Highly cross-linked polyethylene acetabular liners retrieved four to five years after revision surgery: A report of two cases

Matthew G. Teeter, Douglas D.R. Naudie, Kory D. Charron, David W. Holdsworth

Publisher page

Abstract

There is currently considerable interest in the use of highly cross-linked polyethylene (XLPE) acetabular liners for total hip arthroplasty (THA). In literature, only a single retrieval analysis of one type of XLPE liner implanted for greater than four years exists. The purpose of the present report is to quantify surface deviations in two XLPE liners implanted during revision THA and retrieved between four to five years after implantation. The two XLPE acetabular liners (Reflection, Smith and Nephew Inc., Memphis, TN) were retrieved from patients undergoing their second revision surgery, at 4.90 and 4.07 years. The retrieved liners and a new, non-implanted, unworn liner of the same size were scanned using micro-computed tomography (micro-CT). Articular surface deviation maps were created by comparing the retrievals to the unworn liner, based on the liner geometry obtained from micro-CT. The linear penetration rates were found to be 0.018 and 0.008 mm/year. Localized scratches and pits with deviations greater than 0.205 mm were also found on the articular surfaces of both liners. The XLPE liners retrieved from the two cases demonstrated low linear penetration rates. Regions with greater focal deviations were also apparent, likely due to third-body wear. The results are consistent with previously published clinical follow-ups of other XLPE liners.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Orthopedic Infections and Treatments · Total Knee Arthroplasty Outcomes

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (16)

Show source-provided references
1. Wear is reduced in THA performed with highly cross-linked polyethylene10.1007/s11999-008-0661-1 ↗
2. Reduction of osteolysis with use of Marathon cross-linked polyethylene: a concise follow-up, at a minimum of five years, of a previous report10.2106/JBJS.F.00991 ↗
3. Improvement in the assessment of wear of total knee replacements using coordinate-measuring machine techniques10.1243/09544119JEIM289 ↗
4. Validation of a micro-CT technique for measuring volumetric wear in retrieved acetabular liners10.1002/jbm.b.30318 ↗
5. Clinical performance of a Durasol highly cross-linked polyethylene acetabular liner for total hip arthroplasty at five years
6. A randomized prospective evaluation of outcomes after total hip arthroplasty using cross-linked marathon and non-cross-linked Enduron polyethylene liners10.1016/j.arth.2006.05.002 ↗
7. Retrieval analyses of highly cross-linked polyethylene acetabular liners four and five years after implantation10.1302/0301-620X.89B8.18809 ↗
8. Wear rate of highly cross-linked polyethylene in total hip arthroplasty. A randomized controlled trial10.2106/JBJS.H.00244 ↗
9. Evaluation and management of the infected total hip and knee10.3928/01477447-20080601-22 ↗
10. Surface analysis of early retrieved acetabular polyethylene liners: a comparison of conventional and highly crosslinked polyethylenes10.1016/j.arth.2003.08.003 ↗
11. In vivo and in vitro surface changes in a highly cross-linked polyethylene10.1016/S0883-5403(03)00303-6 ↗
12. Retrieval analysis of failed constrained acetabular liners10.1016/j.arth.2009.05.002 ↗
13. 3D surface deviation maps for analysis of retrieved polyethylene acetabular liners using micro-computed tomography10.1016/j.arth.2009.11.001 ↗
14. The accuracy and precision of a micro computer tomography volumetric measurement technique for the analysis of in-vitro tested total disc replacements10.1243/09544119JEIM515 ↗
15. Whittaker, J.P., Charron, K.D., McCalden, R.W., MacDonald, S.J., Bourne, R.B., 2009. Comparison of steady state femoral head penetration rates between 2 highly cross-linked polyethylenes in total hip arthroplasty, J. Arthroplasty (in press).10.1016/j.arth.2009.05.006 ↗
16. Oxidative degradation in highly cross-linked and conventional polyethylene after 2 years of real-time shelf aging10.1016/j.biomaterials.2005.11.010 ↗