Long-term fretting corrosion performance of modular head-neck junctions with self-reinforced composite gaskets from PEEK and UHMWPE
Source abstract
Abstract
Soft gasket-like polymer films may provide multiple advantages in inhibiting fretting corrosion between metal-hard surfaces in total joint implants. Self-reinforced composites (SRC's) made from either poly(ether ether ketone), SRC-PEEK, or ultra high molecular weight polyethylene, SRC-PE, were fabricated and tested to investigate their ability to limit or prevent mechanically assisted corrosion in modular taper devices. Hot compaction was used to create nominally 100 μm thick unidirectional composite gaskets. These gaskets were placed on the trunnions of modular head-neck tapers and seated with 4000 N. One million cycle potentiostatic fretting corrosion tests (3000 N, R = 0.1, 15 Hz, -0.05 V) were employed to assess the ability of these SRCs to reduce or prevent fretting corrosion damage in the modular taper junction. Fretting currents and head-neck micromotion were evaluated. The results of testing, along with pull-off tests and optical and scanning electron microscopic analysis showed that SRC gaskets reduced or eliminated fretting corrosion currents, with the SRC-PEEK performing better than the SRC-PE. Fretting currents were low for SRC's compared to metal-metal tapers. No wear through of the gaskets was noted and minimal wear damage was seen in the SRC-PEEK gaskets. SRC-PE gaskets demonstrated greater deformation and damage compared to the SRC-PEEK gaskets. Pull-off loads for the SRC-PEEK were higher than SRC-PE and not statistically different than the control metal-metal junctions. There was evidence of fatigue cracks forming at the high stress concentration junctions for the SRC-PEEK at the thread form corners of the trunnion, but no loss of integrity was observed. SRC-PEEK gaskets show promise as a method to eliminate modular taper fretting corrosion.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Cardiac Structural Anomalies and Repair · Titanium Alloys Microstructure and Properties
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (31)
Show source-provided references
1. Concerns with modularity in total hip arthroplasty
2. In vitro evaluation of cytotoxicity and corrosion behavior of commercially pure titanium and Ti-6Al-4V alloy for dental implants10.4103/0972-4052.197936 ↗
3. Arthroplasty of the hip: a new operation10.1016/S0140-6736(61)92063-3 ↗
4. Properties of self-reinforced ultra-high-molecular-weight polyethylene composites10.1016/S0142-9612(96)00194-9 ↗
5. Metal-on-metal local tissue reaction is associated with corrosion of the head taper junction10.1016/j.arth.2012.03.019 ↗
6. Fretting-corrosion of materials used as orthopaedic implants10.1016/j.wear.2004.11.027 ↗
7. Gilbert · 1997
8. Medical implant corrosion: electrochemistry at metallic biomaterial surfaces
9. A tribocorrosion model linking fretting mechanics, currents and potentials: model development and experimental comparison10.1002/jbm.b.34643 ↗
10. Gilbert JL, Sivan S, Mali S, “Corrosion of modular tapers in total joint replacements: a critical assessment of design, materials, surfaces structure, mechanics, electrochemistry and biology”, ASTM Special Technical Publication on Implant Modularity, STP 1591, Eds. Greenwald, Kurtz, Lemons, Mihalko, ASTM Int 2015, pp.192-223..10.1520/STP159120140135 ↗
11. Spinal implant debris-induced osteolysis10.1097/00007632-200310151-00006 ↗
12. Current concepts review-corrosion of metal orthopaedic implants10.2106/00004623-199802000-00015 ↗
13. Fretting corrosion of Si3N4 vs CoCrMo femoral heads on Ti‐6Al‐V trunnions10.1002/jor.24681 ↗
14. Modeling of anodic current transients resulting from oxide rupture of plastically strained β+ α titanium10.1149/1.1838564 ↗
15. PEEK biomaterials in trauma, orthopedic, and spinal implants10.1016/j.biomaterials.2007.07.013 ↗
16. Do ceramic femoral heads reduce taper fretting corrosion in hip arthroplasty? A retrieval study10.1007/s11999-013-3096-2 ↗
17. Cranioplasty with customized titanium and PEEK implants in a mechanical stress model10.1089/neu.2011.1794 ↗
18. Absorbable self-reinforced polylactide (SR-PLA) composite rods for fracture fixation: strength and strength retention in the bone and subcutaneous tissue of rabbits10.1007/BF00702943 ↗
19. Construction of a tribocorrosion test apparatus for the hip joint: validation, test methodology and analysis10.1016/j.wear.2011.01.085 ↗
20. Early failures in total hip arthroplasty—a changing paradigm10.1016/j.arth.2013.12.024 ↗
21. Wear of CoCrMo alloys used in metal-on-metal hip joints: a tribocorrosion appraisal10.1016/j.wear.2012.11.061 ↗
22. Ouellette ES. Novel Methods and Self-Reinforced Composite Materials for Assessment and Prevention of Mechanically Assisted Corrosion in Modular Implants (Doctoral dissertation, Syracuse University)..
23. Properties and corrosion performance of self-reinforced composite PEEK for proposed use as a modular taper gasket10.1007/s11999-016-4861-9 ↗
24. Self-reinforced poly (ether ether ketone) and polyethylene composite gaskets for prevention of mechanically-assisted corrosion in modular taper junctions: seating, micromotion and short-term fretting corrosion10.1016/j.jmbbm.2021.104454 ↗
25. Fretting-corrosion between 316L SS and PMMA: influence of ionic strength, protein and electrochemical conditions on material wear. Application to orthopaedic implants10.1016/j.wear.2011.01.082 ↗
26. Preoperative malalignment increases risk of failure after total knee arthroplasty10.2106/JBJS.K.00607 ↗
27. PEEK dental implants: a review of the literature10.1563/AAID-JOI-D-11-00002 ↗
28. Compliant interfaces and fretting corrosion in modular taper junctions: the micromechanics of contact10.1016/j.triboint.2020.106437 ↗
29. Interfacial compliance, energy dissipation, frequency effects, and long‐term fretting corrosion performance of Ti‐6Al‐4V/CoCrMo interfaces10.1002/jbm.a.37299 ↗
30. Biodegradable self-reinforced composite materials; manufacturing structure and mechanical properties10.1016/0267-6605(92)90081-4 ↗
31. In vitro fretting crevice corrosion damage of CoCrMo alloys in phosphate buffered saline: debris generation, chemistry and distribution10.1016/j.actbio.2020.07.052 ↗