Fiber-reinforced silicone for tracheobronchial stents: An experimental study
Source abstract
Abstract
A trachea is a tubular structure composed of smooth muscle that is reinforced with cartilage rings. Some diseases can cause sagging in smooth muscle and cartilaginous tissue. The end result is reduction (narrowing) of the trachea diameter. A solution to this problem is the use of tracheal stents, which are small tubular devices made of silicone. One is inserted into the trachea to prevent or correct its constriction. The purpose of tracheal stent use is to maintain cartilage support that would otherwise be lost in the airway. Current tracheal stent models present limitations in terms of shape and characteristics of the silicone used in their production. One of the most important is the large thickness of the wall, which makes its placement difficult; this mainly applies to pediatric patients. The wall thickness of the stent is closely related to the mechanical properties of the material. This study aims to test the reinforcement of silicone with three kinds of fibers, and then stents that were produced using fiber with the best compressive strength characteristics. Silicone samples were reinforced with polypropylene (PP), polyamide (PA), and carbon fiber (CF) at concentrations of 2% and 4% (vol%), which then underwent tensile strength and Shore A hardness testing. Samples with fiber showed good characteristics; surface analyses were carried out and they were used to produce stents with an internal diameter of 11 or 13mm and a length of 50mm. Stents underwent compression tests for qualitative evaluation. Samples with 2% and 4% CF blends showed the best mechanical performance, and they were used to produce stents. These samples presented similar compressive strengths at low deformation, but stents with a 4% CF blend exhibited improved compressive strength at deformations greater than 30-50% of their diameter (P ≤ 0.05). The addition of 2% and 4% CF blends conferred greater mechanical strength and resistance to the silicone matrix. This is particularly true at low deformation, which is the condition where the stent is used when implanted. In the finite element compression strength tests, the stent composite showed greater compression strength with the addition of fiber, and the results were in accordance with mechanical compression tests performed on the stents. In vivo tests showed that, after 30 days of post-implantation in sheep trachea, an inflammatory process occurred in the region of the trachea in contact with the stent composite and with the stent without fiber (WF). This response is a common process during the first few days of implantation.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Tracheal and airway disorders · Vascular Anomalies and Treatments · Electrospun Nanofibers in Biomedical Applications
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (29)
Show source-provided references
1. Studies on the biocompatibility of materials: fibroblast reorganization of substratum-bound fibronectin on surfaces varying in wettability10.1002/(SICI)1097-4636(199603)30:3<385::AID-JBM13>3.0.CO;2-J ↗
2. Foreign body reaction to biomaterials10.1016/j.smim.2007.11.004 ↗
3. ASTM Standard D2240, 2010. Standard Test Method for Rubber Property—Durometer Hardness. [21 March 2014] Available from: 〈http://www.astm.org/DownloadStandardB.html?ASTM%20HC=&DESIGNATION=D2240&AdID=&Split=&Campaign=Individual%20Standards%207&gclid=CN7sufXmo70CFaVxOgodOnUA2A〉.
4. ASTM Standard D3182, 2012. Standard Practice for Rubber—Materials, Equipment, and Procedures for Mixing Standard Compounds and Preparing Standard Vulcanized Sheets. [21 March 2014] Available from: 〈http://www.astm.org/Standards/D3182.htm〉.
5. ASTM Standard D412, 2013. Standard Test Methods for Vulcanized Rubber and Thermoplastic Elastomers—Tension. [21 March 2014] Available from: 〈http://www.astm.org/DownloadStandardB.html?ASTM%20HC=ASTM&DESIGNATION=D412&AdID=&Split=&Campaign=Individual%20Standards%207&gclid=CM3Y87Tmo70CFfNxOgodvhcAIQ〉.
6. ASTM Standard D5289, 2012. Standard Test Method for Rubber Property—Vulcanization Using Rotorless Cure Meters [21 March 2014] Available from: 〈http://www.astm.org/DownloadStandardA.html?ASTM%20HC=ASTM&DESIGNATION=D5289&AdID=&Split=&Campaign=Individual%20Standards%207&gclid=COnkqOzlo70CFeRlOgodezIAWg〉.
7. Biomaterials: a primer for surgeons10.1053/j.sempedsurg.2006.07.007 ↗
8. Multiscale aspects of mechanical properties of biological materials10.1016/j.jmbbm.2010.12.018 ↗
9. On studying the interaction between different stent models and rabbit tracheal tissue: numerical, endoscopic and histological comparison10.1007/s10439-015-1504-3 ↗
10. Elastomeric biomaterials for tissue engineering10.1016/j.progpolymsci.2012.05.003 ↗
11. Bioinspired polydimethylsiloxane-based composites with high shear resistance against wet tissue10.1016/j.jmbbm.2016.01.014 ↗
12. Experimental comparative study of bioactive composite with polymeric matrix for applications to oculoplastic surgery for tissue replacement
13. Effect of sterilization on the mechanical properties of silicone rubbers
14. Goldin · 1997
15. Effect of tulle on the mechanical properties of a maxillofacial silicone elastomer10.4012/dmj.27.775 ↗
16. Hin · 2004 · 1
17. ISO 10993-5, 1992. Biological evaluation of medical devices - Part 5: Tests for cytotoxicity: in vitro methods. [21 March 2014] Available from: 〈http://www.iso.org/iso/catalogue_detail.htm?Csnumber=18964〉.
18. Reconstrução da estrutura facial por biomateriais: revisão de literatura10.1590/S1983-51752010000300029 ↗
19. In vivo measurement of skin surface strain and sub-surface layer deformation induced by natural tissue stretching10.1016/j.jmbbm.2016.05.035 ↗
20. Biomateriais com aplicação na regeneração óssea – método de análise e perspectivas futuras
21. Airway stenting10.1007/s00595-004-2942-y ↗
22. Santos, L.A., 2002. Desenvolvimento de cimento de fosfato de calcio reforçado por fibras para uso na area medico-odontologica. Doctorate thesis. University of Campinas: UNICAMP. 〈http://www.bibliotecadigital.unicamp.br/document/?Code=vtls000239660〉. (accessed 21 March 2014).
23. Novel silicone stent to treat tracheobronchial lesions: results of 35 patients10.1177/0218492310388428 ↗
24. Effects on wettability by surfactant accumulation/depletion in bulk Polydimethylsiloxane (PDMS)10.1016/j.snb.2005.12.019 ↗
25. Superior in vitro biological response and mechanical properties of an implantable nanostructured biomaterial: nanohydroxyapatite-silicone rubber composite10.1016/j.actbio.2009.04.029 ↗
26. Vergani · 2008
27. Relative importance of surface wettability and charged functional groups on NIH 3T3 fibroblast attachment, spreading, and cytoskeletal organization10.1002/(SICI)1097-4636(19980905)41:3<422::AID-JBM12>3.0.CO;2-K ↗
28. Development of a modified Dumon stent for tracheal applications: an experimental study in dogs10.1590/S1806-37132008000100005 ↗
29. Yaszemski, M.J., Trantolo, D.J., Lewandrowski K.U., Hasirci, V., Altobelli, D.E., Wise, D.L., 2004. Tissue Engineering and Novel Delivery Systems. New York.10.1201/9780203913338 ↗