Evaluation of a vibration modeling technique for the in-vitro measurement of dental implant stability

Chester Jar, Andrew Archibald, Monica Gibson, Lindsey Westover

Abstract

The Advanced System for Implant Stability Testing (ASIST) is a device currently being developed to noninvasively measure implant stability by estimating the mechanical stiffness of the bone-implant interface, which is reported as the ASIST Stability Coefficient (ASC). This study's purpose was to determine whether changes in density, bonding, and drilling technique affect the measured vibration of a dental implant, and whether they can be quantified as a change in the estimated BII stiffness. Stability was also measured using RFA, insertion torque (IT) and the pullout test. Bone-level tapered implants (4.1 mm diameter, 10 mm length) were inserted in polyurethane foam as an artificial bone substitute. Samples were prepared using different bone densities (20, 30, 40 PCF), drilling sequences, and superglue to simulate a bonded implant. Measurements were compared across groups at a significance level of 0.05. The ASC was able to indicate changes in each factor as a change in the interfacial stiffness. IT and pullout force values also showed comparable increases. Furthermore, the relative difference in ISQ values between experimental groups was considerably smaller than the ASC. While future work should be done using biological bone and in-vivo systems, the results of this in-vitro study suggest that modelling of the implant system with a vibration-based approach may provide a noninvasive method of assessing the mechanical stability of the implant.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Dental Implant Techniques and Outcomes · Orthopaedic implants and arthroplasty · Bone Tissue Engineering Materials

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (44)

Show source-provided references
1. 2021
2. 2023
3. Methods used to assess implant stability: current status
4. Quantifying the influence of bone density and thickness on resonance frequency analysis: an in vitro study of biomechanical test materials
5. Evaluation of the correlation between insertion torque and primary stability of dental implants using a block bone test10.5051/jpis.2013.43.1.30 ↗
6. Influence of clinical parameters on the primary stability of a tapered dental implant: a retrospective analysis10.11607/jomi.8682 ↗
7. Bone density and its importance in orthodontics10.1016/j.jobcr.2013.01.001 ↗
8. Primary stability determination by means of insertion torque and RFA in a sample of 4,135 implants10.1111/j.1708-8208.2010.00302.x ↗
9. Under-drilling versus hybrid osseodensification technique: differences in implant primary stability and bone density of the implant bed walls10.3390/ma13020390 ↗
10. The effect of undersizing and tapping on bone to implant contact and implant primary stability: a histomorphometric study on bovine ribs10.4047/jap.2018.10.3.227 ↗
11. Influence of implant shape, surface morphology, surgical technique and bone quality on the primary stability of dental implants10.1016/j.jmbbm.2012.10.010 ↗
12. G*Power 3: a flexible statistical power analysis program for the social, behavioral, and biomedical sciences10.3758/BF03193146 ↗
13. Non-linear 3D finite element analysis of full-arch implant-supported fixed dentures10.1016/j.msec.2014.02.021 ↗
14. Effects of the healing chambers in implant macrogeometry design in a low-density bone using conventional and undersized drilling10.4103/jispcd.JISPCD_96_21 ↗
15. Influence of the implant abutment on the periotest value: an in vivo study
16. Factors influencing implant stability10.4103/jdi.jdi_14_18 ↗
17. Novel design of additive manufactured hollow porous implants10.1016/j.dental.2020.08.011 ↗
18. An analytical model to measure dental implant stability with the Advanced System for Implant Stability Testing (ASIST)10.1016/j.jmbbm.2023.106238 ↗
19. Simulation of impact test for determining “health” of percutaneous bone anchored implants10.1115/1.2241685 ↗
20. Does the manual insertion torque of smartpegs affect the outcome of implant stability quotients (ISQ) during resonance frequency analysis (RFA)?10.1186/s40729-019-0195-1 ↗
21. Challenges of using resonance frequency analysis to identify stability of a dental implant placed in the mandible10.11607/jomi.8579 ↗
22. Stimulating effect of implant loading on surrounding bone10.1111/j.1600-0501.2004.01000.x ↗
23. The application of resonance frequency measurements to study the stability of titanium implants during healing in the rabbit tibia10.1034/j.1600-0501.1997.080310.x ↗
24. Analysis of primary stability of dental implants inserted in different substrates using the pullout test and insertion torque10.1155/2013/194987 ↗
25. Effect of implant design and bone density in primary stability10.1590/S1677-32252013000300002 ↗
26. Primary stability of a conical implant and a hybrid, cylindric screw-type implant in vitro
27. Implant stability measurements using resonance frequency analysis: biological and biomechanical aspects and clinical implications10.1111/j.1600-0757.2008.00267.x ↗
28. 2017
29. The effects of bone density and crestal cortical bone thickness on micromotion and peri-implant bone strain distribution in an immediately loaded implant: a nonlinear finite element analysis10.5051/jpis.2016.46.3.152 ↗
30. A dynamic analytical model for impact evaluation of percutaneous implants10.1115/1.2970061 ↗
31. Influence of the surgical technique and surface roughness on the primary stability of an implant in artificial bone with a density equivalent to maxillary bone: a laboratory study10.1111/j.1600-0501.2008.01692.x ↗
32. Implant micromotion is related to peak insertion torque and bone density10.1111/j.1600-0501.2008.01679.x ↗
33. An in vitro evaluation on polyurethane foam sheets of the insertion torque, removal torque values, and resonance frequency analysis (RFA) of a self-tapping threads and round apex implant10.1177/0262489320971796 ↗
34. Influence of simulated bone-implant contact and implant diameter on secondary stability: a resonance frequency in vitro study10.1111/clr.12195 ↗
35. A quantitative approach for the bone-implant osseointegration assessment based on ultrasonic elastic guided waves10.3390/s19030454 ↗
36. Investigation to predict primary implant stability using frictional resistance torque of tap drilling10.5037/jomr.2022.13401 ↗
37. Identifiable range of osseointegration of dental implants through resonance frequency analysis10.1016/j.medengphy.2010.07.015 ↗
38. Advanced system for implant stability testing (ASIST)10.1016/j.jbiomech.2016.09.043 ↗
39. Comparison of implant stability measurement devices for bone-anchored hearing aid systems10.1016/j.prosdent.2017.02.021 ↗
40. A vibratory, subresonant diagnostic device to measure dental implant stability via angular stiffness
41. Analytical model for dental implant insertion torque10.1016/j.jmbbm.2022.105223 ↗
42. A three-dimensional finite element study on the biomechanical behaviour of an FGBM dental implant in surrounding bone10.1016/j.jbiomech.2006.11.019 ↗
43. Modal analysis for implant stability assessment: sensitivity of this methodology for different implant designs10.1016/j.dental.2018.05.016 ↗
44. A noncontact detection technique for interfacial bone defects and osseointegration assessment surrounding dental implants10.1016/j.measurement.2014.05.039 ↗