Dynamic finite element analysis of hip replacement edge loading: Balancing precision and run time in a challenging model
Source abstract
Abstract
An important aspect in evaluating the resilience of hip replacement designs is testing their performance under adverse conditions that cause edge loading of the acetabular liner. The representation of edge loading conditions in finite element models is computationally challenging due to the changing contact locations, need for fine meshes, and dynamic nature of the system. In this study, a combined mesh and mass-scaling sensitivity study was performed to identify an appropriate compromise between convergence and solution time of explicit finite element analysis in investigating edge loading in hip replacement devices. The optimised model was then used to conduct a sensitivity test investigating the effect of different hip simulator features (the mass of the translating fixture and mediolateral spring damping) on the plastic strain in the acetabular liner. Finally, the effect of multiple loading cycles on the progressive accumulation of plastic strain was then also examined using the optimised model. A modelling approach was developed which provides an effective compromise between mass-scaling effects and mesh refinement for a solution time per cycle of less than 1 h. This 'Recommended Mesh' model underestimated the plastic strains by less than 10%, compared to a 'Best Estimate' model with a run time of ∼190 h. Starting with this model setup would therefore significantly reduce any new model development time while also allowing the flexibility to incorporate additional complexities as required. The polyethylene liner plastic strain was found to be sensitive to the simulator mass and damping (doubling the mass or damping had a similar magnitude effect to doubling the swing phase load) and these should ideally be described in future experimental studies. The majority of the plastic strain (99%) accumulated within the first three load cycles.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Hip disorders and treatments · Total Knee Arthroplasty Outcomes
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (32)
Show source-provided references
1. Notch fatigue of ultrahigh molecular weight polyethylene (UHMWPE) used in total joint replacements10.1016/j.jmbbm.2016.02.014 ↗
2. 2018
3. In vivo assessment of total hip femoral head separation from the acetabular cup during 4 common daily activities10.3928/01477447-20110427-06 ↗
4. Friction of total hip replacements with different bearings and loading conditions10.1002/jbm.b.30691 ↗
5. In vivo determination and comparison of total hip arthroplasty kinematics for normal, preoperative degenerative, and postoperative implanted hips10.1016/j.arth.2019.08.057 ↗
6. In vivo” determination of hip joint separation and the forces generated due to impact loading conditions10.1016/S0021-9290(00)00239-6 ↗
7. Computationally efficient modelling of hip replacement separation due to small mismatches in component centres of rotation
8. Etchels · 2023
9. In vivo comparison of hip mechanics for minimally invasive versus traditional total hip arthroplasty10.1016/j.clinbiomech.2007.09.015 ↗
10. The creep and wear of highly cross-linked polyethylene: a three-year randomised, controlled trial using radiostereometric analysis10.1302/0301-620X.90B5.20545 ↗
11. Ceramic liner fracture after cementless alumina-on-alumina total hip arthroplasty10.1097/BLO.0b013e3180303e87 ↗
12. Contact mechanics of modular metal-on-polyethylene total hip replacement under adverse edge loading conditions10.1016/j.jbiomech.2014.08.015 ↗
13. Critical damping
14. 2014
15. Jahani · 2017
16. Importance of dynamics in the finite element prediction of plastic damage of polyethylene acetabular liners under edge loading conditions10.1016/j.medengphy.2021.07.010 ↗
17. Capsular ligaments provide a passive stabilizing force to protect the hip against edge loading10.1302/2046-3758.109.BJR-2020-0536.R1 ↗
18. In vivo comparison of hip separation after metal-on-metal or metal-on-polyethylene total hip arthroplasty10.2106/00004623-200210000-00015 ↗
19. Effects of mass scaling on finite element simulation of the cold roll-beating forming process
20. An in vivo determination of total hip arthroplasty pistoning during activity10.1054/arth.2000.6637 ↗
21. Material properties of ultra-high molecular weight polyethylene: comparison of tension, compression, nanomechanics and microstructure across clinical formulations10.1016/j.jmbbm.2018.03.029 ↗
22. A novel method to measure rim deformation in UHMWPE acetabular liners10.1016/j.medengphy.2018.04.023 ↗
23. Quantifying cross-shear under translation, rolling, and rotation, and its effect on UHMWPE wear’, Wear
24. Influence of polyethylene creep behavior on wear in total hip arthroplasty10.1002/jor.20042 ↗
25. In vivo wear and migration of highly cross-linked polyethylene cups a radiostereometry analysis study10.1016/j.arth.2004.09.040 ↗
26. Dynamic femoral head translations in dysplastic hips10.1016/j.clinbiomech.2017.05.003 ↗
27. 2019
28. Application of viscoelastic fracture model and non-uniform crack initiation at clinically relevant notches in crosslinked UHMWPE10.1016/j.jmbbm.2012.07.008 ↗
29. Wear of large diameter ceramic-on-ceramic hip bearings under standard and microseparation conditions', Biotribology
30. Finite element analysis of polyethylene wear in total hip replacement : a literature review10.1177/0954411919872630 ↗
31. Wear and deformation of ceramic-on-polyethylene total hip replacements with joint laxity and swing phase microseparation10.1243/09544110360579367 ↗
32. UHMWPE acetabular cup creep deformation during the run-in phase of THA's life cycle10.1016/j.jmbbm.2018.07.015 ↗