Explant analysis of a Discocerv cervical disc: A case study for a ceramic-on-ceramic cervical disc
Source abstract
Abstract
Explant analyses are key to better understanding the effectiveness of medical implants in replacing natural joints. For the first time, an explanted Discocerv cervical disc was examined. The implant utilised the articulation of a caudal zirconia cup (inferior component) and a cephalic alumina head (superior component). The articulating surface of the superior alumina head had an average surface roughness of 0.016 ± 0.003 μm (Sa) and the articulating surface of the inferior zirconia cup had an average surface roughness of 0.015 ± 0.002 μm (Sa). Both articulating surfaces had negative skewness, indicating the removal of local peaks. The difference between the average surface roughness of the components was not significant (p-value: 0.741). Dark grey marks were observed on both of the articulating surfaces, which were found to be adhered titanium debris that was generated due to component impingement. This titanium debris may explain the small amount of metallosis that was reported at explantation. Some transfer of zirconium to the alumina articulating surface was also seen.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Cervical and Thoracic Myelopathy · Orthopaedic implants and arthroplasty · Spine and Intervertebral Disc Pathology
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (28)
Show source-provided references
1. 2011
2. Explant analysis of total disc replacement10.1053/j.semss.2006.03.012 ↗
3. 2021
4. Wear at the taper‐trunnion junction of contemporary ceramic‐on‐ceramic hips shown in a multistation hip simulator10.1002/jbm.b.34213 ↗
5. Bhushan · 2002
6. Biotribology of artificial hip joints10.5312/wjo.v6.i1.77 ↗
7. Single level cervical arthroplasty with the Discocerv prosthesis: a preliminary report
8. 2022
9. Viscoelastic cervical total disc replacement devices: design concepts10.1016/j.spinee.2020.08.007 ↗
10. Examination of failed ex vivo metal-on-metal metatarsophalangeal prosthesis and comparison with theoretically determined lubrication regimes10.1016/j.wear.2006.11.045 ↗
11. Quantification of self-polishing in vivo from explanted metal-on-metal total hip replacements10.1016/j.triboint.2010.04.007 ↗
12. Retrieval analysis of an explanted NuNec cervical disc: a case report10.1016/j.biotri.2020.100150 ↗
13. Explant analysis of a Maestro™ wrist prosthesis and calculation of its lubrication regime10.1016/j.jmbbm.2020.103933 ↗
14. Retrieval analysis of an explanted Mobi-C cervical disc replacement: a case study10.1016/j.medengphy.2021.02.007 ↗
15. Early clinical and radiographical results of keel-less and shallow keel cervical disc replacement10.4103/1793-5482.151501 ↗
16. Ceramic fracture following cervical disc arthroplasty10.2106/JBJS.K.00077 ↗
17. Artificial disc replacement in spine surgery10.21037/atm.2019.08.26 ↗
18. Biomechanics of cervical disc arthroplasty—a review of concepts and current technology10.14444/7087 ↗
19. Total cervical disc replacement with the Discocerv® (Cervidisc Evolution) cervical prosthesis: early results of a second generation10.1007/s00590-007-0242-0 ↗
20. Metallosis is commonly associated with magnetically controlled growing rods; results from an independent multicentre explant database10.1007/s00586-021-06750-2 ↗
21. Ceramic-on-ceramic joints: a suitable Alternative material combination?
22. Topographical analysis of the femoral components of ex vivo total knee replacements10.1007/s10856-012-4815-z ↗
23. Midterm osteolysis-induced aseptic failure of the M6-C™ cervical total disc replacement secondary to polyethylene wear debris10.1007/s00586-021-07094-7 ↗
24. Does design matter? Cervical disc replacements under review10.1007/s10143-016-0765-0 ↗
25. Anterior bone loss after cervical disc replacement: a systematic review10.12998/wjcc.v8.i21.5284 ↗
26. Cervical radiculopathy: epidemiology, etiology, diagnosis and treatment10.1097/BSD.0000000000000284 ↗
27. Analysis of re-operations after cervical total disc replacement in a consecutive series of 535 patients receiving the ProDisc-C device10.1007/s00586-020-06399-3 ↗
28. Fatal cardiomyopathy after revision total hip replacement for fracture of a ceramic liner10.1302/0301-620X.95B1.30060 ↗