Strains in trussed spine interbody fusion implants are modulated by load and design
Source abstract
Abstract
Titanium cages with 3-D printed trussed open-space architectures may provide an opportunity to deliver targeted mechanical behavior in spine interbody fusion devices. The ability to control mechanical strain, at levels known to stimulate an osteogenic response, to the fusion site could lead to development of optimized therapeutic implants that improve clinical outcomes. In this study, cages of varying design (1.00 mm or 0.75 mm diameter struts) were mechanically characterized and compared for multiple compressive load magnitudes in order to determine what impact certain design variables had on localized strain. Each cage was instrumented with small fiducial sphere markers (88 total) at each strut vertex of the truss structure, which comprised of 260 individual struts. Cages were subjected to a 50 N control, 1000 N, or 2000 N compressive load between contoured loading platens in a simulated vertebral fusion condition, during which the cages were imaged using high-resolution micro-CT. The cage was analyzed as a mechanical truss structure, with each strut defined as the connection of two vertex fiducials. The deformation and strain of each strut was determined from 50 N control to 1000 N or 2000 N load by tracking the change in distance between each fiducial marker. As in a truss system, the number of struts in tension (positive strain) and compression (negative strain) were roughly equal, with increased loads resulting in a widened distribution (SD) compared with that at 50 N tare load indicating increased strain magnitudes. Strain distribution increased from 1000 N (+156 ± 415 με) to 2000 N (+180 ± 605 με) in 1.00 mm cages, which was similar to 0.75 mm cages (+132 ± 622 με) at 1000 N load. Strain amplitudes increased 42%, from 346με at 1000 N to 492με at 2000 N, for 1.00 mm cages. At 1000 N, strain amplitude in 0.75 mm cages (481με) was higher by 39% than that in 1.00 mm cages. These amplitudes corresponded to the mechanobiological range of bone homeostasis+formation, with 63 ± 2% (p < .05 vs other groups), 72 ± 3%, and 73 ± 1% of struts within that range for 1.00 mm at 1000 N, 1.00 mm at 2000 N, and 0.75 mm at 1000 N, respectively. The effective compressive modulus for both cage designs was also dependent on strut diameter, with modulus decreasing from 12.1 ± 2.3 GPa (1.25 mm) to 9.2 ± 7.5 GPa (1.00 mm) and 3.8 ± 0.6 GPa (0.75 mm). This study extended past micro-scale mechanical characterization of trussed cages to compare the effects of design on cage mechanical behavior at moderate (1000 N) and strenuous (2000 N) load levels. The findings suggest that future cage designs may be modulated to target desired mechanical strain regimes at physiological loads.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Spine and Intervertebral Disc Pathology · Dental Implant Techniques and Outcomes
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (27)
Show source-provided references
1. Bone response to in vivo mechanical loading in two breeds of mice10.1007/s002239900554 ↗
2. Intervertebral cages for degenerative spinal diseases10.1016/S1529-9430(03)00004-4 ↗
3. Mechanical stress and osteogenesis in vitro10.1002/jbmr.5650071406 ↗
4. Ex vivo loading of trussed implants for spine fusion induces heterogeneous strains consistent with homeostatic bone mechanobiology10.1016/j.jbiomech.2016.10.051 ↗
5. Bone remodeling I: a theory of adaptive elasticity10.1007/BF00041724 ↗
6. Bone-loading response varies with strain magnitude and cycle number10.1152/jappl.2001.91.5.1971 ↗
7. The osteocyte: an endocrine cell … and more10.1210/er.2012-1026 ↗
8. Mechanotransduction and the functional response of bone to mechanical strain10.1007/BF00302070 ↗
9. Biomechanical comparisons of different posterior instrumentation constructs after two-level ALIF: a finite element study10.1016/j.medengphy.2009.12.002 ↗
10. Biomechanics of lateral lumbar interbody fusion constructs with lateral and posterior plate fixation: laboratory investigation10.3171/2013.11.SPINE13617 ↗
11. Bone's mechanostat: a 2003 update10.1002/ar.a.10119 ↗
12. In vitro biomechanical investigation of the stability and stress-shielding effect of lumbar interbody fusion devices
13. Systematic and random errors in compression testing of trabecular bone10.1002/jor.1100150115 ↗
14. Biomechanics of trabecular bone10.1146/annurev.bioeng.3.1.307 ↗
15. Multiaxial and non-proportional loading responses, anisotropy and modeling of Ti–6Al–4V titanium alloy over wide ranges of strain rates and temperatures10.1016/j.ijplas.2006.08.006 ↗
16. Kiapour, A., Vijay, G., Ferrara, L., Hunt, J., 2011. Subsidence evaluation of 4-WEB, a novel cross strut based, interbody cage design, International Society for the Study of the Lumbar Spine (ISSLS), Gothenburg, Sweden.
17. Mechanobiology of bone tissue10.1016/j.patbio.2004.12.005 ↗
18. Mow · 2004 · 736
19. Towards a better understanding of low-back pain: a review of the mechanics of the lumbar disc10.1093/rheumatology/14.3.129 ↗
20. Biomechanics of lateral plate and pedicle screw constructs in lumbar spines instrumented at two levels with laterally placed interbody cages10.1016/j.spinee.2013.03.048 ↗
21. Biomechanics and mechanobiology of trabecular bone: a review10.1115/1.4029176 ↗
22. Mechanical evaluation of porous titanium (Ti6Al4V) structures with electron beam melting (EBM)10.1016/j.jmbbm.2009.10.006 ↗
23. Challenges to bone formation in spinal fusion10.1016/j.jbiomech.2010.10.021 ↗
24. Analysis and measurement of lumbar trunk loads in tasks involving bends and twists10.1016/0021-9290(82)90021-5 ↗
25. Three rules for bone adaptation to a mechanical stimuli10.1016/S8756-3282(98)00118-5 ↗
26. Biomechanical comparison of posterior lumbar interbody fusion and transforaminal lumbar interbody fusion by finite element analysis
27. Multiscale fluid-structure interaction modelling to determine the mechanical stimulation of bone cells in a tissue engineered scaffold10.1007/s10237-014-0599-z ↗