journal article·2007-06

Evaluation of high‐resolution and mobile display systems for digital radiology in dark and bright environments using human and computational observers

Rebecca Vogel, Anindita Saha, Kish Chakrabarti, Aldo Badano

Publisher page

Abstract

Abstract— As digital display systems replace film traditionally used for reading radiographic images, resource‐intensive acceptance testing must be performed to ensure that quality meets and maintains desired specifications. If machine observers can replace human readers, whose performances are highly variable, the results will be more consistent and less costly. To be effective, however, the automated observers must track human performance. An approach for a model observer, validated with human readers, for the evaluation of the visibility of low‐contrast small targets in high‐resolution and mobile displays under different ambient illumination, will be described. The displays were tested using CDMAM‐like digital phantoms containing disks of varying diameters and contrasts on a flat background. For this task, we find the best indicator of display performance to be the display's ability to represent small luminance contrast, not resolution or pixel size. The results confirm that high‐resolution systems perform better under low illumination while illuminance has a minor impact on the mobile‐display performance. Finally, the results show that the machine observer tracks the performance of human readers. Machine observers with proper validation can replace humans in the acceptance testing procedures, saving the testers both time and money.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Digital Radiography and Breast Imaging · Advanced X-ray and CT Imaging · Optical Imaging and Spectroscopy Techniques

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (10)

Show source-provided references
1. A.Badano M.J.Flynn andJ.Kanicki “High‐fidelity display performance ”High‐Fidelity Medical Imaging Displays(SPIE Press Bellingham WA (2004) pp. 56.10.1117/3.2265081 ↗
2. A.Badano “Display systems ”AAPM/RSNA Tutorial on Equipment Selection: PACS Equipment Overview(pp. 879–889). RSNA (RadioGraphics 2004 No. 24:3).10.1148/rg.243035133 ↗
3. K.R.Bijkerk J.M.Lindeijer andM.A. O.Thijssen “Evaluation of the phantom image ”CDMAM‐Phantom type 3.2 A(University Hospital Nijmegen St. Radbound University 1993) pp.10–12.
4. 10.1118/1.135530810.1118/1.1355308 ↗
5. R.M.Gagne B.D.Gallas andK.J.Myers “Toward objective and quantitative evaluation of imaging systems using images of phantoms” (American Association of Physicists in Medicine 2006).10.1118/1.2140117 ↗
6. R.V.Metter M.Heath andL.Fletcher‐Heath “Applying the ‘European protocol for the quality control of the physical and technical aspects of mammography screening ’Threshold Contrast Visibility Assessment to Digital Systems(East Kodak Company Rochester NY 1999).
7. Myers K. J. · 2000 · 559
8. E.Sameiet al.“General guidelines for purchasing and acceptance testing of PACS equipment ”AAPM/RSNA Tutorial on Equipment Selection: PACS Equipment OverviewRSNA RadioGraphics No. 24:1 pp.313–334.10.1148/rg.241035137 ↗
9. J.A.Thomas K.Chakrabarti R.Kaczmarek andA.Romanyukha “Contrast‐detail phantom scoring methodology ”American Association of Physicists in Medicine(2005).10.1118/1.1862097 ↗
10. K.C.Young J.J.Cook andJ.M.Oduko “Automated and human determination of threshold contrast for digital mammography systems ”Digital Mammography(Springer 2006) pp.266–280.10.1007/11783237_37 ↗