New scanning tunneling microscopy tip for measuring surface topography
Source abstract
Abstract
Electron-beam deposition inside scanning electron microscope (SEM) is used to prepare a new scanning tunneling microscope (STM) tip [electron-beam deposited (EBD) tip] which has a submicron diameter and a long straight sidewall. An EBD tip as small as less than 0.1 μm in diameter can be formed with careful beam focusing. This EBD tip is formed to be capable of providing a low-distortion STM image of deep grooves, and is quite useful in topographic measurements of microfabricated patterns.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Electron and X-Ray Spectroscopy Techniques · Anodic Oxide Films and Nanostructures · Optical Coatings and Gratings
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
ULSI Research Center, Toshiba Corporation, Saiwai-ku, Kawasaki 210, Japan
Electrotechnical Laboratory, Tsukuba Research Center, Tsukuba 305, Japan
References
Reference metadata was not supplied for this record.