journal article·2023-02

A hazardous boundary of Poly(L-lactic acid) braided stent design: Limited elastic deformability of polymer materials

Junjie Li, Jie Cheng, Xue Hu, Jiaming Liu, Yuan Tian, Gensheng Wu, Li Chen, Yi Zhang, Gutian Zhao, Zhonghua Ni

Publisher page

Abstract

Poly (L-lactic acid) (PLLA) braided stents, which are expected to replace metal stents, are promising in peripheral vascular therapy due to their superior biocompatibility. Although various design ideas have been proposed and investigated on metal stents, few researches are related to the design theory of PLLA braided stent. In this article, mechanical performance of PLLA braided stents with different parameters was systematically evaluated, and a design theory based on material properties was proposed. Different from metal materials, the risk of filament deformation beyond elastic zone should be evaluated and controlled in PLLA stent design. The findings were obtained through combination study of experiments and simulations. Design parameters, including pitch angle and stent diameter, played a crucial role in mechanical performance of PLLA braided stent. The deformation of PLLA stents with larger pitch angles and stent diameters was in elastic zone and thus presented better mechanical performance with satisfactory resilience. This work could provide meaningful suggestions for preparing bioresorbable braided stents with suitable design parameters.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Electrospun Nanofibers in Biomedical Applications · Bone Tissue Engineering Materials · Coronary Interventions and Diagnostics

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (36)

Show source-provided references
1. Endovascular stent-graft applications in iatrogenic vascular injuries10.1007/s00270-003-0038-5 ↗
2. 'An investigation into the crystalline morphology transitions in poly-L-lactic acid (PLLA) under uniaxial deformation in the quasi-solid-state regime10.1016/j.eurpolymj.2018.01.031 ↗
3. 'Using 3D printing and femtosecond laser micromachining to fabricate biodegradable peripheral vascular stents with high structural uniformity and dimensional precision10.1007/s00170-021-07446-z ↗
4. 'Chemical vs thermal accelerated hydrolytic degradation of 3D-printed PLLA/PLCL bioresorbable stents: characterization and influence of sterilization
5. 'A comparison of balloon- and self-expanding stents10.1080/136457002760273386 ↗
6. 'Update on vascular endoprostheses (stents): from experimental studies to clinical practice10.1590/S1677-54492008000400010 ↗
7. Biodegradable vascular stents with high tensile and compressive strength: a novel strategy for applying monofilaments via solid-state drawing and shaped-annealing processes10.1039/C7BM00011A ↗
8. 'Biocompatibility properties of a new braided biodegradable urethral stent: a comparison with a biodegradable spiral and a braided metallic stent in the rabbit urethra10.1111/j.1464-410X.2006.06000.x ↗
9. 'Elastic recovery of polymeric braided stents under cyclic loading: preliminary assessment10.1016/j.jmbbm.2019.06.018 ↗
10. 'A study of the geometrical and mechanical properties of a self-expanding metallic stent-theory and experiment10.1002/jab.770040111 ↗
11. 'A review on design characteristics and fabrication methods of metallic cardiovascular stents
12. 'Effects of annealing constraint methods on poly(L‐lactic acid) monofilaments for application in stents annealing10.1002/pat.5266 ↗
13. 'Mixed-braided stent: an effective way to improve comprehensive mechanical properties of poly (L-lactic acid) self-expandable braided stent10.1016/j.jmbbm.2022.105123 ↗
14. 'Evaluation of resistance to radial cyclic loads of poly(L-lactic acid) braided stents with different braiding angles10.1016/j.ijbiomac.2022.07.107 ↗
15. 'Use of the gore tigris vascular stent in advanced femoropopliteal peripheral arterial disease10.1016/j.jvir.2017.12.006 ↗
16. The Carotid Wallstent for the endovascular treatment of carotid artery stenosis10.1016/j.inat.2017.08.002 ↗
17. 'A finite element investigation on design parameters of bare and polymer-covered self-expanding wire braided stents10.1016/j.jmbbm.2020.104305 ↗
18. Bio-resorbable polymer stents: a review of material progress and prospects10.1016/j.progpolymsci.2018.05.002 ↗
19. 'Braided or laser-cut self-expanding nitinol stents for the common femoral vein in patients with post-thrombotic syndrome10.1016/j.jvsv.2020.08.027 ↗
20. 'Uniform expansion of a polymeric helical stent
21. 'Fatigue and durability of Nitinol stents10.1016/j.jmbbm.2007.08.001 ↗
22. 'Physical and mechanical degradation behaviour of semi-crystalline PLLA for bioresorbable stent applications10.1016/j.jmbbm.2021.104409 ↗
23. 'Influence of design parameters on the mechanical behavior and porosity of braided fibrous stents10.1016/j.matdes.2015.07.051 ↗
24. 'Viscoelastic braided stent: finite element modelling and validation of crimping behaviour10.1016/j.matdes.2017.02.044 ↗
25. 'The development of bioresorbable composite polymeric implants with high mechanical strength10.1038/nmat5016 ↗
26. 'Self-expanding nitinol stents: material and design considerations10.1007/s00330-003-2022-5 ↗
27. Selection of helical braided flow diverter stents based on hemodynamic performance and mechanical properties10.1136/neurintsurg-2016-012561 ↗
28. Computational and experimental investigation into mechanical performances of Poly-L-Lactide Acid (PLLA) coronary stents10.1016/j.jmbbm.2016.08.033 ↗
29. 'Mechanical response of a metallic aortic stent-Part I: pressure-diameter relationship10.1115/1.1782650 ↗
30. 'Influence of thermal annealing on the mechanical properties of PLLA coiled stents10.1007/s13239-014-0189-3 ↗
31. 'Analytical methods for braided stents design and comparison with FEA10.1016/j.jmbbm.2021.104560 ↗
32. 'Braided bioresorbable cardiovascular stents mechanically reinforced by axial runners10.1016/j.jmbbm.2018.09.003 ↗
33. 'Effect of dynamic and static loading during in vitro degradation of a braided composite bioresorbable cardiovascular stent10.1016/j.matlet.2019.04.097 ↗
34. 'Influence of parameters on mechanical properties of poly (L-lactic acid) helical stents10.1002/jbm.b.35031 ↗
35. 'Effects of constraint between filaments on the radial compression properties of poly (l-lactic acid) self-expandable braided stents10.1016/j.polymertesting.2020.106963 ↗
36. 'The impact of wire stent fabrication technique on the performance of stent placement