journal article·2016-09Open access

Detection of degradation in polyester implants by analysing mode shapes of structure vibration

Hassan Samami, Jingzhe Pan

Abstract

This paper presents a numerical study on using vibration analysis to detect degradation in degrading polyesters. A numerical model of a degrading plate sample is considered. The plate is assumed to degrade following the typical behaviour of amorphous copolymers of polylactide and polyglycolide. Due to the well-known autocatalytic effect in the degradation of these polyesters, the inner core of the plate degrades faster than outer surface region, forming layers of materials with varying Young׳s modulus. Firstly the change in molecular weight and corresponding change in Young׳s modulus at different times are calculated using the mathematical models developed in our previous work. Secondly the first four mode shapes of transverse vibration of the plate are calculated using the finite element method. Finally the curvature of the mode shapes are calculated and related to the spatial distribution of the polymer degradation. It is shown that the curvature of the mode shapes can be used to detect the onset and distribution of polymer degradation. The level of measurement accuracy required in an experiment is presented to guide practical applications of the method. At the end of this paper a demonstration case of coronary stent is presented showing how the method can be used to detect degradation in an implant of sophisticated structure.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Structural Health Monitoring Techniques · Corrosion Behavior and Inhibition · Concrete Corrosion and Durability

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (28)

Show source-provided references
1. Finite-element analysis of a stenotic artery revascularization through a stent insertion10.1080/10255840108908007 ↗
2. Coronary artery stents: identification and evaluation10.1136/jcp.2004.024174 ↗
3. Finite element analysis and stent design: reduction of dogboning10.3233/THC-2006-144-506 ↗
4. Vibration of cracked structures: a state of the art review10.1016/0013-7944(94)00175-8 ↗
5. A summary review of vibration-based damage identification methods10.1177/058310249803000201 ↗
6. Doebling · 1996
7. Mechanical properties of coronary stents determined by using finite element analysis10.1016/S0021-9290(01)00026-4 ↗
8. Vibration-based damage identification methods: a review and comparative study10.1177/1475921710365419 ↗
9. A comparison of the conformability of everolimus-eluting bioresorbable vascular scaffolds to metal platform coronary stents
10. Finite element analysis of covered microstents10.1016/j.jbiomech.2004.06.008 ↗
11. Finite element analysis of degradation of biodegradable medical devices
12. The comprehensive finite element model for stenting: the influence of stent design on the outcome after coronary stent placement
13. Cardiovascular stent design and vessel stresses: a finite element analysis10.1016/j.jbiomech.2004.07.022 ↗
14. Correction: redefining the septal L-strut in septal surgery
15. Bioabsorbable coronary stents
16. Coronary stents: a materials perspective10.1016/j.biomaterials.2006.11.042 ↗
17. Mechanical behavior of coronary stents investigated through the finite element method10.1016/S0021-9290(02)00033-7 ↗
18. Narracott, A.J., Lawford, P.V., Gunn, J.P., Hose, D.R., 2007. Balloon folding affects the symmetry of stent deployment: experimental and computational evidence. In: The proceedings of the 29th Annual International Conference of the IEEE, Engineering in Medicine and Biology Society (EMBS), pp. 3069–307310.1109/IEMBS.2007.4352976 ↗
19. The evolution of cardiovascular stent materials and surfaces in response to clinical drivers: a review10.1016/j.actbio.2008.11.012 ↗
20. Bioabsorbable coronary stents10.1161/CIRCINTERVENTIONS.109.859173 ↗
21. Pan · 2014
22. Damage detection from changes in curvature mode shapes10.1016/0022-460X(91)90595-B ↗
23. Detection of structural damage through changes in frequency: a review10.1016/S0141-0296(96)00149-6 ↗
24. A constitutive law for degrading bioresorbable polymers10.1016/j.jmbbm.2016.02.026 ↗
25. Autocatalytic hydrolysis of amorphous-made polylactides: effects of l-lactide content, tacticity, and enantiomeric polymer blending10.1016/S0032-3861(01)00752-2 ↗
26. Analysis of the transient expansion behavior and design optimization of coronary stents by finite element method10.1016/j.jbiomech.2004.11.003 ↗
27. Yang, A.S., Jow, K., Morrison, A.D., Limon, T.A., 2013. Polymeric stent. Google Patents.