In-vitro characterization of self-expandable textile transcatheter aortic valves
Source abstract
Abstract
Objective This study aims at assessing the global dynamic behavior, closing energy and turbulence characteristics of self-expandable textile (inclined and straight yarn) transcatheter aortic valves (TAV) versus bioprosthetic TAVs. Methods Two self-expandable textile TAVs one with inclined yarn textile and another with straight yarn textile leaflets were assessed in a pulse duplicator and compared with a self-expandable commercial bioprosthetic TAV under physiological pressure and flow. Particle Image Velocimetry and high-speed imaging were performed. Effective orifice areas (EOA), leakage fractions (LF), Pinwheeling indices (PI), closing energy (E), viscous shear stresses (VSS) and Reynolds shear stresses (RSS) were calculated. Results (a) EOAs and LFs were 2.27 ± 0.03 cm2, 31.7 ± 0.6%; 2.25 ± 0.08 cm2, 26.6 ± 0.7%; and 1.63 ± 0.01 cm2, 29.1 ± 1.25% for inclined textile, bioprosthetic and straight textile TAV respectively (p < 0.0001). (b) Following same order, PIs were significantly different going from 1.16 ± 0.21%, 8.48 ± 0.8% and 8.865 ± 0.58% with the exception of CoreValve and straight yarn valve (p = 0.37); (c) E is lowest for straight textile TAV (0.0024 ± 0.0017 J), followed by bioprosthetic valve (0.00259 ± 0.0011 J) and then 45° Oriented Yarn Valve (0.00334 ± 0.03 J) (d) At peak systole, the highest RSS distribution was with the Straight textile TAV reaching up to 330Pa. The bioprosthetic TAV shows the smallest range with RSS reaching around 230Pa and the inclined textile TAV up to 280Pa. VSS limits were comparable among the 3 valves ranging between 5.2Pa and 5.7Pa. Conclusion Hemodynamic similarities were found between the textile self-expandable valves and the bioprosthetic valve. This study constitutes another step towards showing the potential that textile valves have to become an alternative for the biological ones.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Cardiac Valve Diseases and Treatments · Aortic Disease and Treatment Approaches · Cardiovascular and Diving-Related Complications
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (59)
Show source-provided references
1. Rethinking turbulence in blood
2. A general two-equation macroscopic turbulence model for incompressible flow in porous media10.1016/S0017-9310(96)00370-5 ↗
3. Correlation of thrombosis growth rate to pathological wall shear rate during platelet accumulation10.1002/bit.24537 ↗
4. Long-term durability of transcatheter aortic valve prostheses10.1016/j.jacc.2018.10.078 ↗
5. Valve-in-valve outcome: design impact of a pre-existing bioprosthesis on the hydrodynamics of an Edwards Sapien XT valve10.1093/ejcts/ezw317 ↗
6. Valve-in-valve outcome: design impact of a pre-existing bioprosthesis on the hydrodynamics of an Edwards Sapien XT valve
7. Characterization of hemodynamic forces induced by mechanical heart valves: Reynolds vs. viscous stresses10.1007/s10439-007-9411-x ↗
8. Estimation of shear stress-related blood damage in heart valve prostheses--in vitro comparison of 25 aortic valves10.1177/039139889001300507 ↗
9. A discussion on the threshold limit for hemolysis related to Reynolds shear stress
10. An in vitro evaluation of the impact of eccentric deployment on transcatheter aortic valve hemodynamics10.1007/s10439-014-1008-6 ↗
11. Total ellipse of the heart valve: the impact of eccentric stent distortion on the regional dynamic deformation of pericardial tissue leaflets of a transcatheter aortic valve replacement10.1098/rsif.2015.0737 ↗
12. Sinus hemodynamics in representative stenotic native bicuspid and tricuspid aortic valves: an in-vitro study10.3390/fluids3030056 ↗
13. Reduction of pressure gradient and turbulence using vortex generators in prosthetic heart valves10.1007/s10439-018-02128-6 ↗
14. Aortic sinus flow stasis likely in valve-in-valve transcatheter aortic valve implantation10.1016/j.jtcvs.2017.03.053 ↗
15. Stented valve dynamic behavior induced by polyester fiber leaflet material in transcatheter aortic valve devices10.1016/j.jmbbm.2018.06.038 ↗
16. On the significance of systolic flow waveform on aortic valve energy loss10.1007/s10439-018-2102-y ↗
17. Spatiotemporal complexity of the aortic sinus vortex as a function of leaflet calcification10.1007/s10439-019-02224-1 ↗
18. Effect of severe bioprosthetic valve tissue ingrowth and inflow calcification on valve-in-valve performance10.1016/j.jbiomech.2018.04.039 ↗
19. Sinus hemodynamics variation with tilted transcatheter aortic valve deployments10.1007/s10439-018-02120-0 ↗
20. A turbulence in vitro assessment of On-X and St Jude Medical prostheses
21. Impact of patient-specific morphologies on sinus flow stasis in transcatheter aortic valve replacement: an in vitro study10.1016/j.jtcvs.2018.05.086 ↗
22. Textile heartvalve prosthesis: manufacturing process and prototype performances10.1177/0040517508092007 ↗
23. Textile for heart valve prostheses: fabric long‐term durability testing10.1002/jbm.b.31490 ↗
24. Textile heart valve: novel shaping process and material performances10.1080/10426914.2011.552018 ↗
25. Textile heart valve: first in-vivo experiment in the aortic position
26. In vitro hemodynamic assessment of a novel polymeric transcatheter aortic valve10.1016/j.jmbbm.2019.06.016 ↗
27. Blood cell damage by mechanical forces
28. Hoffman · 1992
29. ISO · 2013 · 105
30. Transcatheter fiber heart valve: effect of crimping on material performances10.1002/jbm.b.33330 ↗
31. Incidence and risk factors of hemolysis after transcatheter aortic valve implantation with a balloon-expandable valve10.1016/j.amjcard.2015.02.059 ↗
32. New trends in experimental turbulence research10.1146/annurev.fl.07.010175.001515 ↗
33. Transcatheter or surgical aortic-valve replacement in intermediate-risk patients10.1056/NEJMoa1514616 ↗
34. Red blood cell damage by shear stress10.1016/S0006-3495(72)86085-5 ↗
35. Simulated transcatheter aortic valve deformation: a parametric study on the impact of leaflet geometry on valve peak stress10.1002/cnm.2814 ↗
36. Turbulence characteristics downstream of bileaflet aortic valve prostheses10.1115/1.429643 ↗
37. A reevaluation and discussion on the threshold limit for hemolysis in a turbulent shear flow10.1016/S0021-9290(01)00084-7 ↗
38. Transcatheter aortic-valve replacement with a balloon-expandable valve in low-risk patients10.1056/NEJMoa1814052 ↗
39. Effect of aortic annulus size and prosthesis oversizing on the hemodynamics and leaflet bending stress of transcatheter valves: an in vitro study
40. Simulation of long-term fatigue damage in bioprosthetic heart valves: effects of leaflet and stent elastic properties10.1007/s10237-013-0532-x ↗
41. Valve type, size, and deployment location affect hemodynamics in an in vitro valve-in-valve model10.1016/j.jcin.2016.05.030 ↗
42. Coronary flow impacts aortic leaflet mechanics and aortic sinus hemodynamics10.1007/s10439-015-1260-4 ↗
43. Flow simulations over porous media–Comparisons with experiments10.1016/j.compfluid.2017.03.002 ↗
44. Trauma of the erythrocyte membrane associated with low shear stress10.1016/S0006-3495(73)86056-4 ↗
45. AHA/ACC guideline for the management of patients with valvular heart disease: a report of the American College of cardiology/American heart association task force on practice guidelines10.1016/j.jacc.2014.02.536 ↗
46. Two-dimensional color-mapping of turbulent shear stress distribution downstream of two aortic bioprosthetic valves in vitro10.1016/0021-9290(92)90262-Y ↗
47. Shear stress related blood damage in laminar Couette flow10.1046/j.1525-1594.2003.07103.x ↗
48. Transcatheter aortic valve replacement using a self-expanding bioprosthesis in patients with severe aortic stenosis at extreme risk for surgery10.1016/j.jacc.2014.02.556 ↗
49. Transcatheter aortic-valve replacement with a self-expanding valve in low-risk patients10.1056/NEJMoa1816885 ↗
50. Clinical valve thrombosis and subclinical leaflet thrombosis following transcatheter aortic valve replacement: is there a need for a patient-tailored anti-thrombotic therapy?10.3389/fcvm.2019.00044 ↗
51. Novel polymeric valve for transcatheter aortic valve replacement applications: in vitro hemodynamic study10.1007/s10439-018-02119-7 ↗
52. Human red blood cell hemolysis in a turbulent shear flow: contribution of Reynolds shear stresses
53. Durability of transcatheter and surgical bioprosthetic aortic valves in patients at lower surgical risk10.1016/j.jacc.2018.10.083 ↗
54. Fluid mechanics of heart valves and their replacements10.1146/annurev-fluid-122414-034314 ↗
55. Deformation and fragmentation of human red blood cells in turbulent shear flow10.1016/S0006-3495(75)85787-0 ↗
56. Fiber heart valve prosthesis: early in vitro fatigue results10.1002/jbm.b.33393 ↗
57. Heart valves from polyester fibers: a preliminary 6-month in vivo study10.1515/bmt-2016-0242 ↗
58. Fluid mechanics of heart valves10.1146/annurev.bioeng.6.040803.140111 ↗
59. Heart valves from polyester fibers vs. biological tissue: comparative study in vitro10.1007/s10439-016-1632-4 ↗