Characterizing fretting damage in different test media for cardiovascular device durability testing

J.D. Weaver, L. Ramirez, S. Sivan, M. Di Prima

Abstract

In vitro durability tests of cardiovascular devices are often used to evaluate the potential for fretting damage during clinical use. Evaluation of fretting damage is important because severe fretting can concentrate stress and lead to the loss of structural integrity. Most international standards call for the use of phosphate buffered saline (PBS) for such tests although there has been little evidence to date that the use of PBS is appropriate in terms of predicting the amount of fretting damage that would occur in vivo. In order to determine an appropriate test media for in vitro durability tests where fretting damage is being evaluated, we utilized an in vitro test that is relevant to cardiovascular devices both in terms of dimensions and materials (nitinol, cobalt-chromium, and stainless steel) to characterize fretting damage in PBS, deionized water (DIW), and heparinized porcine blood. Overall, tests conducted in blood were found to have increased levels of fretting damage over tests in DIW or PBS, although the magnitude of this difference was smaller than the variability for each test media. Tests conducted in DIW and PBS led to mostly similar amounts of fretting damage with the exception of one material combination where DIW had greatly reduced damage compared to PBS and blood. Differences in fretting damage among materials were also observed with nitinol having less fretting damage than stainless steel or cobalt-chromium. In general, evaluating fretting damage in PBS or DIW may be appropriate although caution should be used when selecting test media and interpreting results given some of the differences observed across different materials.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Cardiac Structural Anomalies and Repair · Cardiac Valve Diseases and Treatments

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (24)

Show source-provided references
1. Fretting corrosion in saline and serum10.1002/jbm.820150404 ↗
2. Shear-induced hemolysis: species differences10.1111/aor.12459 ↗
3. In vitro assessment of serum-saline ratios for fluid simulator testing of highly modular spinal implants with articulating surfaces10.1016/S1935-9810(08)70036-7 ↗
4. Clinical device-related article surface characterization of explanted endovascular stents: evidence of in vivo corrosion10.1002/jbm.b.31698 ↗
5. In-vivo corrosion and local release of metallic ions from vascular stents into surrounding tissue
6. Multilaboratory study of flow-induced hemolysis using the FDA benchmark nozzle model10.1111/aor.12368 ↗
7. Multi-scale mechanical investigation of stainless steel and cobalt-chromium stents10.1016/j.jmbbm.2014.09.010 ↗
8. Stent overlapping and geometric curvature influence the structural integrity and surface characteristics of coronary nitinol stents10.1016/j.jmbbm.2012.11.006 ↗
9. The influence of complexing agent and proteins on the corrosion of stainless steels and their metal components
10. Incidence and long-term clinical impact of late-acquired stent fracture after sirolimus-eluting stent implantation in narrowed coronary arteries10.1016/j.amjcard.2017.03.259 ↗
11. A new type of wear-resistant material: pseudo-elastic TiNi alloy10.1016/S0043-1648(98)00269-5 ↗
12. Exploration of TiNi shape memory alloy for potential application in a new area: tribological engineering10.1088/0964-1726/9/5/317 ↗
13. Development of standard tests to examine viscoelastic properties of blood of experimental animals for pediatric mechanical support device evaluation
14. The behavior of stainless steels in physiological solution containing complexing agent studied by X‐ray photoelectron spectroscopy10.1002/1097-4636(200011)52:2<404::AID-JBM22>3.0.CO;2-Z ↗
15. Wear mechanisms of nitinol under reciprocating sliding contact10.1016/j.wear.2014.02.018 ↗
16. The pig as a model in blood coagulation and fibrinolysis research
17. Review: stent fracture in the drug-eluting stent era10.1016/j.carrev.2016.06.002 ↗
18. Rapid platelet accumulation leading to thrombotic occlusion10.1007/s10439-011-0296-3 ↗
19. Prevalence and clinical impact of stent fractures after femoropopliteal stenting10.1016/j.jacc.2004.11.026 ↗
20. Effect of wire fretting on the corrosion resistance of common medical alloys
21. Trépanier, C., Gong, X.-.Y., Ditter, T., Pelton, A., Neely, Y., Grishaber, R., 2008. Effect of wear and crevice on the corrosion resistance of overlapped stents. SMST-2006. In: Proceedings of the International Conference on Shape Memory and Superelastic Technologies. Menlo Park, CA: SMST Society.
22. Structural fatigue of pseudoelastic NiTi shape memory wires10.1016/j.msea.2003.11.058 ↗
23. Research combining experiment and FEM analysis on sliding wear behaviors and mechanisms of TiNi alloy10.1016/j.wear.2017.01.093 ↗
24. A brief review of bio-tribology in cardiovascular devices10.1016/j.bsbt.2015.11.002 ↗