Abstract

Background and objective Fatigue failure of the humeral stem is a severe long-term failure after shoulder arthroplasty, causing harm to patients and resulting in complex revision surgeries. However, there are few studies on humeral stem fatigue testing, and corresponding testing standards have not been established. Therefore, this study aims to investigate the fatigue performance of the humeral stem by establishing an efficient numerical simulation method. Methods Material properties are obtained by uniaxial tensile and fatigue tests. A parameterized static analysis program was written, and an automated fatigue numerical simulation platform was established using Abaqus, Fe-safe, and Isight in combination, enabling the establishment of a numerical simulation method for the fatigue performance of the humeral stem. Result Standard testing conditions include an 8 mm diameter humeral stem, a 40-21B humeral head, an 8° tilt angle, and a 2 mm fillet radius. Further research found that the fatigue life of the humeral stem decreases with increasing patient weight, and patients should control their weight after surgery. Conclusions The established automated fatigue numerical simulation platform avoids repetitive operations and efficiently completes large-scale calculations, guiding preoperative humeral stem selection and testing.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Shoulder Injury and Treatment · Total Knee Arthroplasty Outcomes · Metal and Thin Film Mechanics

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (34)

Show source-provided references
1. Glenohumeral contact forces10.1243/0954411001535660 ↗
2. Mechanical testing of shoulder prostheses and recommendations for glenoid design10.1067/mse.2000.105451 ↗
3. Glenohumeral contact forces10.1243/0954411001535660 ↗
4. Realistic loads for testing hip implants
5. The three-dimensional geometry of the proximal humerus: implications for surgical technique and prosthetic design10.1302/0301-620X.79B5.0790857 ↗
6. Fatigue analysis of cemented hip prosthesis: damage accumulation scenario and sensitivity analysis10.1016/S0142-1123(01)00204-3 ↗
7. Design, manufacture, and fatigue analysis of lightweight hip implants
8. Mid- to long-term follow-up of total shoulder arthroplasty using a keeled glenoid in young adults with primary glenohumeral arthritis10.1016/j.jse.2012.09.016 ↗
9. Experimental research on uniaxial tensile behavior and cyclic deformation behavior of TA17
10. Fatigue lifetime prediction of a reduced-diameter dental implant system: numerical and experimental study10.1016/j.dental.2018.06.002 ↗
11. The influence of rotator cuff disease on the results of shoulder arthroplasty for primary osteoarthritis: results of a multicenter study10.2106/00004623-200212000-00018 ↗
12. A three-dimensional non-linear finite element study of the effect of cement-prosthesis debonding in cemented femoral total hip components10.1016/0021-9290(91)90021-E ↗
13. The effects of glenoid component alignment variations on cement mantle stresses in total shoulder arthroplasty10.1016/j.jse.2004.04.008 ↗
14. Python-based secondary development of ABAQUS and its application in ana⁃ lyzing shallow-buried unsymmetrical loading tunnels
15. Parametric design optimisation of proximal humerus plates based on finite element method10.1007/s10439-018-02160-6 ↗
16. Secondary development of ABAQUS and its application to static and dynamic analyses of earth-rockfill dam
17. The effects of static, dynamic and fatigue behavior on three-dimensional shape optimization of hip prosthesis by finite element method10.1016/j.matdes.2006.08.012 ↗
18. Development and validation of statistical models of femur geometry for use with parametric finite element models10.1007/s10439-015-1307-6 ↗
19. The thermomechanical behavior for aluminum alloy under uniaxial tensile loading10.1016/S0921-5093(99)00069-6 ↗
20. Fatigue analysis method of used shaft parts based on ABAQUS/fe-Safe
21. 2nd: the classic: articular replacement for the humeral head10.1007/s11999-011-1944-5 ↗
22. Comparison of fatigue behaviour of eight different hip stems: a numerical and experimental study10.4236/jbise.2011.410080 ↗
23. Wyss UP: hip stem fatigue test prediction10.1016/j.ijfatigue.2008.10.005 ↗
24. Surgery for shoulder osteoarthritis: a cochrane systematic review10.3899/jrheum.101008 ↗
25. Surgery for shoulder osteoarthritis: a Cochrane systematic review10.3899/jrheum.101008 ↗
26. Bone–cement interface of the glenoid component: stress analysis for varying cement thickness10.1016/j.clinbiomech.2005.03.010 ↗
27. Total shoulder arthroplasty versus hemiarthroplasty for glenohumeral arthritis: a systematic review of the literature at long-term follow-up10.4103/0973-6042.118915 ↗
28. Implant survivorship and clinical outcomes of the Bigliani-Flatow shoulder prosthesis at a mean of five years: a post-marketing surveillance study from three centres10.1016/j.jajs.2014.11.001 ↗
29. Biomechanical evaluation of all-polyethylene pegged bony ingrowth glenoid fixation techniques on implant micromotion
30. Mental health, preoperative disability, and postoperative outcomes in patients undergoing shoulder arthroplasty10.1016/j.jse.2018.02.066 ↗
31. Glenohumeral motion in patients with rotator cuff tears: a comparison of asymptomatic and symptomatic shoulders10.1016/S1058-2746(00)90002-8 ↗
32. Glenoid implant orientation and cement failure in total shoulder arthroplasty: a finite element analysis10.1016/j.jse.2012.09.007 ↗
33. An integrated system for durability and reliability synthesis using iSIGHT and FE-Fatigue10.4271/2003-01-1220 ↗
34. Fatigue failure of a shoulder hemiarthroplasty stem: a case report10.1016/S1058-2746(03)00049-1 ↗