Abstract

To develop a new manufacturing process for total knee arthroplasty (TKA) femoral components, we investigated the effects of fine microstructures and precipitates on the wear rate of ultrahigh-molecular-weight polyethylene (UHMWPE) inserts against a laser-sintered Co-28Cr-6Mo alloy femoral component measured using a knee joint simulator. The tensile and fatigue strengths of the laser-sintered Co-28Cr-6Mo alloy were higher than those of cast Co-28Cr-6Mo and aged Zr-2.5Nb alloys. The laser-sintered Co-28Cr-6Mo alloy had finer microstructures and precipitates of the pi (π)-phase [(Cr, Mo)12Co8(C, N)4; lattice constants: a = b = c = 0.636 nm] in the grains and grain boundaries. The volumetric wear rate (7.16 ± 1.9 mm3/million cycles) of conventional UHMWPE (CPE, non-highly crosslinked) against a laser-sintered Co-28Cr-6Mo femoral component was lower than that of previously reported CPE inserts with Genesis Ⅱ TKA devices. Since the effect of precipitates of the π-phase on the increase in the wear rate was negligible, laser sintering is a promising new manufacturing technology for femoral components.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Advanced materials and composites · Total Knee Arthroplasty Outcomes

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (41)

Show source-provided references
1. A comparison between electromechanical and pneumatic-controlled knee simulators for the investigation of wear of total knee replacements10.1177/0954411917696519 ↗
2. Standard specification for wrought cobalt–28chromium–6molybdenum alloy for surgical implants
3. Comparison of wear in a total knee replacement under different kinematic conditions10.1023/A:1012894023657 ↗
4. Oxidized zirconium femoral components reduce polyethylene wear in a knee wear simulator10.1097/01.blo.0000148576.70780.13 ↗
5. Wear of polyethylene against oxidized zirconium femoral components10.1016/j.arth.2011.06.002 ↗
6. (i) Wear of polyethylene in artificial knee joints. Mini-symposium: total knee replacement–practical considerations10.1054/cuor.2001.0190 ↗
7. FDA Guidance · 2017
8. Effect of conformity and contact stress on wear in fixed-bearing total knee prostheses10.1016/j.jbiomech.2009.05.010 ↗
9. Comparison between force-controlled and displacement-controlled in-vitro wear testing on a widely used TKR implant
10. ISO 5832-4 · 2014
11. ISO 5832-12 · 2019
12. ISO 14243-2 · 2016
13. ISO 14243-3 · 2014
14. JIS G 1258-0 · 2007
15. JIS G 1258-2 · 2014
16. JIS G 1228 · 1997
17. JIS G 1211-3 · 2011
18. JIS Z 2241 · 2009
19. JIS T 7402–2 · 2005
20. JIS T 6121 · 2013
21. JIS T 6115 · 2013
22. JIS T 0309 · 2009
23. Comparison of wear of mobile and fixed bearing knees tested in a knee simulator10.1016/S0043-1648(03)00272-2 ↗
24. JSMS-SD-06-08 · 2008
25. β-manganese-type phase precipitated in high chromium-high nickel austenitic steels containing nitrogen10.2355/isijinternational1966.13.226 ↗
26. Comparison of long-term numerical and experimental total knee replacement wear during simulated gait loading10.1016/j.jbiomech.2006.07.027 ↗
27. Wear of fixed bearing and rotating platform mobile bearing knees subjected to high levels of internal and external tibial rotation10.1023/A:1012850224565 ↗
28. The influence of design, materials and kinematics on the in vitro wear of total knee replacements10.1016/j.jbiomech.2004.02.015 ↗
29. Knee simulator wear of vitamin E stabilized irradiated ultrahigh molecular weight polyethylene10.1016/j.arth.2011.03.006 ↗
30. Design of an advanced bearing system for total knee arthroplasty10.3390/lubricants3020475 ↗
31. 2014
32. 2015
33. Scattering-angle selective STEM images for a Co-Cr-Mo alloy fabricated by selective laser melting10.1384/jsa.26.212 ↗
34. Precipitates in biomedical Co-Cr alloys10.1007/s11837-013-0567-6 ↗
35. In-vitro and in-silico investigations on the influence of contact pressure on cross-linked polyethylene wear in total knee replacements10.1016/j.wear.2015.02.048 ↗
36. Development trends of custom-made orthopedic implants10.1007/s10047-011-0584-6 ↗
37. Chemical, physical, and mechanical properties and microstructures of laser-sintered Co-25Cr-5Mo-5W (SP2) and W-free Co-28Cr-6Mo alloys for dental application10.3390/ma12244039 ↗
38. Effects of knee simulator control method and radiation dose on UHMWPE wear rate, and relationship between wear rate and clinical revision rate in National Joint Registry10.1016/j.jmbbm.2018.09.034 ↗
39. Wear, delamination, and fatigue resistance of melt-annealed highly crosslinked UHMWPE cruciate-retaining knee inserts under activities of daily living10.1002/jor.21104 ↗
40. Wear of UHMWPE tibial inserts under simulated obese patient conditions
41. Differences of knee anthropometry between Chinese and white men and women10.1016/j.arth.2009.11.020 ↗