journal article·2006-11

Color measurement methods for medical displays

Anindita Saha, Hongye Liang, Aldo Badano

Publisher page

Abstract

Abstract— The effect of different measurement methods on the characterization of display color, maximum color difference, and luminance uniformity of medical liquid‐crystal displays are reported. We use a telescopic colorimeter and a custom‐designed collimated probe with an internal lens attached to a spectrometer. The maximum color‐difference variations were found to be between 0.0047 to 0.0073, in the same range as variations among methods, displays, and screen locations.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Color Science and Applications · Ergonomics and Musculoskeletal Disorders

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (11)

Show source-provided references
1. Video Electronics Standards Association (VESA) Flat Panel Display Measurements Standard Working Group. Flat panel display measurements standard version 2.0. Technical report VESA May 2003.
2. P.A.Downen “A review of popular FPD measurement standards ”Proc ADEAC 5–8 (2004).
3. Standard Panel Working Group (SPWG).SPWG Notebook Panel Specification Version 3.5 Technical report SPWG(March 2005).
4. TCO Development. TCO '05 Desktop Computers. Technical Report TCO(June 2005).
5. 10.1118/1.186115910.1118/1.1861159 ↗
6. 10.1889/1.183243610.1889/1.1832436 ↗
7. E.F.KelleyandA.Badano “Characterization of luminance probe for accurate contrast measurements in medical displays ” Technical Report NISTIR 6974 NIST (2003).10.6028/NIST.IR.6974 ↗
8. R.VanEngen K.Young H.Bosmans andM.Thijssen “Addendum on Digital Mammography ” European Reference Organisation for Quality Assured Breast Screening and Diagnostic Services 3rd ed. (November 2003).
9. 10.1364/AO.39.00205910.1364/AO.39.002059 ↗
10. 10.1364/AO.43.00499910.1364/AO.43.004999 ↗
11. The slit plate was provided by Edward F. Kelley (NIST).