No effect in primary stability after increasing interference fit in cementless TKA tibial components
Source abstract
Abstract
Cementless total knee arthroplasty (TKA) implants rely on interference fit to achieve initial stability. However, the optimal interference fit is unknown. This study investigates the effect of using different interference fit on the initial stability of tibial TKA implants. Experiments were performed on human cadaveric tibias using a low interference fit of 350 μm of a clinically established cementless porous-coated tibial implant and a high interference fit of 700 μm. The Orthoload peak loads of gait and squat were applied to the specimens with a custom-made load applicator. Micromotions and gaps opening/closing were measured at the bone-implant interface using Digital Image Correlation (DIC) in 6 regions of interest (ROIs). Two multilevel linear mixed-effect models were created with micromotions and gaps as dependent variables. The results revealed no significant differences for micromotions between the two interference fits (gait p = 0.755, squat p = 0.232), nor for gaps opening/closing (gait p = 0.474, squat p = 0.269). In contrast, significant differences were found for the ROIs in the two dependent variables (p < 0.001), where more gap closing was seen in the posterior ROIs than in the anterior ROIs during both loading configurations. This study showed that increasing the interference fit from 350 to 700 μm did not influence initial stability.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Orthopedic Infections and Treatments
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (39)
Show source-provided references
1. Revision total knee arthroplasty in the young patient: is there trouble on the horizon?10.2106/JBJS.M.00131 ↗
2. Bellemans · 2005
3. 2008
4. Evaluation of interference fit and bone damage of an uncemented femoral knee implant10.1016/j.clinbiomech.2017.10.022 ↗
5. Experimental pre-clinical assessment of the primary stability of two cementless femoral knee components10.1016/j.jmbbm.2017.07.043 ↗
6. Experimental and computational analysis of micromotions of an uncemented femoral knee implant using elastic and plastic bone material models10.1016/j.jbiomech.2017.07.023 ↗
7. The effect of surface morphology on the primary fixation strength of uncemented femoral knee prosthesis: a cadaveric study10.1016/j.arth.2014.09.030 ↗
8. Standardized loads acting in kne implants10.1371/journal.pone.0086035 ↗
9. Micromotion of cementless tibial baseplates: keels with adjuvant pegs offer more stability than pegs alone10.1016/j.arth.2014.02.016 ↗
10. Micromotion of cementless tibial baseplates under physiological loading conditions10.1016/j.arth.2011.06.010 ↗
11. Differences in stiffness between a cementless and cancellous bone into varying amounts of of the interface porous implant vivo in dogs due implant motion
12. Optimal interference of the tibial component of the cementless oxford unicompartmental knee replacement10.1302/2046-3758.73.BJR-2017-0193.R1 ↗
13. Analysis of bone-prosthesis interface micromotion for cementless tibial prosthesis fixation and the influence of loading conditions10.1016/j.jbiomech.2009.12.006 ↗
14. Initial stability of cemented vs cementless tibial components under cyclic load10.1016/j.arth.2017.03.039 ↗
15. Have the causes of revision for total and unicompartmental knee arthroplasties changed during the past two decades?10.1007/s11999-017-5316-7 ↗
16. Biomechanical behaviours of the bone–implant interface: a review10.1098/rsif.2019.0259 ↗
17. The future of biologic coatings for orthopaedic implants10.1016/j.biomaterials.2013.01.074 ↗
18. Influence of limb alignment on mediolateral loading in total knee replacement10.2106/JBJS.K.00927 ↗
19. Micromotion and migration of cementless tibial trays under functional loading conditions10.1016/j.arth.2020.07.017 ↗
20. Predicting proximal femoral strength using structural engineering models10.1097/01.blo.0000164400.37905.22 ↗
21. The epidemiology of failure in total knee arthroplasty10.1302/0301-620X.98B1.36293 ↗
22. Mediolateral force distribution at the knee joint shifts across activities and is driven by tibiofemoral alignment10.1302/0301-620X.99B6.BJJ-2016-0713.R1 ↗
23. Loading of the knee joint during activities of daily living measured in vivo in five subjects10.1016/j.jbiomech.2010.03.046 ↗
24. Comparison of cementless and hybrid cemented total knee arthroplasty10.3928/01477447-20130327-16 ↗
25. The effect of the preoperative bone quality on the fixation of the tibial component in total knee arthroplasty10.1054/arth.2000.6617 ↗
26. Micromotion in cemented rotating platform total knee arthroplasty: cemented tibial stem versus hybrid fixation10.1007/s00402-005-0082-5 ↗
27. A prospective evaluation of a largely cementless total knee arthroplasty cohort without patellar resurfacing: 10-year outcomes and survivorship
28. 2019
29. Validation of model-predicted tibial tray-synthetic bone relative motion in cementless total knee replacement during activities of daily living10.1016/j.jbiomech.2018.06.024 ↗
30. Preoperative bone mineral density of the proximal tibia and migration of the tibial component after uncemented total knee arthroplasty10.1016/S0883-5403(99)90206-1 ↗
31. Cementless knee arthroplasty
32. The statistical relationship between varus deformity, surgeon's experience, BMI and tourniquet time for computer assisted total knee replacements10.1016/j.knee.2008.09.008 ↗
33. The effect of different interference fits on the primary fixation of a cementless femoral component during experimental testing10.1016/j.jmbbm.2020.104189 ↗
34. Experimental determination of friction characteristics at the trabecular bone/porous-coated metal interface in cementless implants10.1002/jbm.820270205 ↗
35. Rates of total joint replacement in the United States: future projections to 2020–2040 using the national inpatient sample10.3899/jrheum.170990 ↗
36. Micromotion analysis of the fixation of total knee tibial component10.1016/0045-7949(95)00029-G ↗
37. Measuring bone quality10.1007/s11926-013-0373-8 ↗
38. Tibio-femoral contact force distribution is not the only factor governing pivot location after total knee arthroplasty
39. Validation and sensitivity of model-predicted proximal tibial displacement and tray micromotion in cementless total knee arthroplasty under physiological loading conditions10.1016/j.jmbbm.2020.103793 ↗