Wear and biological effects of a semi-constrained total disc replacement subject to modified ISO standard test conditions
Source abstract
Abstract
Development of pre-clinical testing methodologies is an important goal for improving prediction of artificial replacement joint performance and for guiding future device design. Total disc replacement wear and the potential for osteolysis is a growing concern, therefore a parametric study on the effects on wear of altered kinematics and loading was undertaken. A standard ISO testing protocol was modified in order to study the wear behaviour of lumbar total disc replacements when subject to low cross shear input kinematics, reduced axial loading and smaller flexion-extension magnitude. Volumetric wear, bearing surface topography, and wear debris biological reactivity were assessed. The ISO standard results were expected, however, the very low cross shear test produced a level of wear approximately two orders of magnitude higher than that reported for zero cross shear motions on UHMWPE bearings. When the osteolytic potential of the wear particles was calculated, all total disc replacement simulations had lower predicted osteolytic potential compared to total hip replacements, as a consequence of the generally lower wear rates found.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Hip disorders and treatments
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (31)
Show source-provided references
1. Intervertebral disc arthroplasty10.1097/01.brs.0000146460.11591.8a ↗
2. Contact and rubbing of flat surfaces
3. The influence of contact stress on the wear of UHMWPE for total replacement hip prostheses10.1016/0043-1648(94)07032-6 ↗
4. Bible, J.E., et al., 2010. Normal functional range of motion of the lumbar spine during 15 activities of daily living. J. Spinal Disord. Tech. 23(2): 106–112.10.1097/BSD.0b013e3181981823 ↗
5. P32. comparison of wear rates: metal/UHMWPE and metal-on-metal total disc arthroplasty
6. Low back three-dimensional joint forces, kinematics, and kinetics during walking10.1016/S0268-0033(98)00069-2 ↗
7. Retrieval analysis of a ProDisc-L total disc replacement10.1097/BSD.0b013e31816dd2b6 ↗
8. Comparison of daily motion of the cervical and lumbar spine to ASTM F2423-11 and ISO 18192-1.2011 standard testing
9. Chronic failure of a lumbar total disc replacement with osteolysis. Report of a case with nineteen-year follow-up10.2106/JBJS.G.01712 ↗
10. Influence of mechanical behavior on the wear of 4 clinically relevant polymeric biomaterials in a hip simulator10.1016/S0883-5403(00)90647-8 ↗
11. A stratified approach to pre-clinical tribological evaluation of joint replacements representing a wider range of clinical conditions advancing beyond the current standard10.1039/c2fd00001f ↗
12. A novel method for the prediction of functional biological activity of polyethylene wear debris10.1243/0954411011533599 ↗
13. Wear and biological activity of highly crosslinked polyethylene in the hip under low serum protein concentrations10.1243/09544119JEIM99 ↗
14. Biotribological evaluation of artificial disc arthroplasty devices: influence of loading and kinematic patterns during in vitro wear simulation10.1007/s00586-008-0840-5 ↗
15. Introduction to lumbar total disc replacement: factors that affect tribological performance10.1243/13506501JET187 ↗
16. Wear simulation of total disc arthroplasties: Sensitivity to device design and test parameters
17. Biological reactions to wear debris in total joint replacement10.1243/0954411001535219 ↗
18. The role of macrophages in osteolysis of total joint replacement10.1016/j.biomaterials.2004.04.035 ↗
19. Quantification of the effect of cross-shear on the wear of conventional and highly cross-linked UHMWPE10.1016/j.jbiomech.2007.09.005 ↗
20. Retrieval analysis of motion preserving spinal devices and periprosthetic tissues10.1016/j.esas.2009.11.003 ↗
21. Nechtow, W., et al., 2006. IVD replacement mechanical performance depends strongly on input parameters. In: Trans 52nd Annual Meeting of the Orthopeadic Research Society, Chicago, IL.
22. Nechtow, W., et al., 2008. Vertebral motion segment dynamics influence Prodisc-L wear performance. In 54th, Orthopaedic Research Society. San Fransisco: Transactions of the Annual Meeting-Orthopaedic Research Society-CD-Rom edition-; Poster no. 1928
23. Polyethylene wear debris and long-term clinical failure of the Charite disc prosthesis: a study of 4 patients. (erratum appears in Spine. 2007 Apr 20;32(9):1052)10.1097/01.brs.0000251370.56327.c6 ↗
24. Periprosthetic tissue reactions observed at revision of total intervertebral disc arthroplasty10.1016/j.biomaterials.2008.12.071 ↗
25. Identification of nanometre-sized ultra-high molecular weight polyethylene wear particles in samples retrieved in vivo10.1302/0301-620X.90B8.20737 ↗
26. Tipper, J., et al., Quantitative comparison of UHMWPE wear particles generated from ProDisc-L total disc replacements tested under altered simulation inputs. In: Orthopaedic Research Society, Annual Meeting, 2013: San Antonio, Texas.
27. Vicars · 2007
28. The effect of anterior–posterior shear load on the wear of ProDisc-L TDR10.1007/s00586-010-1396-8 ↗
29. The effect of anterior–posterior shear on the wear of charite total disc replacement10.1097/BRS.0b013e31823cbd6e ↗
30. A unified theory of wear for ultra-high molecular weight polyethylene in multi-directional sliding10.1016/S0043-1648(00)00522-6 ↗
31. Orientation softening in the deformation and wear of ultra-high molecular weight polyethylene10.1016/S0043-1648(96)07362-0 ↗