journal article·2023-07

Insight on the in vivo wear characteristics of goat artificial cervical disc implanted for 6 months

Junhui Li, Song Wang, Fengxian Wang, Xing Yu, Lin Xu

Publisher page

Abstract

Since the prosthesis may suffer overload or extreme motion during the daily activities, some special failure modes may be found in service. In order to give an insight on the in vivo stability of artificial cervical disc, the wear characteristics of goat prosthesis were studied after implanted in goat animal for 6 months. The prosthesis was designed with a ball-on-socket structure under the material combination of PE-on-TC4. The X-ray examination was performed to monitor the in vivo wear process. The worn morphology and wear debris were analyzed in detail by EDX and SEM. The result indicated that goat prosthesis revealed good safety and effectiveness during 6-month in vivo wear test. The wear damage occurred only on nucleus pulposus component with the dominant failure mode of surface fatigue and deformation. The damage distribution and wear severity was seriously uneven with a trend that the closer to the edge, the more severe the wear. For example, slippage phenomenon caused a wide and curved severe ploughing damage on the edge. Three kinds of debris were found including bone debris, carbon-oxygen compound debris and PE wear debris. Both bone debris and carbon-oxygen compound debris came from superior endplate while PE wear debris came from nucleus pulposus. The debris proportion for endplate was 82% for bone debris, 15% for carbon-oxygen compound debris and 3% for PE debris while for nucleus pulposus it was 8% for carbon-oxygen compound debris and 92% for PE debris. The size range of PE debris for nucleus pulposus was 0.1-100 μm, with an average size of 9.58 ± 16.34 μm. For the bone debris of endplate components, the size range was 0.1-600 μm, with an average size of 49.18 ± 94.54 μm. After wear test, the equivalent elastic modulus of nucleus pulposus increased from 28.55 MPa to 38.25 MPa. The results of FT-IR spectrum showed that the functional groups on the surface of polyethylene have not changed significantly after wear test. The results indicated that there were some differences in wear characteristics of wear morphology and wear debris between in vivo wear and in vitro wear.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Tribology and Wear Analysis · Spine and Intervertebral Disc Pathology

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (30)

Show source-provided references
1. Three-dimensional finite element simulation and experimental validation of sliding wear
2. In vitro method to quantify and visualize mechanical wear in human meniscus subjected to joint loading10.1016/j.jmbbm.2022.105338 ↗
3. Anatomization of wear behaviour of materials for total hip arthroplasty bearing surfaces: a review
4. Occurrence of periodontitis and dental wear in dairy goats10.1016/j.smallrumres.2019.05.004 ↗
5. Prospective animal study of 5 FDA-approved cervical artificial discs10.1016/j.spinee.2019.05.131 ↗
6. Wear and corrosion characteristics of nano-crystalline tantalum nitride coatings deposited on CoCrMo alloy for hip joint applications10.1016/j.matchar.2021.111516 ↗
7. Third body damage and wear in arthroplasty bearing materials: a review of laboratory methods
8. A novel hip joint prosthesis with uni-directional articulations for reduced wear10.1016/j.jmbbm.2021.105072 ↗
9. Minimally invasive hydrogel nucleoplasty in a goat model of moderate severity disc degeneration10.1016/j.spinee.2021.05.196 ↗
10. Titanium versus polyetheretherketone versus structural allograft in anterior cervical discectomy and fusion: a systematic review10.1016/j.bas.2022.100923 ↗
11. Cervical total disc replacement: available implant size matters10.1016/j.spinee.2021.05.190 ↗
12. PEEK as a replacement for CoCrMo in knee prostheses: pin-on-disc wear test of PEEK-on-polyethylene articulations10.1016/j.biotri.2021.100189 ↗
13. Aseptic and septic prosthetic joint loosening: impact of biomaterial wear on immune cell function, inflammation, and infection10.1016/j.biomaterials.2021.121127 ↗
14. Radiographic evaluation of linear wear of bipolar hemiarthroplasty devices in vivo10.1016/j.jcot.2018.07.016 ↗
15. The effect of anterior cervical discectomy and fusion procedure duration on patient-reported outcome measures10.1016/j.wneu.2022.05.016 ↗
16. Cervical artificial disc replacement with ProDisc-C: clinical and radiographic outcomes with long-term follow-up10.1016/j.jocn.2013.09.013 ↗
17. Delamination-fretting wear failure evaluation at HAp-Ti-6Al–4V interface of uncemented artificial hip implant10.1016/j.jmbbm.2021.104657 ↗
18. A minimum of 15-year results of cementless total hip arthroplasty using a 28-mm metal-on-metal articulation10.1016/j.arth.2019.03.010 ↗
19. A comparative study on fretting wear and frictional heating behavior of PEEK composites for artificial joint applications10.1016/j.polymertesting.2022.107552 ↗
20. How does bicycling affect the longevity of Total Hip Arthroplasty? A finite element wear analysis
21. Clinical and radiological mid- to long-term investigation of anterior lumbar stand-alone fusion: incidence of reoperation and adjacent segment degeneration10.1016/j.bas.2022.100924 ↗
22. Isolation and characterisation of wear debris surrounding failed total ankle replacements10.1016/j.actbio.2023.01.051 ↗
23. Isolation and characterisation of wear debris surrounding failed total ankle replacements10.1016/j.actbio.2023.01.051 ↗
24. In vivo volumetric and linear wear measurement of reverse shoulder arthroplasty at minimum 5-year follow-up10.1016/j.jse.2019.11.031 ↗
25. Vicars · 2017 · 7 · 246
26. Artificial discs in cervical disc replacement: a meta-analysis for comparison of long-term outcomes10.1016/j.wneu.2019.10.032 ↗
27. The biotribological behavior of an artificial cervical disc model with ball-on-socket contact type under different material configurations10.1007/s11249-016-0767-7 ↗
28. The biotribological behavior of an artificial cervical disc model with ball-on-socket contact type under different material configurations10.1007/s11249-016-0767-7 ↗
29. Wear assessment of a Ti–6Al–4V motion-preserving porous artificial-cervical-joint fabricated by SLM after surface carburization10.1016/j.ceramint.2022.05.296 ↗
30. Effect of osteoporosis on adjacent segmental degeneration after posterior lumbar interbody fusion under whole body vibration10.1016/j.wneu.2021.06.031 ↗