journal article·2014-06

Three-dimensional and chemical changes on the surface of a 3-year clinically retrieved oxidized titanium dental implant

Raúl González-García, Alberto Monje, María Coronada Fernández-Calderón, Margarita Hierro-Oliva, María Luisa González-Martín, Florencio Monje

Publisher page

Abstract

Objective To report the main topographical features in the micro- and nano-scales and to assess implant chemical changes of the surface of a 3-year clinically retrieved oxidized titanium dental implant, and compare them with a similar, unused implant. Materials and methods The surface of the oxidized titanium dental implants was assessed by surface electron microscopy (SEM) analysis at increasing magnifications. X-ray photoelectron spectroscopy (XPS) measurement was performed to analyze the implants surface chemistry. XPS spectra were acquired before and after sputtering with an Ar(+) ion etching of 3keV. Results With a length of 10-40μm, and a width of 0.05-0.1μm, numerous cracks were ubiquitous along the implant surfaces. Chimney-like structures formed micropores between 1 and 5μm, with up to 40% of them partially or totally broken in the retrieved implant. In relation to chemical composition, Ti and O were predominant in both the unused and in the retrieved implant. N was present in high concentrations (11.49at%) at the retrieved implant surface, in contrast with those observed for the unused implant (1.14at%). Also, C was present in higher concentrations in the retrieved implant surface, while drastically decreased following the sputter-cleaning process. Conclusion While cracks were ubiquitous present from the manufacturing, broken chimney-like structures forming micropores in the clinically retrieved implant may be attributable to excessive mechanical friction forces during the insertion of the implant. Chemical composition of the implant surface may be subjected to changes because of the in vivo environment, with increase of N and C, and decrease of Ca and P.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Dental Implant Techniques and Outcomes · Bone Tissue Engineering Materials · Titanium Alloys Microstructure and Properties

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (17)

Show source-provided references
1. Experimental studies on oxidized implants. A histomorphometrical and biomechanical analysis
2. The optimum pore size for the fixation of porous surfaced metal implants by the ingrowth of bone10.1097/00003086-198007000-00045 ↗
3. An overview of the corrosion aspect of dental implants (titanium and its alloys)10.4103/0970-9290.49068 ↗
4. Interface mechanics of porous titanium implants10.1002/jbm.820150111 ↗
5. Surface characterization analysis of failed dental implants using scanning electron microscopy10.3109/00016357.2011.569507 ↗
6. Immediate occlusal loading of Brånemark MkIV TiUnite™ implants placed predominantly in soft bone: 1-year results of a prospective, clinical study10.1111/j.1708-8208.2003.tb00015.x ↗
7. Anchorage of titanium implants with different surface characteristics: an experimental study in rabbits10.1111/j.1708-8208.2000.tb00002.x ↗
8. Histologic evaluation of bone response to oxidized and turned titanium micro-implants in human jawbone
9. XPS, AES and SEM analysis of recent dental implants10.1016/j.actbio.2009.01.049 ↗
10. Bioactive metals: preparation and properties10.1023/B:JMSM.0000011809.36275.0c ↗
11. Biomechanical comparison of the sandblasted and acid-etched and the machined and acid-etched titanium surface for dental implants10.1002/jbm.10063 ↗
12. Changes in surface compositon of the Ti–6Al–4V implant alloy by cultured macrophage cells10.1016/j.apsusc.2003.09.027 ↗
13. Immediate loading of Brånemark System with TiUnite and machined surfaces in the posterior mandible: a randomised, open-ended trial10.1111/j.1708-8208.2003.tb00016.x ↗
14. Histology of retrieved immediately and early loaded oxidized implants. Light microscopic observations after 5 to 9 months of loading in the posterior mandible of 5 cases10.1111/j.1708-8208.2003.tb00020.x ↗
15. The human bone-oxidized titanium implant interface: a light microscopic, scanning electron microscopic, back-scatter scanning electron microscopic, and energy-dispersive X-ray study of clinically retrieved dental implants
16. Bone reactions to oxidized titanium implants with electrochemical anion sulphuric acid and phosphoric acid incorporation10.1111/j.1708-8208.2002.tb00156.x ↗
17. Biotribocorrosion of CrCoMo orthopaedic implant materials: assessing the formation and effect of the biofilm10.1016/j.triboint.2007.02.019 ↗