The impact of conical design of dental implant and osteotomy shape on the insertion process and primary stability
Source abstract
Abstract
This study investigated the impact of dental implant shape (conical vs. cylindrical) and pilot-hole geometry (conical vs. cylindrical) on primary stability. Finite element analysis was employed to simulate the insertion process and push-in test. Three implantation scenarios were examined: a cylindrical implant in a cylindrical pilot-hole (cy-cy model), a conical implant in a cylindrical pilot-hole (co-cy model), and a conical implant in a conical pilot-hole (co-co model). These configurations were evaluated based on the maximum insertion torque, insertion energy, stiffness, and holding power of the implant-bone construct. Given peri-implant damage during the insertion process and push-in test, elastic-plastic-damage properties were used to model bone material behavior. Results showed that despite lower implant-bone engagement in the co-cy scenario, its maximum insertion torque was 81 % higher than the cy-cy model. However, the stiffness and the holding power were 31 % and 44 % lower, respectively. Additionally, the co-co scenario demonstrated a significant increase in maximum insertion torque compared to other models while maintaining stiffness and holding power comparable to the cy-cy configuration, making it the most favorable model. This study also highlighted the importance of insertion energy as a potential indicator of implant-bone construct engagement, suggesting its consideration alongside other traditional stability factors. Further research, including in-vitro experiments, is necessary to refine pilot-hole geometry based on implant body design, ultimately enhancing primary stability and improving clinical outcomes in dental implantation.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Dental Implant Techniques and Outcomes · Orthodontics and Dentofacial Orthopedics · Dental Radiography and Imaging
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (58)
Show source-provided references
1. Investigation on primary stability of dental implants through considering peri-implant bone damage, caused by small and large deformations: a validated non-linear micro finite element study10.1016/j.jmbbm.2023.106062 ↗
2. Three-dimensional finite element investigation into effects of implant thread design and loading rate on stress distribution in dental implants and anisotropic bone10.3390/ma14226974 ↗
3. A biomechanical assessment of the relation between the oral implant stability at insertion and subjective bone quality assessment10.1111/j.1600-051X.2007.01047.x ↗
4. Influence of dental implant diameter and bone quality on the biomechanics of single-crown restoration. A finite element analysis10.3390/dj9090103 ↗
5. Improving dental implant stability by optimizing thread design: simultaneous application of finite element method and data mining approach10.1016/j.prosdent.2023.06.034 ↗
6. Methods used to assess implant stability: current status
7. Technique for placement of oxidized titanium implants in compromised maxillary bone: prospective study of 290 implants in 126 consecutive patients followed for a minimum of 3 years after loading
8. Effect of a modified stepped osteotomy on the primary stability of dental implants in low-density bone: a cadaver study10.11607/jomi.3720 ↗
9. Biomaterials and biomechanics of oral and maxillofacial implants: current status and future developments
10. Multiscale analyses of the bone-implant interface10.1177/0022034514566029 ↗
11. In vitro biomechanical and histological assessment of pilot hole diameter for positive-profile external skeletal fixation pins in Canine Tibiae10.1111/j.1532-950X.1996.tb01443.x ↗
12. Biomechanical evaluation of undersized drilling on implant biomechanical stability at early implantation times10.1016/j.joms.2012.10.008 ↗
13. Influence of stepped osteotomy on primary stability of implants inserted in low-density bone sites: an In vitro Study10.11607/jomi.4852 ↗
14. Influence of underpreparation on primary stability of implants inserted in poor quality bone sites: an in vitro study10.1016/j.joms.2015.01.029 ↗
15. Development of a new implant primary stability parameter: insertion torque revisited10.1111/j.1708-8208.2011.00392.x ↗
16. Modeling dental implant insertion10.1016/j.jmbbm.2017.01.021 ↗
17. Dental implants with immediate loading using insertion torque of 30 ncm: a systematic review10.1097/ID.0000000000000444 ↗
18. Effects of different implant osteotomy preparation sizes on implant stability and bone response in the Minipig mandible
19. Effect of surgical techniques on primary implant stability and peri-implant bone10.1016/j.joms.2007.04.017 ↗
20. Biomechanical and histological evaluation of four different implant macrogeometries in the early osseointegration process: an in vivo animal study10.1016/j.jmbbm.2021.104935 ↗
21. Effects of the technique and drill design used during the osteotomy on the thermal and histological stimulation10.1038/s41598-020-77762-z ↗
22. The Frialit-2 implant System: Five-Year clinical experience in single-tooth and immediately postextraction applications
23. Evaluation of insertion energy as novel parameter for dental implant stability10.3390/jcm9092977 ↗
24. Dynamic modelling and simulation of dental implant insertion process—A finite element study10.1016/j.finel.2011.03.005 ↗
25. Effects of biomechanical properties of the bone-implant interface on dental implant stability: from in silico approaches to the patient's mouth10.1146/annurev-bioeng-071813-104854 ↗
26. The effect of static bone strain on implant stability and bone remodeling10.1016/j.bone.2011.07.003 ↗
27. The influence of bone volume fraction and ash fraction on bone strength and modulus10.1016/S8756-3282(01)00467-7 ↗
28. Increase of pullout strength of spinal pedicle screws with conical core: biomechanical tests and finite element analyses10.1016/j.orthres.2004.11.002 ↗
29. The role of primary stability for successful immediate loading of dental implants. A literature review10.1016/j.jdent.2010.05.013 ↗
30. Measurement of bone damage caused by quasi-static compressive loading-unloading to explore dental implants stability: simultaneous use of in-vitro tests, μ-CT images, and digital volume correlation10.1016/j.jmbbm.2022.105566 ↗
31. Comparison of stability between cylindrical and conical type Mini-Implants: mechanical and histologic properties10.2319/0003-3219(2008)078[0692:COSBCA]2.0.CO;2 ↗
32. Development of predictable stability test for assessment of optimum loading time in dental implant10.4047/jkap.2008.46.6.628 ↗
33. Microstructure-based numerical computational method for the insertion torque of dental implant10.1016/j.jmbbm.2019.06.012 ↗
34. The effect of osteotomy dimension on implant insertion torque, healing mode, and osseointegration indicators: a study in dogs10.1097/ID.0000000000000476 ↗
35. Dental implants' stability dependence on rotational speed and feed-rate of drilling: In-vivo and ex-vivo investigations10.1016/j.jbiomech.2021.110696 ↗
36. Optimization of thread configuration in dental implants through regulating the mechanical stimuli in neighboring bone10.1016/j.cmpb.2023.107376 ↗
37. Prediction of insertion torque and stiffness of a dental implant in bovine trabecular bone using explicit micro-finite element analysis10.1016/j.jmbbm.2019.06.024 ↗
38. A nonlinear homogenized finite element analysis of the primary stability of the bone–implant interface10.1007/s10237-018-1038-3 ↗
39. Finite element analysis of a new dental implant design optimized for the desirable stress distribution in the surrounding bone region10.3390/prosthesis2030019 ↗
40. The relationship between implant stability quotient values and implant insertion variables: a clinical study10.1111/j.1365-2842.2011.02255.x ↗
41. An overview of basic concepts of finite element analysis and its applications in orthodontics
42. Modelling dental implant extraction by pullout and torque procedures10.1016/j.jmbbm.2017.04.010 ↗
43. Comparison of Push-In versus pull-out tests on bone-implant interfaces of rabbit Tibia dental implant healing model10.1111/j.1708-8208.2011.00357.x ↗
44. The effects of implant surface roughness and surgical technique on implant fixation in an in vitro model10.1111/j.1600-0501.2005.01202.x ↗
45. Evaluation of the primary stability of two types of dental implants through computational modeling of insertion process and pull-out test
46. Screw insertion in trabecular bone causes peri-implant bone damage10.1016/j.medengphy.2016.01.006 ↗
47. Current trends to measure implant stability10.4103/0972-4052.176539 ↗
48. Timing of loading and effect of micromotion on bone-dental implant interface: review of experimental literature10.1002/(SICI)1097-4636(199822)43:2<192::AID-JBM14>3.0.CO;2-K ↗
49. Step-wise analysis of the dental implant insertion process using the finite element technique10.1111/j.1600-0501.2007.01427.x ↗
50. Application of the finite element method in dental implant research10.1080/10255840600837074 ↗
51. Homogenized finite element simulations can predict the primary stability of dental implants in human jawbone10.1016/j.jmbbm.2024.106688 ↗
52. Peroperative estimation of bone quality and primary dental implant stability10.1016/j.jmbbm.2018.12.035 ↗
53. The effect of implant design and bone quality on insertion torque, resonance frequency analysis, and insertion energy during implant placement in low or low- to medium-density bone10.11607/ijp.4063 ↗
54. Elastic modulus and hardness of cortical and trabecular bovine bone measured by nanoindentation10.1016/S1003-6326(06)60293-8 ↗
55. An explicit micro-FE approach to investigate the post-yield behaviour of trabecular bone under large deformations10.1002/cnm.3188 ↗
56. Biochemistry, collagen synthesis
57. The effects of flute shape and thread profile on the insertion torque and primary stability of dental implants10.1016/j.medengphy.2011.09.021 ↗
58. Effects of implant thread design on primary stability—a comparison between single- and double-threaded implants in an artificial bone model10.1186/s40729-020-00239-1 ↗