journal article·2021-11

The different failure modes of the connecting elements of the modular hip arthroplasty revision stem Revitan

Christopher Butler Ransohoff, Ronald Wanner, Theo Solinger, Emanuel Gautier, Henk Eijer, Peter Wahl

Publisher page

Abstract

Introduction Fracture of the femoral stem is the cause of ~1 % of revisions after total hip arthroplasty. The risk increases intrinsically with modularity, whereby modularity is particularly useful in revision arthroplasty. We present 7 cases of failure of a specific modular, tapered, fluted, titanium alloy Revitan stem and analyse the different failure modes. Methods Retrospective review of all Revitan stems revised at our institutions due to implant failure and analysis of clinical presentation, diagnostic workup, and failure mode. The retrieved components were analyzed by optical and scanning electron microscopy. Results A total of 7 cases were included. There was a significant time lag between symptom onset and correct diagnosis. Conventional radiographs and low-dose CT scans (CT scout imaging) were decisive for diagnosis. All failures occurred at the level of the connection between the proximal component and the distal part of the stem. Three different failure modes were identified: loosening of the proximal component, fatigue fracture of the connection pin, and distal loosening of the connection pin. No alterations of the microstructure or deviation from manufacturing specifications regarding dimensions were observed. Failure was caused by mechanical overload. Conclusion Conventional radiographs are the mainstay in identifying failed modular stems. Repeated radiographs and low-dose CT scans may be helpful additions. No single modification of the connection will address all possible failure modes. Modularity of revision stems offers advantages up until insertion of the definitive stem. Monoblock definitive stems might overcome the potential mechanical weaknesses of modularity and should be considered in relatively young, heavy and active patients.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Hip and Femur Fractures · Total Knee Arthroplasty Outcomes

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (50)

Show source-provided references
1. Revision total hip arthroplasty after removal of a fractured well-fixed extensively porous-coated femoral component using a trephine10.1302/0301-620X.97B9.35037 ↗
2. Australian Orthopaedic Association · 2017
3. Removal of the well-bonded distal part of a non-cylindrical broken femoral stem (Autophor 900S) with hollow trephine reamers--report of two cases10.1080/00016470216321 ↗
4. High rate of fracture in the cementless modular Extreme (Mark I) femoral prosthesis in revision total hip arthroplasty: 33 cases at more than 5 years' follow-up10.1016/j.otsr.2013.08.007 ↗
5. Standardized loads acting in hip implants10.1371/journal.pone.0155612 ↗
6. Femoral revision: distal fixation with fluted, tapered grit-blasted stems10.1054/arth.2002.32463 ↗
7. Femoral revision with the Wagner SL revision stem : evaluation of one hundred and twenty-nine revisions followed for a mean of 4.8 years10.2106/00004623-200107000-00007 ↗
8. Fractures of distally-fixed femoral stems after revision arthroplasty10.1302/0301-620X.87B10.16528 ↗
9. Fracture of the femoral prosthesis in total hip replacement according to Charnley10.3109/17453677708994812 ↗
10. Fracture of the femoral component after total hip replacement. An analysis of fifty-eight cases10.2106/00004623-198163070-00005 ↗
11. Fracture of femoral prostheses in total hip replacement. A clinical study10.1097/00003086-197509000-00014 ↗
12. Exchangeable femoral neck (Dual-Modular) THA prostheses have poorer Survivorship than other designs: a nationwide cohort of 324,108 patients10.1007/s11999-017-5260-6 ↗
13. Corrosion between the components of modular femoral hip prostheses10.1302/0301-620X.74B4.1624507 ↗
14. The role of proximal femoral support in stress development within hip prostheses10.1097/00003086-200403000-00024 ↗
15. Assessment of a press-fit proximal femoral modular reconstruction implant (PFMR((R))) at 14.5 years. A 48-case series with a disturbing rate of implant fracture10.1016/j.otsr.2017.11.024 ↗
16. Mid term results with the curved modular tapered, fluted titanium Revitan stem in revision hip replacement10.1302/0301-620X.96B7.33280 ↗
17. Improved outcome in femoral revision arthroplasty with tapered fluted modular titanium stems10.1097/01.blo.0000238875.86519.cf ↗
18. Letter regarding "Short proximal components in modular revision stems carry a higher risk for stem fractures" by Falko Herold, Hubert Nötzli, and Henk Eijer10.1177/1120700020931026 ↗
19. Ti based biomaterials, the ultimate choice for orthopaedic implants–a review10.1016/j.pmatsci.2008.06.004 ↗
20. In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations. The effect of crevice, stress, motion, and alloy coupling10.1002/jbm.820271210 ↗
21. Radiographic bone regeneration and clinical outcome with the Wagner SL revision stem: a 5-year to 12-year follow-up study10.1016/j.arth.2006.04.029 ↗
22. Fracture of a femoral revision stem following a technical failure
23. Short proximal components in modular revision stems carry a higher risk for stem fractures10.1177/1120700019884049 ↗
24. Which length should the neck segment of modular revision stems have?
25. The rationale of cementless revision of cemented arthroplasty failures
26. Utility of trephine reamers in revision hip arthroplasty10.1016/j.arth.2013.04.009 ↗
27. Modular tapered titanium stems in revision arthroplasty of the hip: the Risk and Causes of Stem Fracture10.1302/0301-620X.98B1.36442 ↗
28. Simplified extraction of broken femoral stems. A brief technical note10.1016/S0883-5403(89)80073-7 ↗
29. Mechanical failure of 113 uncemented modular revision femoral components10.1302/0301-620X.102B5.BJJ-2019-1333.R2 ↗
30. Factors influencing taper failure of modular revision hip stems10.1016/j.medengphy.2018.02.001 ↗
31. Modified extended trochanteric osteotomy for the treatment of vancouver B2/B3 periprosthetic fractures of the femur10.1016/j.arth.2017.02.079 ↗
32. Removal of well-fixed fixed femoral stems10.1016/j.otsr.2015.06.029 ↗
33. Outcome of revision hip arthroplasty done without cement10.2106/00004623-199407000-00002 ↗
34. Femoral revision with a fluted, tapered, modular stem seventy patients followed for a mean of 3.9 years10.1016/j.arth.2005.08.022 ↗
35. Distal locking stem for revision femoral loosening and peri-prosthetic fractures10.1007/s00264-010-1182-6 ↗
36. The window technique for the removal of broken femoral stems in total hip replacement
37. Mechanical properties of biomedical titanium alloys10.1016/S0921-5093(97)00806-X ↗
38. Fatigue fracture in dual modular revision total hip arthroplasty stems: failure analysis and computed tomography diagnostics in two cases10.1016/j.arth.2013.09.008 ↗
39. Modular femoral stem-sleeve junction failure after primary total hip arthroplasty10.1016/j.arth.2008.09.006 ↗
40. Registry · 2020
41. Femoral revision hip arthroplasty: a comparison of two stem designs10.1007/s11999-009-1145-7 ↗
42. Understanding the surface features of fatigue fractures: how they describe the failure cause and the failure history10.1361/15477020522924 ↗
43. Quantitative assessment of walking activity after total hip or knee replacement10.2106/00004623-199801000-00010 ↗
44. Revision total hip arthroplasty: the limits of fully coated stems
45. High survival of modular tapered stems for proximal femoral bone defects at 5 to 10 years followup10.1007/s11999-012-2552-8 ↗
46. An in vitro comparison of the primary stability of 2 tapered fluted femoral stem designs10.1016/j.arth.2015.09.036 ↗
47. [Revision prosthesis for the hip joint in severe bone loss]
48. The transfemoral approach for revision of total hip replacement10.1007/BF03180945 ↗
49. Removal of an osteointegrated broken uncemented femoral stem after hip arthroplasty-technical note10.1186/s13018-021-02365-x ↗
50. Fracture of a modular femoral neck after total hip arthroplasty: a case report10.2106/JBJS.I.01033 ↗