journal article·2024-09Open access

Unveiling the fatigue life of NiTi endodontic files: An integrated computational–experimental study

Shravan Kumar Subramanian, Vivek Joshi, Sahil Kalra, Sondipon Adhikari

Abstract

Nickel-titanium (NiTi) rotary files used in root canal treatments experience fatigue and shear damage due to the complex curved geometries and operating conditions encountered within the root canal. This can lead to premature file fracture, causing severe complications. A comprehensive understanding of how different factors contribute to file damage is crucial for improving their functional life. This study investigates the combined effects of root canal curvature radius, file canal curvature, and rotational speed on the fatigue life and failure modes of NiTi endodontic files through an integrated computational and experimental approach. Advanced finite element simulations precisely replicating the dynamic motion of files inside curved canal geometries were conducted. Critical stress/strain values were extracted and incorporated into empirical fatigue models to predict the functional life of endodontic files. Extensive experiments with files rotated inside artificial curved canals at various canal curvatures and speeds provided validation. Increasing the canal curvature beyond 60∘ and shorter curvature radii below 5 mm dramatically reduced the functional life of the endodontic file, especially at rotational speeds over 360 rpm. The Coffin-Manson fatigue model based on strain amplitude showed the closest agreement with experiments. Shear stresses dominated damage at low canal curvatures, while the combined shear-fatigue loading effects were prominent at higher canal curvatures. This conclusive study elucidates how operational parameters like canal curvature radii, canal curvature, and rotational speed synergistically influence the fatigue damage processes in NiTi files. The findings offer valuable guidelines to optimize these factors, significantly extending the functional life of endodontic files and reducing the risk of intra-operative failures. The validated computational approach provides a powerful tool for virtual testing and estimation of the functional life of the new file designs before manufacturing.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Endodontics and Root Canal Treatments · Occupational health in dentistry · Dental materials and restorations

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (58)

Show source-provided references
1. Numerical and analytical studies of low cycle fatigue behavior of 316 LN austenitic stainless steel10.1115/1.4045897 ↗
2. Kinematic effects of nickel–titanium instruments with reciprocating or continuous rotation motion: A systematic review of in vitro studies10.1016/j.joen.2016.04.002 ↗
3. Impact of thermomechanical texture on the superelastic response of nitinol implants10.1016/j.jmbbm.2011.05.013 ↗
4. Cyclic fatigue of different reciprocating endodontic instruments using matching artificial root canals at body temperature in vitro10.3390/ma17040827 ↗
5. Influence of heat treatment of nickel–titanium instruments on cyclic fatigue resistance in simulated curved canals10.1055/s-0042-1747952 ↗
6. Advancing nitinol: From heat treatment to surface functionalization for nickel–titanium (NiTi) instruments in endodontics10.1016/j.bioactmat.2022.09.008 ↗
7. Fatigue testing of a NiTi rotary instrument. Part 1: Strain–life relationship10.1111/j.1365-2591.2007.01262.x ↗
8. A numerical method for predicting the bending fatigue life of NiTi and stainless steel root canal instruments10.1111/j.1365-2591.2010.01838.x ↗
9. An improved model of 3-dimensional finite element analysis of mechanical behavior of endodontic instruments10.1016/j.tripleo.2009.10.039 ↗
10. Improvement of the fatigue resistance of NiTi endodontic files by surface and bulk modifications10.1111/j.1365-2591.2010.01759.x ↗
11. Low cycle fatigue lifetime prediction of superplastic shape memory alloy structures: Application to endodontic instruments10.1016/j.jmbbm.2023.106122 ↗
12. Comparison of the mechanical behavior between controlled memory and superelastic nickel–titanium files via finite element analysis10.1016/j.joen.2013.07.030 ↗
13. The relation between impact strength and flexural strength of dental materials10.1016/j.jmbbm.2021.104658 ↗
14. Cyclic fatigue resistance of ProTaper next nickel–titanium rotary files10.1111/iej.12244 ↗
15. Cumulative fatigue damage and life prediction theories: A survey of the state of the art for homogeneous materials10.1016/S0142-1123(97)00081-9 ↗
16. Movement kinematics and cyclic fatigue of NiTi rotary instruments: A systematic review10.1111/iej.12613 ↗
17. Low-cycle fatigue life of superelastic NiTi wires10.1016/j.ijfatigue.2008.03.014 ↗
18. Evaluation of stress distribution in nickel–titanium rotary instruments with different geometrical designs subjected to bending and torsional load: A finite element study10.1186/s42269-020-00377-x ↗
19. Cyclic fatigue analysis of twisted file rotary NiTi instruments used in reciprocating motion10.1111/j.1365-2591.2012.02036.x ↗
20. Mechanical behavior of ProTaper universal F2 finishing file under various curvature conditions: A finite element analysis study10.1016/j.joen.2011.06.003 ↗
21. Resistance to flexural fatigue of reciproc R25 files under continuous rotation and reciprocating movement10.1016/j.joen.2011.12.033 ↗
22. Mechanical properties and metallurgical features of new green niti reciprocating instruments10.3390/ma13173736 ↗
23. Alteration in the inherent metallic and surface properties of nickel–titanium root canal instruments to enhance performance, durability and safety: A focused review10.1111/j.1365-2591.2011.01957.x ↗
24. Stress generation during pecking motion of rotary nickel–titanium instruments with different pecking depth10.1016/j.joen.2017.04.013 ↗
25. An analytical model to measure dental implant stability with the advanced system for implant stability testing (ASIST)10.1016/j.jmbbm.2023.106238 ↗
26. Opto-magnetic imaging spectroscopy in analyzing rotary NiTi endodontic instruments10.1016/j.jmbbm.2023.105789 ↗
27. Design of shape memory alloy actuated intelligent parabolic antenna for space applications10.1088/1361-665X/aa7468 ↗
28. Development of shape memory alloy actuator integrated flexible poly-ether-ether-ketone antenna with simultaneous beam steering and shaping ability10.1177/1045389X18798946 ↗
29. Assessment of dentinal damage during canal preparation using reciprocating and rotary files10.1016/j.joen.2014.02.015 ↗
30. Evaluation of the cyclic fatigue and torsional resistance of novel nickel–titanium rotary files with various alloy properties10.1016/j.joen.2016.07.015 ↗
31. Factors influencing the dimensional accuracy of additively manufactured dental models: A systematic review of in vitro studies10.1016/j.jmbbm.2023.106057 ↗
32. Comparison of forces generated during root canal shaping and residual stresses of three nickel–titanium rotary files by using a three-dimensional finite-element analysis10.1016/j.joen.2008.03.015 ↗
33. The role of pecking motion depths in dynamic cyclic fatigue resistance: In vitro study10.1016/j.identj.2024.01.014 ↗
34. Influence of surrounding temperature and angle of file access on cyclic fatigue resistance of two single file nickel–titanium instruments10.1111/aej.12473 ↗
35. Correlation between experimental cyclic fatigue resistance and numerical stress analysis for nickel–titanium rotary files10.1016/j.joen.2011.03.025 ↗
36. Influence of kinematics on the cyclic fatigue resistance of replicalike and original brand rotary instruments10.1016/j.joen.2020.05.001 ↗
37. A study on the influence of Ni–Ti M-wire in the flexural fatigue life of endodontic rotary files by using finite element analysis10.1016/j.msec.2014.03.061 ↗
38. Shot peening increases resistance to cyclic fatigue fracture of endodontic files10.1038/s41598-021-92382-x ↗
39. Rotary NiTi instrument fracture and its consequences10.1016/j.joen.2006.06.008 ↗
40. Impact of pecking amplitude on cyclic fatigue of new nickel–titanium files10.1002/cre2.811 ↗
41. Defects in ProTaper S1 instruments after clinical use: Longitudinal examination10.1111/j.1365-2591.2005.00991.x ↗
42. Mechanical properties of a novel nickel–titanium root canal instrument: Stationary and dynamic tests10.1016/j.joen.2020.03.016 ↗
43. Finite element analysis comparing WaveOne, WaveOne gold, reciproc and reciproc blue responses with bending and torsion tests10.1016/j.jmbbm.2018.10.016 ↗
44. Cyclic torsional loading of pseudoelastic NiTi shape memory alloys: Damping and fatigue failure10.1016/j.msea.2005.10.037 ↗
45. Rotation or reciprocation: A contemporary look at NiTi instruments10.1038/sj.bdj.2012.268 ↗
46. Cyclic fatigue testing of nickel–titanium endodontic instruments10.1016/S0099-2399(97)80250-6 ↗
47. Fatigue analysis of NiTi rotary endodontic files through finite element simulation: Effect of root canal geometry on fatigue life10.3390/jcm10235692 ↗
48. Effects of R-phase on mechanical responses of a nickel–titanium endodontic instrument: Structural characterization and finite element analysis
49. Defects in rotary nickel–titanium files after clinical use10.1097/00004770-200003000-00008 ↗
50. Failure mechanism of ProTaper Ni–Ti rotary instruments during clinical use: Fractographic analysis10.1111/j.1365-2591.2006.01065.x ↗
51. Influence of environmental temperature, heat-treatment and design on the cyclic fatigue resistance of three generations of a single-file nickel–titanium rotary instrument10.1007/s10266-018-0399-5 ↗
52. The effect of experimental conditions in root dentin microcracks detection by micro-computed tomography10.1016/j.jmbbm.2022.105108 ↗
53. Effect of different axial speed patterns on cyclic fatigue resistance of rotary nickel–titanium instruments10.1186/s12903-022-02639-8 ↗
54. Cyclic fatigue resistance of RaCe and mtwo rotary files in continuous rotation and reciprocating motion10.1016/j.joen.2013.12.010 ↗
55. Influence of the number of pecking motions at working length on the shaping ability of single-file systems in long oval-shaped curved canals10.1016/j.joen.2022.01.009 ↗
56. Comparative research on the accumulative damage rules under multiaxial block loading spectrum for 2024-T4 aluminum alloy10.1016/j.ijfatigue.2012.11.004 ↗
57. An overview of the mechanical properties of nickel–titanium endodontic instruments10.1111/etp.12045 ↗
58. Influence of the pecking motion frequency on the cyclic fatigue resistance of endodontic rotary files10.3390/jcm9010045 ↗