journal article·2010-10

In vitro evaluation of mechanical characteristics of a novel Patent Ductus Arteriosus (PDA) occlusion device with a new lateral pressure measurement study

Megha Agrawal, Kunal Khilar, Jayesh R. Bellare

Publisher page

Abstract

Persistence of the ductus arteriosus (DA), connecting the aorta to the pulmonary artery, after birth leads to a congenital heart disease known as Patent Ductus Arteriosus (PDA). The objective of this study was to propose a new intraductal design for PDA occlusion and determine the efficacy of the new device by means of an in vitro mechanical evaluation. A novel intraductal design device called the Novel design 36 (ND36) was proposed, fabricated and tested. Mechanical tests included tests to determine the mechanical characteristics, force-displacement characteristics, stiffness, device slenderization, minimum diameter attained, insertion force required to pull device at site and lateral pressure exerted on the PDA walls. Comparative analysis was carried out with the Amplatzer Duct Occluder (AGA Medical Corporation, Golden Valley, MN). The results showed that the ND36 when fully stretched was shorter than the ADO implying an easier in vivo movement. The minimum diameter attained by ND36 is 2.52 mm which is compatible with standard catheter bores of 6-8 Fr (2-2.7 mm) bore sizes. A new setup was devised in order to carry out comparative lateral pressure measurements between the ND36 and ADO. The results showed that the ND36 exerted a pressure of 10 KPa (=75 mm Hg) and the ADO exerted 80 KPa (=600 mm Hg) over the walls of the tubes which is a physiologically promising result for the ND36 as it shows that the ND36 would exert lesser stress on the walls of the duct. The results suggest that the novel intraductal design (ND36) would prove to be a safer, reliable and effective device for PDA occlusion.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Cardiovascular Conditions and Treatments · Congenital Heart Disease Studies · Mechanical Circulatory Support Devices

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (12)

Show source-provided references
1. Agrawal, M., 2009. Patent ductus arteriosus closure device: development and testing. M.Tech Thesis. IIT, Mumbai.
2. Device closure of patent ductus arteriosus
3. Natural history of persistent ductus arteriosus10.1136/hrt.30.1.4 ↗
4. Transcatheter Amplatzer occlusion and surgical closure of patent ductus arteriosus: comparison of effectiveness and costs in a low-income country10.1007/s00246-009-9440-3 ↗
5. Interventional pediatric cardiology: closure of a very large patent ductus arteriosus using the Amplatzer duct occluder
6. Safety and efficacy of using 0.052-inch Gianturco coil for closure of large (≥4 mm) patent ductus arteriosus
7. Transcatheter closure of the patent ductus arteriosus using an Amplatzer duct occluder in adults10.1536/jhj.42.533 ↗
8. Interventional pediatric cardiology: transcatheter closure of patent ductus arteriosus in Chinese adults: immediate and long-term results
9. Nonsurgical closure of patent ductus arteriosus: clinical application of the Rashkind PDA occluder system
10. Congenital heart diseases in India: a status report10.1007/BF02724185 ↗
11. Shah, V.Y., 2007. Patent ductus arteriosus closure device: modeling, simulation and manufacturing methods. M.Tech Thesis. IIT, Mumbai.
12. Transcathetral closure of patent ductus arteriosus