Versailles Project on Advanced Materials and Standards interlaboratory study on intensity calibration for x-ray photoelectron spectroscopy instruments using low-density polyethylene

Benjamen P. Reed, David J. H. Cant, Steve J. Spencer, Abraham Jorge Carmona-Carmona, Adam Bushell, Alberto Herrera-Gómez, Akira Kurokawa, Andreas Thissen, Andrew G. Thomas, Andrew J. Britton, Andrzej Bernasik, Anne Fuchs, Arthur P. Baddorf, Bernd Bock, Bill Theilacker, Bin Cheng, David G. Castner, David J. Morgan, David Valley, Elizabeth A. Willneff, Emily F. Smith, Emmanuel Nolot, Fangyan Xie, Gilad Zorn, Graham C. Smith, Hideyuki Yasufuku, Jeffery L. Fenton, Jian Chen, Jonathan D. P. Counsell, Jörg Radnik, Karen J. Gaskell, Kateryna Artyushkova, Li Yang, Lulu Zhang, Makiho Eguchi, Marc Walker, Mariusz Hajdyła, Mateusz M. Marzec, Matthew R. Linford, Naoyoshi Kubota, Orlando Cortazar-Martínez, Paul Dietrich, Riki Satoh, Sven L. M. Schroeder, Tahereh G. Avval, Takaharu Nagatomi, Vincent Fernandez, Wayne Lake, Yasushi Azuma, Yusuke Yoshikawa, Alexander G. Shard

Abstract

We report the results of a Versailles Project on Advanced Materials and Standards interlaboratory study on the intensity scale calibration of x-ray photoelectron spectrometers using low-density polyethylene (LDPE) as an alternative material to gold, silver, and copper. An improved set of LDPE reference spectra, corrected for different instrument geometries using a quartz-monochromated Al Kα x-ray source, was developed using data provided by participants in this study. Using these new reference spectra, a transmission function was calculated for each dataset that participants provided. When compared to a similar calibration procedure using the NPL reference spectra for gold, the LDPE intensity calibration method achieves an absolute offset of ∼3.0% and a systematic deviation of ±6.5% on average across all participants. For spectra recorded at high pass energies (≥90 eV), values of absolute offset and systematic deviation are ∼5.8% and ±5.7%, respectively, whereas for spectra collected at lower pass energies (<90 eV), values of absolute offset and systematic deviation are ∼4.9% and ±8.8%, respectively; low pass energy spectra perform worse than the global average, in terms of systematic deviations, due to diminished count rates and signal-to-noise ratio. Differences in absolute offset are attributed to the surface roughness of the LDPE induced by sample preparation. We further assess the usability of LDPE as a secondary reference material and comment on its performance in the presence of issues such as variable dark noise, x-ray warm up times, inaccuracy at low count rates, and underlying spectrometer problems. In response to participant feedback and the results of the study, we provide an updated LDPE intensity calibration protocol to address the issues highlighted in the interlaboratory study. We also comment on the lack of implementation of a consistent and traceable intensity calibration method across the community of x-ray photoelectron spectroscopy (XPS) users and, therefore, propose a route to achieving this with the assistance of instrument manufacturers, metrology laboratories, and experts leading to an international standard for XPS intensity scale calibration.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Electron and X-Ray Spectroscopy Techniques · X-ray Spectroscopy and Fluorescence Analysis · Nuclear Physics and Applications

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

National Physical Laboratory 1 , Hampton Road, Teddington TW11 0LW, United Kingdom

CINVESTAV-Unidad Queretaro 2 , Queretaro 76230, Mexico

Thermo Fisher Scientific (Surface Analysis) 3 , East Grinstead RH19 1XZ, United Kingdom

National Metrology Institute of Japan (NMIJ), National Institute of Advanced Industrial Science and Technology (AIST) 4 , 1-1-1 Higashi, Tsukuba, Ibaraki 305-8565, Japan

SPECS Surface Nano Analysis GmbH 5 , Voltastraße 5, 13355 Berlin, Germany

School of Materials, Photon Science Institute and Sir Henry Royce Institute, Alan Turing Building, University of Manchester 6 , Oxford Road, Manchester M13 9PL, United Kingdom

Versatile X-ray Spectroscopy Facility, School of Chemical and Process Engineering, University of Leeds 7 , Leeds LS2 9JT, United Kingdom

Academic Centre for Materials and Nanotechnology, AGH University of Science and Technology 8 , 30-059 Kraków, Poland

Robert Bosch GmbH, Robert-Bosch-Campus 9 , 71272 Renningen, Germany

Center for Nanophase Materials Sciences, Oak Ridge National Laboratory 10 , 1 Bethel Valley Road, Oak Ridge, Tennessee 37830

Tascon GmbH 11 , Mendelstr. 17, D-48149 Münster, Germany

Medtronic 12 , 710 Medtronic Parkway, LT240, Fridley, Minnesota 55432

Analysis and Testing Center, Beijing University of Chemical Technology 13 , Beijing 100029, People’s Republic of China

National ESCA and Surface Analysis Center for Biomedical Problems, Department of Bioengineering and Chemical Engineering, University of Washington 14 , Seattle, Washington 98195

Cardiff Catalysis Institute, School of Chemistry, Cardiff University 15 , Main Building, Cardiff CF10 3AT, United Kingdom

Physical Electronics Inc. 16 , East Chanhassen, Minnesota 55317

Versatile X-ray Spectroscopy Facility, School of Design, University of Leeds 17 , Leeds LS2 9JT, United Kingdom

Nanoscale and Microscale Research Centre, University of Nottingham 18 , Nottingham NG7 2RD, United Kingdom

CEA-LETI 19 , 17 rue des Martyrs, 38054 Grenoble, France

Instrumental Analysis & Research Center, Sun Yat-sen University 20 , Guangzhou 510275, People’s Republic of China

GE Research 21 , 1 Research Circle, K1 1D7A, Niskayuna, New York 12309

Faculty of Science and Engineering, University of Chester 22 , Thornton Science Park, Chester CH2 4NU, United Kingdom

Materials Analysis Station, National Institute for Materials Science (NIMS) 23 , 1-2-1 Sengen, Tsukuba, Ibaraki 305-0044, Japan

Medtronic 24 , 6700 Shingle Creek Parkway, Brooklyn Center, Minnesota 55430

Kratos Analytical Ltd., Wharfside 25 , Trafford Wharf Road, Manchester M17 1GP, United Kingdom

Funding

Department for Business, Energy and Industrial Strategy, UK Government

References (28)

Show source-provided references
1. Microsc. Today · 2016 · 24 · 1610.1017/S1551929516000080 ↗
2. Surf. Sci. Rep. · 2002 · 47 · 3310.1016/S0167-5729(02)00031-6 ↗
3. J. Electron Spectrosc. · 2000 · 107 · 13110.1016/S0368-2048(00)00095-5 ↗
4. Atom. Data Nucl. Data · 1985 · 32 · 110.1016/0092-640X(85)90016-6 ↗
5. J. Electron Spectrosc. · 1976 · 8 · 12910.1016/0368-2048(76)80015-1 ↗
6. Nature · 2016 · 533 · 35310.1038/nature17990 ↗
7. J. Surf. Anal. · 2019 · 26 · 9410.1384/jsa.26.94 ↗
8. Microsc. Microanal. · 2020 · 26 · 110.1017/S1431927619015332 ↗
9. International Organization for Standardization · 2010
10. International Organization for Standardization · 2005
11. Surf. Interface Anal. · 1993 · 20 · 17910.1002/sia.740200302 ↗
12. Surf. Interface Anal. · 2005 · 37 · 58910.1002/sia.2056 ↗
13. Rev. Sci. Instrum. · 2009 · 80 · 05310810.1063/1.3131631 ↗
14. J. Electron Spectrosc. · 2018 · 222 · 12210.1016/j.elspec.2017.07.004 ↗
15. Acta Phys. Pol. A · 1992 · 81 · 15910.12693/APhysPolA.81.159 ↗
16. Surf. Interface Anal. · 1990 · 15 · 32310.1002/sia.740150505 ↗
17. J. Electron Spectrosc. · 1989 · 48 · 5510.1016/0368-2048(89)80007-6 ↗
18. J. Electron Spectrosc. · 2019 · 233 · 5110.1016/j.elspec.2019.03.008 ↗
19. J. Electron Spectrosc. · 1995 · 71 · 19110.1016/0368-2048(94)02275-5 ↗
20. Surf. Interface Anal. · 1993 · 20 · 24310.1002/sia.740200309 ↗
21. Surf. Interface Anal. · 1990 · 15 · 75110.1002/sia.740151208 ↗
22. Surf. Interface Anal. · 2019 · 51 · 61810.1002/sia.6627 ↗
23. Surf. Interface Anal. · 2014 · 46 · 17510.1002/sia.5406 ↗
24. J. Electron Spectrosc. · 2010 · 182 · 8110.1016/j.elspec.2010.07.009 ↗
25. J. Vac. Sci. technol. A · 2020 · 38 · 06320910.1116/6.0000578 ↗
26. J. Electron Spectrosc. · 1999 · 104 · 7310.1016/S0368-2048(98)00317-X ↗
27. J. Vac. Sci. Technol. A · 2020 · 38 · 04120110.1116/1.5141395 ↗
28. See supplementary material at https://doi.org/10.1116/6.0000577 for (S1) Protocol for VAMAS interlaboratory study circulated to participants; (S2) Table showing important experimental information, percentage values of Δ (%) and Σ (%), and comments which describe issues or observations with the participants’ datasets or direct feedback from the participants; (S2 continued) Figures showing participants’ transmission functions calculated from LDPE and gold; (S3) Supporting figures including: a graphical representation of Ri, Δ, and Σ from Eqs. (2) and (3); and transmission functions calculated at NPL using LDPE prepared with differing surface topographies.