journal article·2016-09

A theoretical model to predict tensile deformation behavior of balloon catheter

Mitsugu Todo, Keiji Yoshiya, Takuya Matsumoto

Publisher page

Abstract

In this technical note, a simple theoretical model was proposed to express the tensile deformation and fracture of balloon catheter tested by the ISO standard using piece-wise linear force-displacement relations. The model was then validated by comparing with the tensile force-displacement behaviors of two types of typical balloon catheters clinically used worldwide. It was shown that the proposed model can effectively be used to express the tensile deformation behavior and easily be handled by physicians who are not familiar with mechanics of materials.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Central Venous Catheters and Hemodialysis · Hemodynamic Monitoring and Therapy · Vascular Procedures and Complications

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (11)

Show source-provided references
1. The mechanical properties of intact and traumatized epidural catherters
2. Balloon rupture during coronary angioplasty causing dissection and intramural hematoma of the coronary artery; a case report
3. Successful transradial retrieval of broken catheter fragment during transradial coronary angiography
4. Percutaneous retrieval of a fractured guide catheter using contralateral snaring
5. A rare instructive complication of balloon catheter fracture during percutaneous coronary intervention10.1007/s12928-015-0322-8 ↗
6. Central venous catheter use Part 1: mechanical complications10.1007/s00134-001-1154-9 ↗
7. Removing stuck, ruptured angioplasty balloon catheter10.1148/radiology.139.1.7208931 ↗
8. Rational use of rotational atherectomy in calcified lesions in the drug-eluting stent era: review of the evidence and current practice10.1016/j.carrev.2014.12.011 ↗
9. Trehan · Catheter. Cardiovasc. Interv. · 2003 · 58 · 473
10. Pinhole balloon rupture and stuck stent: case report of a new and simple bailout technique for incomplete stent dilatation caused by rupture from a highly calcified lesion10.1007/s12928-014-0247-7 ↗
11. The stuck catheter: a hazardous twist to the meaning of permanent catheters10.5301/jva.5000392 ↗