journal article·2023-06

Design of a double acting pneumatic cartilage loading device for magnetic resonance imaging

Jessica C. Küpper, Emily S. Sullivan, Robin J.N. Coope, David R. Wilson

Publisher page

Abstract

Studies of osteoarthritis initiation and progression that measure strain in cartilage require physiological loading levels. Many studies use magnetic resonance (MR) imaging, which necessitates a MR-compatible loading device. In this study, the design and validation of a new device, the cartilage compressive actuator (CCA), is presented. The CCA is designed for high-field (e.g., 9.4 T) small-bore MR scanners, and meets a number of design criteria. These criteria include capability for testing bone-cartilage samples, MR compatibility, constant load and incremental strain application, a water-tight specimen chamber, remote control, and real time displacement feedback. The mechanical components in the final design include an actuating piston, a connecting chamber, and a sealed specimen chamber. An electro-pneumatic system applies compression, and an optical Fibre-Bragg grating (FBG) sensor provides live displacement feedback. A logarithmic relationship was observed between force exerted by the CCA and pressure (R2 = 0.99), with a peak output force of 653 ± 2 N. The relationship between FBG sensor wavelength and displacement was linear when calibrated both outside (R2 = 0.99) and inside (R2 = 0.98) the MR scanner. Average slope was similar between the two validation tests, with a slope of -4.2 nm/mm observed inside the MR scanner and -4.3 to -4.5 nm/mm observed outside the MR scanner. This device meets all design criteria and represents an improvement over published designs. Future work should incorporate a closed feedback loop to allow for cyclical loading of specimens.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Osteoarthritis Treatment and Mechanisms · Knee injuries and reconstruction techniques · Total Knee Arthroplasty Outcomes

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (25)

Show source-provided references
1. The role of ambulatory mechanics in the initiation and progression of knee osteoarthritis10.1097/01.bor.0000240365.16842.4e ↗
2. Articular cartilage deformation determined in an intact tibiofemoral joint by displacement-encoded imaging10.1002/mrm.21927 ↗
3. Review of fiber Bragg grating sensor technology10.1007/s12200-011-0130-4 ↗
4. Comparison of cartilage mechanical properties measured during creep and recovery10.1038/s41598-020-58220-2 ↗
5. Depth-dependent changes in cartilage T2 under compressive strain: a 7T MRI study on human knee cartilage10.1016/j.joca.2020.05.012 ↗
6. European society of biomechanics S.M. Perren award 2022: standardized tibio-femoral implant loads and kinematics10.1016/j.jbiomech.2022.111171 ↗
7. Do dynamic-based MR knee kinematics methods produce the same results as static methods?10.1002/mrm.24425 ↗
8. Implantable sensor technology: measuring bone and joint biomechanics of daily life in vivo10.1186/ar4138 ↗
9. Measuring cartilage morphology with quantitative magnetic resonance imaging10.1055/s-2004-861579 ↗
10. The relation between the biochemical composition of knee articular cartilage and quantitative MRI: a systematic review and meta-analysis10.1016/j.joca.2021.10.016 ↗
11. Medial meniscal displacement and strain in three dimensions under compressive loads: MR assessment10.1002/jmri.24461 ↗
12. Patterns of cartilage stiffness on normal and degenerate human femoral heads10.1016/0021-9290(71)90049-2 ↗
13. Prevalence trends of site-specific osteoarthritis from 1990 to 2019: findings from the global burden of disease study 201910.1002/art.42089 ↗
14. Existing comorbidities in people with osteoarthritis: a retrospective analysis of a population-based cohort in alberta, Canada10.1136/bmjopen-2019-033334 ↗
15. Functional cartilage MRI T2 mapping: evaluating the effect of age and training on knee cartilage response to running10.1016/j.joca.2009.11.011 ↗
16. Functional in situ assessment of human articular cartilage using MRI: a whole-knee joint loading device10.1007/s10237-017-0932-4 ↗
17. Functional MRI mapping of human meniscus functionality and its relation to degeneration10.1038/s41598-020-59573-4 ↗
18. Functional imaging in OA: role of imaging in the evaluation of tissue biomechanics10.1016/j.joca.2014.05.016 ↗
19. Toward an MRI-based method to measure non-uniform cartilage deformation: an MRI-cyclic loading apparatus system and steady-state cyclic displacement of articular cartilage under compressive loading10.1115/1.1560141 ↗
20. Thickness of human articular cartilage in joints of the lower limb10.1136/ard.58.1.27 ↗
21. Loading and knee alignment have significant influence on cartilage MRI T2 in porcine knee joints10.1016/j.joca.2010.05.002 ↗
22. Temple · 2016
23. A multi-purpose force-controlled loading device for cartilage and meniscus functionality assessment using advanced MRI techniques10.1016/j.jmbbm.2019.103428 ↗
24. The measurement of joint mechanics and their role in osteoarthritis genesis and progression
25. Yung · 2020 · 2750