Effect of head size, head material, and radiation dose on the repeatability of CT-RSA measurements of femoral head penetration

Jennifer S. Polus, Bart L. Kaptein, Brent A. Lanting, Matthew G. Teeter

Abstract

BACKGROUND: The risk of early revision of total hip arthroplasty (THA) for polyethylene wear is now low, but there remains a need to perform wear measurements in patients for clinical surveillance. The gold standard of wear measurements has been radiostereometric analysis (RSA), which has limited availability. The use of computed tomography (CT) to perform THA wear measurement was described a decade ago and found to have acceptable accuracy and precision, but high radiation dose was a concern. Additionally, the use of larger femoral head sizes and ceramic femoral heads has risen in the past decade. The objectives of the study were to determine the effect of femoral head size, femoral head material, and lowered radiation dose on femoral head penetration measurement repeatability. METHODS: A cadaveric hip was implanted with a cementless THA implant system. CT scans were acquired at a conventional radiation dose and at a reduced dose and repeated for a 32 mm and 36 mm cobalt-chromium femoral heads and a 32 mm ceramic femoral head. Apparent translation of the femoral head versus the acetabular cup was measured between the repeated scans using a CT-RSA software, where deviations from zero indicated measurement precision. RESULTS: The mean and standard deviation of translations in all planes was <0.200 mm. There was no effect for 3D translation of increasing cobalt-chromium head size (p = 0.2252). Cobalt-chromium heads had superior repeatability compared to ceramic heads at reduced dose (p = 0.022), but not at conventional dose (p = 0.1265). Further, superior repeatability was achieved with the reduced dose scan for the cobalt-chromium head (p = 0.0058), however there was no difference between doses for the ceramic head (p = 0.8148). DISCUSSION: CT-based wear measurement repeatability is excellent and consistent with prior literature even when implementing a larger femoral head, a ceramic femoral head, or reducing radiation dose to 25% of a conventional clinical scan.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Bone and Joint Diseases

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (43)

Show source-provided references
1. 2008
2. The lawrence D. Dorr surgical techniques & technologies award: using big heads and small acetabular components with highly cross-linked polyethylene in total hip arthroplasty: is it safe?10.1016/j.arth.2021.01.067 ↗
3. The effect of conformity and plastic thickness on contact stresses in metal-backed plastic implants10.1115/1.3138543 ↗
4. Review of periprosthetic osteolysis in total joint arthroplasty: an emphasis on host factors and future directions10.1002/jor.21554 ↗
5. Iterative reconstruction methods in X-ray CT10.1016/j.ejmp.2012.01.003 ↗
6. The 2012 John Charnley Award: clinical multicenter studies of the wear performance of highly crosslinked remelted polyethylene in THA10.1007/s11999-012-2604-0 ↗
7. Low-dose CT-based implant motion analysis is a precise tool for early migration measurements of hip cups: a clinical study of 24 patients10.1080/17453674.2020.1725345 ↗
8. Precision of CT-based micromotion analysis is comparable to radiostereometry for early migration measurements in cemented acetabular cups10.1080/17453674.2021.1906082 ↗
9. Wear of highly crosslinked polyethylene acetabular components10.3109/17453674.2014.972890 ↗
10. Accuracy of methods to measure femoral head penetration within metal-backed acetabular components10.1002/jor.23356 ↗
11. Highly crosslinked polyethylene liners have negligible wear at 10 Years: a radiostereometric analysis study10.1097/CORR.0000000000002002 ↗
12. Wear, osteolysis, and aseptic loosening following total hip arthroplasty in young patients with highly cross-linked polyethylene: a review of studies with a follow-up of over 15 years10.3390/jcm12206615 ↗
13. Measurement of polyethylene wear in metal-backed acetabular cups. I. Three-dimensional technique10.1097/00003086-199510000-00033 ↗
14. The Otto Aufranc Award. Highly cross-linked polyethylene in total hip arthroplasty: randomized evaluation of penetration rate in cemented and uncemented sockets using radiostereometric analysis10.1097/01.blo.0000150314.70919.e3 ↗
15. Characterization of long-term femoral-head-penetration rates. Association with and prediction of osteolysis10.2106/00004623-200008000-00006 ↗
16. Have newer bearing surfaces changed expectations regarding the longevity of total hip arthroplasty?10.1302/0301-620X.102B7.BJJ-2019-1545.R1 ↗
17. Thin highly cross-linked polyethylene liners combined with large femoral heads in primary total hip arthroplasty show excellent survival and low wear rates at a mean follow-up of 12.8 years10.1302/0301-620X.105B1.BJJ-2022-0812.R1 ↗
18. A new technique for measuring wear in total hip arthroplasty using computed tomography10.1016/j.arth.2012.03.053 ↗
19. In vivo and ex vivo measurement of polyethylene wear in total hip arthroplasty: comparison of measurements using a CT algorithm, a coordinate-measuring machine, and a micrometer10.3109/17453674.2014.913225 ↗
20. Periprosthetic osteolysis: mechanisms, prevention and treatment10.3390/jcm8122091 ↗
21. The problem is osteolysis
22. elastix: a toolbox for intensity-based medical image registration10.1109/TMI.2009.2035616 ↗
23. A new automated way to measure polyethylene wear in THA using a high resolution CT scanner: method and analysis10.1155/2014/528407 ↗
24. Determination of polyethylene wear in total hip replacements with use of digital radiographs10.2106/00004623-199711000-00004 ↗
25. Radiographic methods for the assessment of polyethylene wear after total hip arthroplasty
26. Corrosion concerns? Trends in metal-on-polyethylene total hip arthroplasty revision rates and comparisons against ceramic-on-polyethylene up to 10 Years of follow-up10.1016/j.arth.2020.05.007 ↗
27. Excellent 20-year results of total hip arthroplasty with highly cross-linked polyethylene on cobalt-chromium femoral heads in patients ≤50 years10.1016/j.arth.2023.08.025 ↗
28. The variability of CT scan protocols for total hip arthroplasty: a call for harmonisation10.1530/EOR-22-0141 ↗
29. Sequentially irradiated and annealed highly cross-linked polyethylene: linear vector and volumetric wear in total hip arthroplasty at 10 years10.1016/j.artd.2021.08.002 ↗
30. “Great balls on fire:” known algorithm with a new instrument?10.1080/17453674.2020.1840029 ↗
31. A ten-year radiostereometric analysis of polyethylene wear between oxidized zirconium and cobalt chrome articulations in total hip arthroplasty10.1016/j.arth.2022.02.099 ↗
32. Inducible displacement CT increases the diagnostic accuracy of aseptic loosening in primary total hip arthroplasty10.2340/17453674.2022.5240 ↗
33. Computed tomography-based radiostereometric analysis in orthopedic research: practical guidelines10.2340/17453674.2023.15337 ↗
34. Low wear at 10-year follow-up of a second-generation highly cross-linked polyethylene in total hip arthroplasty10.1016/j.arth.2022.01.016 ↗
35. Superior accuracy of model-based radiostereometric analysis for measurement of polyethylene wear: a phantom study10.1302/2046-3758.18.2000041 ↗
36. Deep learning reconstruction allows low-dose imaging while maintaining image quality: comparison of deep learning reconstruction and hybrid iterative reconstruction in contrast-enhanced abdominal CT10.21037/qims-21-1216 ↗
37. To what extent could the acetabular liner thickness be reduced yet remaining tribologically acceptable in metal-on-vitamin E-diffused crosslinked polyethylene hip arthroplasty?10.1002/jbm.b.35078 ↗
38. Highly crosslinked polyethylene wear rates and acetabular component orientation: a minimum ten-year follow-up10.1302/0301-620X.100B7.BJJ-2017-1457.R3 ↗
39. The use of large metal heads in thin vitamin E-doped cross-linked polyethylene inserts does not increase polyethylene wear in total hip arthroplasty: 5-year results from a randomized controlled trial10.1016/j.arth.2024.02.008 ↗
40. Prosthetic liner wear in total hip replacement: a longitudinal 13-year study with computed tomography10.1007/s00256-018-2878-8 ↗
41. Long-Term wear-related complications of cross-linked versus conventional polyethylene after total hip arthroplasty: a meta-analysis10.1016/j.arth.2022.05.013 ↗
42. Mean 16-year results of total hip arthroplasty with alumina ceramic femoral heads on highly cross-linked polyethylene in patients 50 Years or less10.1016/j.arth.2023.04.041 ↗
43. Radiostereometric analysis using clinical radiographic views: validation measuring total hip replacement wear10.1002/jor.23170 ↗