Biomechanical evaluation of load transfer in Talus focal resurfacing implant: experimental and finite element models
Source abstract
Abstract
INTRODUCTION: Osteochondral lesions (OTLs) are common injuries in ankle sprains, occurring in up to 70 % of patients. Recovery options are limited due to the talus bone's poor vascularization. This study aimed to develop a computational model of an ankle joint to evaluate cartilage stress in the presence of an OTL and assess the biomechanical effectiveness of a focal resurfacing prosthesis in reducing joint stress. MATERIALS AND METHODS: Both in silico and physical model models were developed based on a previous case study involving a patient with a medial OTL in the right ankle. A 10 mm diameter focal resurfacing prosthesis was tested in two positions: proud (+0.5 mm) and recessed (-0.5 mm), as recommended in clinical guidelines. Cartilage stress and trabecular bone strain in the talus were evaluated. The in vitro model, produced via additive manufacturing and instrumented with strain gauges, was used to validate the finite element (FE) model by comparing measured and simulated strains. RESULTS: The FE model showed strong agreement with experimental data, with a correlation coefficient of 0.88. Both the lesion and prosthesis placement influenced talar cartilage stress. The OTL increased stress by approximately 23 % near the lesion site and 8 % in more distal regions. The highest cartilage stress (7.20 MPa) occurred with the prosthesis in the recessed position. Prosthesis positioning significantly affected cartilage stress distribution (p < 0.001). CONCLUSIONS: While fixation of the focal prosthesis remains challenging, placing it slightly proud can help reduce stress on talar cartilage. However, excessive proud positioning increases stress on tibial cartilage and should be avoided. Proper prosthesis placement is critical for optimal stress reduction. Furthermore, fixation in the talus reduces trabecular bone strain, potentially mitigating bone loss and enhancing implant stability.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Foot and Ankle Surgery · Total Knee Arthroplasty Outcomes · Nasal Surgery and Airway Studies
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (46)
Show source-provided references
1. Effect of implantation accuracy on ankle contact mechanics with a metallic focal resurfacing implant10.2106/JBJS.I.00431 ↗
2. 2018
3. Osteochondral lesion of the talus: what are we talking about?
4. Partial hemiresurfacing of osteochondral defects of the talus? Surgical technique and preliminary report from four patients10.1055/s-0044-100731 ↗
5. Biomechanical finite element analysis of typical tibiotalar arthrodesis
6. Finite element modelling and experimental validation of a total implanted shoulder joint10.1016/j.cmpb.2021.106158 ↗
7. Finite element model validation based on an experimental model of the intact shoulder joint10.1016/j.medengphy.2020.11.004 ↗
8. Development and implantation of a universal talar prosthesis10.3389/fsurg.2019.00063 ↗
9. An analysis of changes in in vivo cartilage thickness of the healthy ankle following dynamic activity10.1016/j.jbiomech.2016.05.030 ↗
10. Mid-term results after treatment of complex talus osteochondral defects with HemiCAP implantation10.1016/j.fas.2019.05.003 ↗
11. Results of HemiCAP implantation as a salvage procedure for osteochondral lesions of the talus10.1053/j.jfas.2017.04.001 ↗
12. Bone's mechanostat: a 2003 update10.1002/ar.a.10119 ↗
13. Material models and properties in the finite element analysis of knee ligaments: a literature review10.3389/fbioe.2014.00054 ↗
14. Design of a lightweight universal talus implant using topology optimization10.3389/fbioe.2023.1228809 ↗
15. Osteochondral lesions of the talus: effect of defect size and plantarflexion angle on ankle joint stresses10.1177/0363546511434404 ↗
16. Solid modeling and static finite element analysis of the human tibia
17. Full-field comparisons between strains predicted by QCT-derived finite element models of the scapula and experimental strains measured by digital volume correlation10.1016/j.jbiomech.2020.110101 ↗
18. The effect of talus osteochondral defects of different area size on ankle joint stability: a finite element analysis
19. Investigation of the average shape and principal variations of the human talus bone using statistic shape model10.3389/fbioe.2020.00656 ↗
20. Development and validation of a novel ankle joint musculoskeletal model10.1007/s11517-023-03010-x ↗
21. HemiCAP® implantation after failed previous surgery for osteochondral lesions of the talus10.1016/j.fas.2020.02.008 ↗
22. Instability-associated changes in contact stress and contact stress rates near a step-off incongruity10.2106/JBJS.G.00127 ↗
23. Effects of implant orientation and implant material on tibia bone strain, implant–bone micromotion, contact pressure, and wear depth due to total ankle replacement10.1177/0954411918823811 ↗
24. Short-term results of hemiarthroplasty of the ankle joint for talar-sided cartilage loss10.1177/24730114221148172 ↗
25. Tetrahedral versus hexahedral finite elements in numerical modelling of the proximal femur10.1016/j.medengphy.2005.12.006 ↗
26. Predictions of Birmingham hip resurfacing implant offset - in vitro and numerical models10.1080/10255842.2018.1556973 ↗
27. The effect of osteochondral lesion size and ankle joint position on cartilage behavior - numerical and in vitro experimental results10.1016/j.medengphy.2021.10.014 ↗
28. Reilingh
29. Knee articular cartilage repair and restoration techniques: a review of the literature
30. Rikken · 2022 · 273
31. The biomechanical influence of defected cartilage on the progression of osteochondral lesions of the talus: a three-dimensional finite element analysis10.1111/os.13753 ↗
32. Osteochondritis dissecans10.2106/00004623-199603000-00018 ↗
33. A comparison of hexahedral and tetrahedral finite elements for biphasic analysis of cartilage micromechanics
34. Force and motion analysis of the normal, diseased, and prosthetic ankle joint10.1097/00003086-197709000-00027 ↗
35. A statistical analysis of human talar shape and bone density distribution10.1002/jor.25835 ↗
36. Comparison of hexahedral and tetrahedral elements in finite element analysis of the foot and footwear10.1016/j.jbiomech.2011.05.006 ↗
37. Tan · 2023 · 1
38. Constitutive modeling of cartilaginous tissues: a review
39. Four decades of finite element analysis of orthopaedic devices: where are we now and what are the opportunities?10.1016/j.jbiomech.2014.12.019 ↗
40. Development and experimental validation of a finite element model of total ankle replacement10.1016/j.jbiomech.2013.12.022 ↗
41. Geometric analysis of the talus and development of a generic talar prosthetic10.1016/j.fas.2016.12.002 ↗
42. A feature-based statistical shape model for geometric analysis of the human talus and development of universal talar prostheses10.1111/joa.13552 ↗
43. The use of HemiCAP for the treatment of osteochondral lesions10.1053/j.oto.2014.04.003 ↗
44. Fibula and its ligaments in load transmission and ankle joint stability10.1097/00003086-199609000-00034 ↗
45. Finite element analysis of the biomechanical effect of bone resection depth in the distal tibia after total ankle replacement10.1007/s40846-022-00722-7 ↗
46. Anatomic ankle implant can provide better tibiotalar joint kinematics and loading10.1016/j.medengphy.2022.103789 ↗