journal article·2018-06

Test method for mechanical properties of implantable catheters according to DIN 10555-3

J.D. Busch, H. Schröder, K. Sellenschloh, G. Adam, H. Ittrich, G. Huber

Publisher page

Abstract

To enable causal analysis of port catheter failure, this study aimed to develop an experimental setup for uniaxial tensile tests that addresses the specific requirements of highly elastic medical catheters; and to quantify parameters of the catheters' mechanical competence with respect to effects of artificial aging. Segments of 6F-polyurethane catheters were tested in their native status, after chemical and after mechanical aging. Tension experiments were performed with a rate of 220 mm/min until catheter failure. Material behavior was analyzed based on load cell measurements of the universal test system and an additional optical distance registration. The Young's modulus, the ultimate stress and the ultimate strain were determined. Chemical aging significantly decreased Young's modulus (84%; p = 0.001) and ultimate stress (83%; p < 0.001), whereas mechanical aged samples demonstrated similar results for the Young's modulus (p = 0.772) and a non-significant rise of ultimate stress (13%; p = 0.128). Ultimate strain did not differ significantly regardless of the pretreatment. The results proof reliability, reproducibility and sensitivity to quantify artificial aging induced variations and also promise to detect deviations in material features caused by long-term clinical usage of catheters.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Central Venous Catheters and Hemodialysis · Cardiac pacing and defibrillation studies · Intravenous Infusion Technology and Safety

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (20)

Show source-provided references
1. Mechanical integrity of hemodialysis catheters after exposure to a novel catheter lock solution10.1086/656749 ↗
2. Effect of ethanol/trisodium citrate lock on the mechanical properties of carbothane hemodialysis catheters10.5414/CNP65342 ↗
3. Catheter fracture and embolization associated with arm implantation of the cook vital port10.1016/j.carj.2012.03.001 ↗
4. Radiology-implanted arm ports: a review of the literature10.1016/j.java.2015.02.002 ↗
5. Material failure of silicone catheter lines: a retrospective review of partial and Complete ruptures in 553 patients
6. Cauich-Rodríguez J.V., Hernandez-Sánchez F., 2013. Degradation of Polyurethanes for Cardiovascular Applications.
7. Clinical biodurability of aliphatic polyether based polyurethanes as peritoneal dialysis catheters10.1097/01.MAT.0000065467.87011.3B ↗
8. The effects of prolonged ethanol exposure on the mechanical properties of polyurethane and silicone catheters used for intravascular access10.1086/502607 ↗
9. DIN EN ISO 10555-3:2013-11, 2013. Intravascular catheters - Sterile and single-use catheters - Part 3: Central venous catheters (ISO 10555-3:2013); English Version EN ISO 10555-3.
10. Catheter fracture: a rare complication of totally implantable subclavian venous access devices10.1002/(SICI)1096-9098(199607)62:3<222::AID-JSO14>3.0.CO;2-8 ↗
11. is radiologic placement of an arm port mandatory in oncology patients? Analysis of a large bi-institutional experience10.1002/cncr.23040 ↗
12. McGinn A.N., Comes V.C., 2013. Chemical Resistance and Stress Cracking in Polyurethanes [Internet]. mddionline.com [cited 2016 Nov 15]. Available from: 〈http://www.mddionline.com/article/chemical-resistance-and-stress-cracking-polyurethanes?Page=1〉.
13. Polyurethane central venous catheters, hydrochloric acid and 70% ethanol: a safety evaluation10.1177/0310057X9702500404 ↗
14. 2013 · 563
15. A new peripherally implanted subcutaneous permanent central venous access device for patients requiring chemotherapy10.1200/JCO.1993.11.11.2181 ↗
16. Understanding the biodegradation of polyurethanes: from classical implants to tissue engineering materials10.1016/j.biomaterials.2005.05.079 ↗
17. Biostable polyurethane elastomers
18. Vermette · 2001 · 6
19. Biodegradation of polyurethane under fatigue loading10.1002/jbm.a.10565 ↗
20. Catheter fracture of intravenous ports and its management10.1007/s00268-011-1200-x ↗