journal article·2018-09Open access

Qualified viewing space determination of near‐eye and head‐up displays

Richard L. Austin, Bruce S. Denning, Brian C. Drews, Victor B. Fedoriouk, Robert C. Calpito

Abstract

Abstract A method for determining the extents of a qualified viewing space (QVS) based on repeatable and reproducible luminance measurements of augmented and virtual reality near‐eye displays is described. This QVS mapping can also use other display performance metrics such as (1) Michelson contrast, (2) modulation transfer function, or (3) color as boundary condition parameters. We describe the use of a 4‐mm‐diameter entrance pupil, 1° to 2° field of view tele‐spectroradiometer, to determine the luminance and color uniformity of the virtual image. A 1‐mm‐diameter entrance pupil is used to map the QVS boundaries based on luminance at the center of the virtual image. The luminance measurement results from a pair of binocular augmented reality display glasses in three separate eye relief planes of the QVS of both eyes are presented. The data are further reduced to provide a perimeter profile of the QVS for the 50% of peak luminance boundary points in each eye relief plane.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Visual perception and processing mechanisms · Advanced Optical Imaging Technologies · Ophthalmology and Visual Impairment Studies

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Gamma Scientific San Diego CA USA

References (5)

Show source-provided references
1. D.Felloweset al “Near‐to‐eye display test station ” Proceedings of SPIE Vol.5079 (2003)10.1117/12.487767 ↗
2. R.Draperet al “Design considerations and preliminary performance evaluation for a technology demonstration of off‐the‐visor wide‐field‐of‐view HMD ” Proceedings of SPIE Vol.5079 (2003)10.1117/12.487859 ↗
3. 10.1002/jsid.53710.1002/jsid.537 ↗
4. 10.1002/sdtp.1093710.1002/sdtp.10937 ↗
5. 10.1002/sdtp.1180810.1002/sdtp.11808 ↗