The design, performance, and application of an atomic force microscope-based profilometer
Source abstract
Abstract
Rayleigh–Taylor instabilities during implosions of inertially confined fusion capsules affect capsule performance. During acceleration, surface imperfections grow and can, if large enough, lead to an asymmetric implosion or even shell breakup. For this reason, characterizing the topography of target capsules is extremely important. We have developed a profilometer based on an atomic force microscope combined with a precision rotary air bearing. Averaged one-dimensional (1D) surface height power spectra obtained with this instrument are converted to two-dimensional mode spectra that are used as input to hydrodynamic simulations. We describe the design of the system and its performance in terms of runout and repeatability. We also discuss the simulation of these measurements and the statistics involved in averaging 1D power spectra. Finally, we show the application of this measurement technique to capsules whose surfaces have been modified by laser ablation, resulting in a well-defined surface topography. This special case provides an excellent test for the system since the expected results are exactly calculable.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Laser-Plasma Interactions and Diagnostics · Laser-induced spectroscopy and plasma · Hemoglobin structure and function
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Lawrence Livermore National Laboratory, Livermore, California 94551
References