrist denervation): a review of literature10.2174/1874325001913010217 ↗
38. Topographical analysis of the femoral components of ex vivo total knee replacements10.1007/s10856-012-4815-z ↗
39. Force and pressure transmission through the normal wrist. A theoretical two-dimensional study in the posteroanterior plane10.1016/0021-9290(94)00093-J ↗
40. Surface analysis of early retrieved polyethylene tibial inserts for both knees in total knee replacement10.1016/j.engfailanal.2019.104279 ↗
41. 2009
42. Total wrist arthroplasty: a systematic review of the evidence from the last 5 years10.1177/1753193414539796 ↗