Abstract

Background Implantation of nonabsorbable polypropylene (PP) mesh in the vagina is the main surgical treatment for pelvic organ prolapse (POP); however, clinical outcomes remain controversial and far from satisfactory. In particular, reducing the exposure or erosion of vaginal implants to obtain improved functional reconstruction is challenging. There is an urgent need for the development of new materials and/or products for POP treatment. A nanofibrous biomimetic mesh was recently developed to address this issue. Objective In this study, the basic properties of the newly developed mesh, including structural characteristics, mechanical properties, biological response of human umbilical cord mesenchymal stem cells in vitro, and tissue regeneration and biocompatibility in vivo, were evaluated and compared with those of Gynemesh™PS. Methods Scanning electron microscopy and uniaxial tensile methods were used to evaluate microstructure and mechanical properties, respectively. Mesenchymal stem cell growth on the meshes was observed by fluorescence microscopy to visualize the expression of enhanced red fluorescent protein. Twenty-four mature female Sprague Dawley rats were randomly assigned to two groups: group 1 (nanofibrous biomimetic mesh, Medprin, Germany, n=12) and group 2 (Gynemesh(TM)PS, Ethicon, USA; n=12). The posterior vaginal wall was incised from the introitus, and the mesh was then implanted. Three implants of each type were tested for 1, 4, 8 and 12 weeks. Connective tissue organization, inflammation, vascularization, and regenerated tissue were histologically assessed. Results The nanofibrous biomimetic mesh is a relatively heavy material and exhibited lower porosity than Gynemesh(TM)PS. The new mesh was stiffer than Gynemesh(TM)PS (p<0.001) but supported human umbilical cord mesenchymal stem cell attachment. Erosion of the grafts did not occur in any animal. The nanofibrous biomimetic mesh was encapsulated by a thicker layer of connective tissue and was associated with significantly greater inflammatory scores compared with Gynemesh(TM)PS. At 12 weeks, the vascularization of the new mesh was greater than that of Gynemesh(TM)PS (p<0.05). No significant difference in the thickness of the smooth muscle layer following implantation was observed between the two groups (p>0.05). Conclusions The nanofibrous biomimetic mesh is a candidate for reinforcing pelvic reconstruction. The mesh could be improved by decreasing its weight and stiffness and increasing its porosity. This mesh could serve as a carrier for stem cells in future regenerative medicine and tissue engineering research.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Pelvic floor disorders treatments · Hernia repair and management · Tissue Engineering and Regenerative Medicine

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (21)

Show source-provided references
1. Anterior colporrhaphy versus transvaginal mesh for pelvic-organ prolapse10.1056/NEJMoa1009521 ↗
2. The use of prosthetics in pelvic reconstructive surgery10.1016/j.bpobgyn.2005.08.013 ↗
3. Characterization of heavyweight and lightweight polypropylene prosthetic mesh explants from a single patient10.1177/1553350607306356 ↗
4. The biology behind fascial defects and the use of implants in pelvic organ prolapse repair10.1007/s00192-006-0101-2 ↗
5. FDA Public Health Notification, 2008. Serious Complications Associated with Transvaginal Placement of Surgical Mesh in Repair of Pelvic Organ Prolapse and Stress Urinary Incontinence.
6. FDA Safety Communication, 2011. Update on Serious Complications Associated with Transvaginal Placement of Surgical Mesh for Pelvic Organ Prolapse.
7. Feola · 2011
8. Characterizing the ex vivo textile and structural properties of synthetic prolapse mesh products10.1007/s00192-012-1901-1 ↗
9. Adverse effects of polyvinylidene fluoride-coated polypropylene mesh used for laparoscopic intraperitoneal onlay repair of incisional hernia10.1002/bjs.7082 ↗
10. Enzymatic adaptation in ligaments during immobilization10.1177/036354658401200311 ↗
11. Histologic response of porcine collagen-coated and uncoated polypropylene grafts in a rabbit vagina model10.1016/j.ajog.2007.12.029 ↗
12. Demands and properties of alloplastic implants for the treatment of stress urinary incontinence10.1586/17434440.4.3.349 ↗
13. Functional and morphological evaluation of a low-weight, monofilament polypropylene mesh for hernia repair10.1002/jbm.10119 ↗
14. PVDF as a new polymer for the construction of surgical meshes10.1016/S0142-9612(02)00070-4 ↗
15. Long-term histologic response to synthetic and biologic graft materials implanted in the vagina and abdomen of a rabbit model
16. Mesh implants in hernia repair. Inflammatory cell response in a rat model10.1159/000070045 ↗
17. Signs of genital prolapse in a Swedish population of women 20–59 years of age and possible related factors10.1016/S0002-9378(99)70203-6 ↗
18. Lifetime risk of undergoing surgery for pelvic organ prolapse10.1097/AOG.0b013e3181f73729 ↗
19. Dynamesh in the repair of laparoscopic ventral hernia: a prospective trial10.1007/s10029-013-1090-y ↗
20. Histologic evaluation of human cadaveric fascia lata in a rabbit vagina model10.1007/s00192-005-1321-6 ↗
21. In vitro and in vivo interactions of cells with biomaterials10.1016/0142-9612(88)90063-4 ↗