journal article·1989-07-01Open access

Intercomparison of vacuum standards between the Physikalisch Technische Bundesanstalt (Germany) and the National Physical Laboratory (India) using two spinning rotor gauges

J. K. N. Sharma, Pardeep Mohan, W. Jitschin, P. Röhl

Abstract

The paper describes the results of a comparison of low-pressure standards maintained at the National Physical Laboratory (NPL, India) and the Physikalisch Technische Bundesanstalt (PTB, FRG) over the range 10−3–2 Pa. The intercomparison was carried out by employing spinning rotor gauges shipped with a novel transport device. Both the laboratories, i.e., NPL and PTB, use static expansion systems for pressure generation. The gauge coefficients σ0,eff of the two gauges for Ar agree within ±0.5% for the two laboratories. This result proves agreement of the respective national standards in the above range of pressures. The small deviation of −0.4% between calibrations over a period of one year before and after air-parcel transportation from Berlin to New Delhi and back is indicative of a careful handling of the rotors during operation and a good performance of the employed transport device.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Scientific Measurement and Uncertainty Evaluation · Advanced Frequency and Time Standards · Calibration and Measurement Techniques

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

National Physical Laboratory, New Delhi-110012, India

Physikalisch Technische Bundesanstalt, Berlin, Germany

References