journal article·2007-06

Influence of 8‐bit vs. 11‐bit digital displays on observer performance and visual search: A multi‐center evaluation

Elizabeth A. Krupinski, Khan Siddiqui, Eliot Siegel, Rasu Shrestha, Edward Grant, Hans Roehrig, Jiahua Fan

Publisher page

Abstract

Abstract— Medical‐grade monochrome monitors typically display 8 bits of data. This study determined if 11‐bit displays could improve observer performance and decrease use of window/level. 8‐ and 11‐bit displays from three manufacturers were used at three sites. Six radiologists at each site viewed 100 DR chest images (half with a pulmonary nodule) on both displays. Decisions, confidence, nodule location, viewing time, and window/level use were recorded. There was no significant difference in ROC Az as a function of bit depth. The average Az with 8 bits was 0.8284 and with 11 bits was 0.8253. There was a significant difference in viewing time favoring the 11‐bit displays. Window/level use did not differ. Eye position was recorded on a subset of images at one site. Cumulative dwell times for each decision category were lower with the 11‐bit than with the 8‐bit display. When tested with t‐tests for paired observations, the TP (t = 1.452, p = 0.1507), FN (t = 0.050, p = 0.9609), and FP (t = 0.042, p = 0.9676) were not statistically significant. The difference in the TN decisions was statistically significant (t = 1.926, p = 0.05). 8‐bit displays will not impact negatively diagnostic accuracy, but using 11‐bit displays may improve workflow efficiency.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Radiology practices and education · Digital Radiography and Breast Imaging · Radiation Dose and Imaging

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (35)

Show source-provided references
1. 10.1007/s10278-002-0025-410.1007/s10278-002-0025-4 ↗
2. 10.1007/s10278-002-0021-810.1007/s10278-002-0021-8 ↗
3. 10.1007/s10278-002-0022-710.1007/s10278-002-0022-7 ↗
4. 10.1007/s10278-002-0024-510.1007/s10278-002-0024-5 ↗
5. 10.1007/s10278-002-0026-310.1007/s10278-002-0026-3 ↗
6. E.L.SiegelandB.Reiner “Image workflow ” In: K.J. Dreyer A. Mehta and J.H. Thrall eds. PACS: A Guide to the Digital Revolution(Springer‐Verlag New York NY 2002) pp.161–190.
7. H.Roehrig “The monochrome cathode ray tube display and its performance ”Handbook of Medical Imaging Volume 3: Display and PACS Y. Kim and S.C. Horii eds. (SPIE Press Bellingham WA 2000) pp.155–220.10.1117/3.367205.ch4 ↗
8. F.M.Behlen B.M.Hemminger andS.C.Horii “Displays ”Handbook of Medical Imaging Volume 3: Display and PACS Y. Kim and S.C. Horii eds. (SPIE Press Bellingham WA 2000) pp.403–440.10.1117/3.367205.ch10 ↗
9. T.ChunnandJ.Honeyman Storage and Database Y. Kim and S.C. Horii eds. Handbook of Medical Imaging Volume 3: Display and PACS(SPIE Press Bellingham WA 2000) pp.365–401.10.1117/3.367205.ch9 ↗
10. C.F.Nodine H.L.Kundel J.B.Polikoff andL.C.Toto “Using eye movements to study decision making of radiologists ”Eye Movement Research: Physiological and Psychological Aspects G. Luer U. Lass and J. Shallo‐Hoffman eds. (CJ Hogrefe Publishers Lewiston NY 1987) pp.349–363.
11. C.F.NodineandH.L.Kundel “The cognitive side of visual search ”Eye Movements: From Physiology to Cognition J. K. O'Regan and A. Levy‐Schoen eds. (Elsevier Publishers Amsterdam Netherlands 1987) pp.573–582.10.1016/B978-0-444-70113-8.50081-3 ↗
12. B.I.Reiner N.Aykin E.A.Krupinski F.Sauer andE.L.Siegel “Optimizing the man/machine interface ”Special Session SCAR 2004(2004).
13. B.J.Erickson E.L.Siegel A.Hartman andE.A.Krupinski “The radiology interpretation process: Peering into the mind of the radiologist ”Special Session SCAR 2004(2004).
14. B.J.Erickson C.P.Langlotz E.Burnside andC.C.Jaffe “Radiologist decision support ”Senior‐Level Session SCAR 2004(2004).
15. D.L.Weiss D.Liu andC.P.Langlotz “Speech recognition and structured reporting ”Senior‐Level Session SCAR 2004(2004).
16. M.Shaw P.G.Nagy E.L.Siegel andB.I.Reiner “Productivity and workflow ”Senior‐Level Session SCAR 2004(2004).
17. E.L.Siegel B.Rostenberg M.A.Stein K.C.Johnson andA.Hedge “Design considerations in a filmless enterprise ”Senior‐Level Session SCAR 2004(2004).
18. 10.1117/12.13595610.1117/12.135956 ↗
19. 10.1117/3.35325410.1117/3.353254 ↗
20. 10.1117/12.43550910.1117/12.435509 ↗
21. 10.1097/00004424-199209000-0001510.1097/00004424-199209000-00015 ↗
22. 10.1117/12.53332810.1117/12.533328 ↗
23. 10.1016/S1076-6332(05)80287-910.1016/S1076-6332(05)80287-9 ↗
24. 10.1089/tmj.1.1996.2.10110.1089/tmj.1.1996.2.101 ↗
25. 10.1016/S1076-6332(05)80381-210.1016/S1076-6332(05)80381-2 ↗
26. 10.1016/S1076-6332(05)80780-910.1016/S1076-6332(05)80780-9 ↗
27. 10.1016/S1076-6332(99)80191-310.1016/S1076-6332(99)80191-3 ↗
28. 10.1016/S1076-6332(00)80437-710.1016/S1076-6332(00)80437-7 ↗
29. 10.1016/S1076-6332(03)80308-210.1016/S1076-6332(03)80308-2 ↗
30. 10.3758/BF0320358410.3758/BF03203584 ↗
31. 10.1117/12.59376610.1117/12.593766 ↗
32. K.M.Siddiqui E.L.Siegel B.I.Reiner A.E.Musk F.J.Hooper andR.Moffitt “11‐bitvs.8‐bit monochrome LCD for interpretation of 12‐bit DICOM images ”Proc 90th Annual Meeting of the Radiological Society of North America(2004).
33. K.M.Siddiqui E.L.Siegel B.I.Reiner A.E.Musk J.Johnson andO.Crave “Quantitative differences between 8‐bit and 11‐bit monochrome DICOM images: A visual discrimination model analysis ”Proc 90th Annual Meeting of the Radiological Society of North America(2004).
34. 10.1117/12.46693010.1117/12.466930 ↗
35. 10.1118/1.185503110.1118/1.1855031 ↗