journal article·2022-10

Mechanical properties analysis of medical endodontic instruments based on parameterization

Xiulin Hu, Gongwei Zhao, Jiachen Du, Nanhai Ye

Publisher page

Abstract

To study the relationship between structural parameters and mechanical properties of endodontic instruments, the T02004B25 nickel-titanium endodontic instrument was selected for bending and torsion tests and finite element simulation analysis, which demonstrated the feasibility of simulation analysis method. Then based on the idea of parametric design, the models of the endodontic instruments with different structural parameters (cross-section, pitch, taper) were established, and the bending-torsion performance simulation analysis was completed. The results showed that the mechanical properties of endodontic instruments with different structural parameters are different. It is necessary to find the optimal parameters for different structure parameters of endodontic instruments to maximize their service life.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Endodontics and Root Canal Treatments · Occupational health in dentistry · Engineering Technology and Methodologies

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (25)

Show source-provided references
1. Enhancement of NiTi superelastic endodontic instruments by TiO2 coating10.1016/j.msec.2016.06.031 ↗
2. Comparison of torsional stiffness of nickel-titanium rotary files with different geometric characteristics10.1016/j.joen.2011.05.032 ↗
3. Effect of eccentricity on mechanical properties and chip tolerance of nickel-titanium endodontic instruments
4. Cyclic fatigue resistance of one curve, 2Shape, ProFile vortex, vortex blue, and RaCe nickel-titanium rotary instruments in single and double curvature canals10.1016/j.joen.2018.07.023 ↗
5. Dynamic cyclic and torsional fatigue resistance of TruNatomy compared with different nickel-titanium rotary instruments10.1111/aej.12396 ↗
6. Atomic force microscopic evaluation of nanostructure alterations of rotary NiTi instruments after immersion in irrigating solutions10.1111/iej.12189 ↗
7. The geometric effect of an off-centered cross-section on nickel–titanium rotary instruments: a finite element analysis study10.1016/j.jds.2016.11.005 ↗
8. Phase transformation behaviour and bending properties of hybrid nickel-titanium rotary endodontic instruments10.1111/j.1365-2591.2007.01203.x ↗
9. Stress distribution of three NiTi rotary files under bending and torsional conditions using a mathematic analysis10.1111/j.1365-2591.2008.01481.x ↗
10. Cyclic fatigue of endodontic nickel titanium rotary instruments: static and dynamic tests10.1097/00004770-200206000-00007 ↗
11. Cyclic fatigue resistance of ProTaper universal instruments when subjected to static and dynamic tests10.1016/j.tripleo.2010.05.013 ↗
12. Cyclic fatigue of ProTaper instruments10.1016/j.joen.2006.09.003 ↗
13. Fatigue resistance of engine-driven rotary nickel-titanium endodontic instruments10.1097/00004770-200211000-00005 ↗
14. Mechanical behaviour of nickel-titanium rotary endodontic instruments in simulated clinical conditions: a computational study10.1111/j.1365-2591.2008.01454.x ↗
15. Mechanical behavior of three nickel-titanium rotary files: a comparison of numerical simulation with bending and torsion tests
16. Effect of physical vapor deposition on cutting efficiency of nickel-titanium files10.1097/00004770-200212000-00002 ↗
17. Mechanical and metallurgical properties of various nickel-titanium rotary instruments10.1155/2017/4528601 ↗
18. Standardization · 2008
19. Evidence for reduced fatigue resistance of contemporary rotary instruments exposed to body temperature10.1016/j.joen.2016.01.025 ↗
20. Cyclic fatigue resistance of two variable-taper rotary file systems: ProTaper universal and V-Taper10.1016/j.joen.2009.01.002 ↗
21. Comparative study of torsional and bending properties for six models of nickel-titanium root canal instruments with different cross-sections10.1016/j.joen.2005.08.012 ↗
22. Cyclic fatigue of three types of rotary nickel-titanium files in a dynamic model10.1016/j.joen.2005.10.013 ↗
23. An in vitro study of the torsional properties of new and used K3 instruments10.1046/j.1365-2591.2003.00732.x ↗
24. An update on nickel-titanium rotary instruments in endodontics: mechanical characteristics, testing and future perspective & mdash; an overview10.3390/bioengineering8120218 ↗
25. A paradigm shift for torsional stiffness of nickel-titanium rotary instruments: a finite element analysis10.1016/j.joen.2021.04.017 ↗